SwissPlus® OPBARecalls & reported problems
ZimVie US Corp LLC · Class 2 · Implant
All models of SwissPlus (10)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00889024015838), received between Sep 18, 2023 and Nov 20, 2023 — classified as 2 malfunctions.
The most frequently reported specific device problem is “Unintended Movement” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Unintended Movement2
Patient problems
- No Clinical Signs, Symptoms or Conditions2
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionNov 20, 2023Unintended Movement; Unintended Movement
NO ADDITIONAL OR CORRECTED INFORMATION TO REPORT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18173396 ↗MalfunctionSep 18, 2023Unintended Movement; Unintended Movement
IT WAS REPORTED THAT LOOSENESS SEEMS TO OCCUR WITH THIS ABUTMENT SCREW ABOUT THREE (3) DAYS BEFORE IT FRACTURES. TOOTH SITE #5. THIS EVENT CAPTURES SCREW LOOSENING.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17763735 ↗
Other ABUTMENT, IMPLANT, DENTAL, ENDOSSEOUS devices
- CERTAIN® GOLD-TITE® IUNIHGZimVie US Corp LLC · ABUTMENT, IMPLANT, DENTAL, ENDOSSEOUS151 reports
- Nobel Biocare® 36662Nobel Biocare AB · ABUTMENT, IMPLANT, DENTAL, ENDOSSEOUS2 recalls5 reports
- CERTAIN® IUNIHTZimVie US Corp LLC · ABUTMENT, IMPLANT, DENTAL, ENDOSSEOUS67 reports
- ATLANTIS® 35502Dentsply Ih, Inc. · ABUTMENT, IMPLANT, DENTAL, ENDOSSEOUS66 reports
- ATLANTIS® 35502Dentsply Ih, Inc. · Abutment, implant, dental, endosseous54 reports
- Esthetic Abutment 36669Nobel Biocare AB · ABUTMENT, IMPLANT, DENTAL, ENDOSSEOUS1 recall2 reports
- CERTAIN® BELLATEK® IEDAT5ZimVie US Corp LLC · ABUTMENT, IMPLANT, DENTAL, ENDOSSEOUS1 recall1 report
- Straumann Anatomic Healing Abutments XC (AHA) Healing ComponentsInstitut Straumann AG · Abutment, implant, dental, endosseous1 recall0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page