Skip to content

NA 47-2357-018-06Recalls & reported problems

Zimmer, Inc. · Class 2 · Implant

Data updated Oct 18, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
0
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

No MAUDE adverse event reports in our data reference this device identifier directly.

1 FDA recall record name this device's identifier, 1 still open.

20 other recalls cover devices cleared under the same FDA submission (K070906); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Recalls naming this device

  1. Mar 20, 2024Class IIOpen, Classified

    Thread form issue for the locking holes in that the locking screws may not properly mate with the plate. The improperly mated screw may not be readily recognizable by the user since the screw may not correctly lock, may result in delay in surgery or loss of fixation resulting further intervention

    Z-1716-2024 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K070906. One submission often covers many models, so check the product and lot details in each record.

  • Z-1710-2024 · Mar 20, 2024 · Open, Classified
    ZPLP (Zimmer Periarticular Locking Plate System) Distal Lateral Fibular Plate, Right, 6 Holes, 106 MM, Sterile- Indicated for temporary internal fixation and stabilization of oste
  • Z-1725-2024 · Mar 20, 2024 · Open, Classified
    ZPLP Distal Lateral Fibular Plate, Right, 10 Holes, 158 MM, Nonsterile- Indicated for temporary internal fixation and stabilization of osteotomies and fractures Item Number: 00-235
  • Z-1723-2024 · Mar 20, 2024 · Open, Classified
    ZPLP Distal Lateral Fibular Plate, Right, 6 Holes, 106 MM, Nonsterile- Indicated for temporary internal fixation and stabilization of osteotomies and fractures Item Number: 00-2357
  • Z-1715-2024 · Mar 20, 2024 · Open, Classified
    ZPLP Distal Lateral Fibular Plate, Left, 4 Holes, 80 MM, Sterile- Indicated for temporary internal fixation and stabilization of osteotomies and fractures Item Number: 7-2357-018-0
  • Z-1712-2024 · Mar 20, 2024 · Open, Classified
    ZPLP (Zimmer Periarticular Locking Plate System) Distal Lateral Fibular Plate, Right, 10 Holes, 158 MM, Sterile- Indicated for temporary internal fixation and stabilization of oste
  • Z-1728-2024 · Mar 20, 2024 · Open, Classified
    ZPLP Distal Lateral Fibular Plate, Left, 8 Holes, 132 MM, Nonsterile-Indicated for temporary internal fixation and stabilization of osteotomies and fractures Item Number: 00-2357-0
  • Z-1724-2024 · Mar 20, 2024 · Open, Classified
    ZPLP Distal Lateral Fibular Plate, Right, 8 Holes, 132 MM, Nonsterile- Indicated for temporary internal fixation and stabilization of osteotomies and fractures Item Number: 00-2357
  • Z-1714-2024 · Mar 20, 2024 · Open, Classified
    ZPLP (Zimmer Periarticular Locking Plate System) Distal Lateral Fibular Plate, Right, 16 Holes, 236 MM, Sterile- Indicated for temporary internal fixation and stabilization of oste
  • Z-1718-2024 · Mar 20, 2024 · Open, Classified
    ZPLP(Zimmer Periarticular Locking Plate System) Distal Lateral Fibular Plate, Left, 10 Holes, 158 MM, Sterile- Indicated for temporary internal fixation and stabilization of osteo
  • Z-1721-2024 · Mar 20, 2024 · Open, Classified
    ZPLP (Zimmer Periarticular Locking Plate System) Distal Lateral Fibular Plate, Left, 16 Holes, 236 MM, Sterile- Indicated for temporary internal fixation and stabilization of osteo
  • Z-1709-2024 · Mar 20, 2024 · Open, Classified
    ZPLP (Zimmer Periarticular Locking Plate System) Distal Lateral Fibular Plate, Right, 4 Holes, 80 MM, Sterile- Indicated for temporary internal fixation and stabilization of osteot
  • Z-1717-2024 · Mar 20, 2024 · Open, Classified
    ZPLP (Zimmer Periarticular Locking Plate System) Distal Lateral Fibular Plate, Left, 8 Holes, 132 MM, Sterile- Indicated for temporary internal fixation and stabilization of osteot
  • Z-1720-2024 · Mar 20, 2024 · Open, Classified
    ZPLP Distal Lateral Fibular Plate, Left, 14 Holes, 210 MM, Sterile- Indicated for temporary internal fixation and stabilization of osteotomies and fractures Item Number: 47-2357-01
  • Z-1713-2024 · Mar 20, 2024 · Open, Classified
    ZPLP (Zimmer Periarticular Locking Plate System) Distal Lateral Fibular Plate, Right, 12 Holes, 184 MM, Sterile- Indicated for temporary internal fixation and stabilization of oste
  • Z-1719-2024 · Mar 20, 2024 · Open, Classified
    ZPLP (Zimmer Periarticular Locking Plate System) Distal Lateral Fibular Plate, Left, 12 Holes, 184 MM, Sterile- Indicated for temporary internal fixation and stabilization of oste

…and 5 more.

Other Screw, fixation, bone devices

All Screw, fixation, bone devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page