Skip to content

NA 5983-40-48Recalls & reported problems

Zimmer, Inc. · Class 1

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
12
Deaths reported
0
Injuries reported
3
Recalls naming it
0

Summary

The FDA's MAUDE database contains 12 adverse event reports that list this exact device (DI 00889024218277), received between Feb 4, 2021 and Nov 19, 2025 — classified as 3 injuries, 9 malfunctions.

The most frequently reported specific device problem is “Fracture” (12 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 1 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 1 reportsJan 2025: 0 reportsFeb 2025: 1 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 1 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 2 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Fracture12
  • Mechanical Jam1

Patient problems

  • No Clinical Signs, Symptoms or Conditions9
  • Bone Fracture(s)2
  • Foreign Body In Patient1

Recent adverse event reports

The 12 most recent of 12 reports, in the reporter's own words as published by the FDA.

  • MalfunctionNov 19, 2025Fracture; Fracture

    IT WAS REPORTED THAT DURING A TOTAL KNEE PROCEDURE, THE HEAD OF THE SCREW BROKE OFF IN THE DRIVER. THERE WAS NO PATIENT IMPACT AND NO ADVERSE EVENTS HAVE BEEN REPORTED AS A RESULT OF THE MALFUNCTION. IT WAS REPORTED THAT ALL AVAILABLE INFORMATION HAS BEEN PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23594094 ↗
  • MalfunctionNov 12, 2025Fracture; Fracture

    IT WAS REPORTED THAT SEVERAL THIN METAL FRAGMENTS WERE FOUND WHEN THE INSERTED SCREW WAS REMOVED WITH A POWER TOOL. ALL AVAILABLE INFORMATION HAS BEEN PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23536070 ↗
  • MalfunctionAug 22, 2025Fracture; Fracture

    IT WAS REPORTED THE HEAD OF SCREW BROKE OFF IN USE. NO PATIENT HARM OR IMPACT WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22874290 ↗
  • MalfunctionFeb 24, 2025Fracture; Fracture

    NO FURTHER EVENT INFORMATION AVAILABLE AT THE TIME OF THIS REPORT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21448405 ↗
  • MalfunctionDec 4, 2024Fracture; Fracture

    IT WAS REPORTED THAT DURING THE PROCEDURE ONE OF THE SCREW HEADS FRACTURED. THE HEAD WAS BURRED OUT AND ALL PIECES WERE RETRIEVED. NO PIECES WERE LEFT IN THE PATIENT. IT TOOK ABOUT TWENTY FIVE MINUTES EXTRA BEFORE PROCEEDING.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20846050 ↗
  • MalfunctionMay 21, 2024Fracture; Fracture

    NO FURTHER EVENT INFORMATION AVAILABLE AT THE TIME OF THIS REPORT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19364165 ↗
  • MalfunctionAug 10, 2023Fracture; Mechanical Jam; Fracture; Mechanical Jam

    NO FURTHER EVENT INFORMATION AVAILABLE AT THE TIME OF THIS REPORT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17498302 ↗
  • MalfunctionMar 19, 2022Fracture; Fracture

    IT WAS REPORTED THAT THE HEAD OF THE SCREW FRACTURED OFF UPON SCREWING IT IN ATTEMPTS TO OBTAIN ADDITIONAL INFORMATION HAVE BEEN MADE; HOWEVER, NO MORE INFORMATION IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13824514 ↗
  • MalfunctionSep 27, 2021Fracture; Fracture

    IT WAS REPORTED THAT WHILE THE SURGEON WAS DRILLING THE HEADED SCREW INTO THE FEMORAL CUTTING BLOCK, THE TIP OF THE SCREW CAME OFF. NO ADVERSE EVENTS WERE REPORTED AS A RESULT OF THIS MALFUNCTION. ATTEMPTS HAVE BEEN MADE, HOWEVER, NO ADDITIONAL INFORMATION IS AVAILABLE AT THIS TIME.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12536003 ↗
  • InjuryJun 3, 2021Fracture; Fracture

    NO FURTHER EVENT INFORMATION AVAILABLE AT THE TIME OF THIS REPORT.

    Patient problems: Foreign Body In Patient

    MAUDE report 11931507 ↗
  • InjuryFeb 4, 2021Fracture; Fracture

    IT IS REPORTED THAT DURING KNEE ARTHROPLASTY WHILE THE SURGEON WAS INSERTING A SCREW INTO THE FEMUR THROUGH THE SPACER BLOCK FEMORAL RESECTOR, THE HEAD OF THE SCREW FRACTURED INSIDE THE INSERTER. THE SURGEON ATTEMPTED TO REMOVE THE SCREW FROM THE DEVICE WITH PLIERS, RESULTING IN A TWENTY-FIVE (25) MINUTE DELAY AND AN ANTERIOR MICRO FRACTURE TO THE PATIENT'S FEMUR. ATTEMPTS HAVE BEEN MADE, HOWEVER, NO ADDITIONAL INFORMATION IS AVAILABLE AT THIS TIME.

    Patient problems: Bone Fracture(s)

    MAUDE report 11278118 ↗
  • InjuryFeb 4, 2021Fracture; Fracture

    IT IS REPORTED THAT DURING KNEE ARTHROPLASTY WHILE THE SURGEON WAS INSERTING A SCREW INTO THE FEMUR THROUGH THE SPACER BLOCK FEMORAL RESECTOR, THE HEAD OF THE SCREW FRACTURED INSIDE THE INSERTER. THE SURGEON ATTEMPTED TO REMOVE THE SCREW FROM THE DEVICE WITH PLIERS, RESULTING IN A TWENTY-FIVE (25) MINUTE DELAY AND AN ANTERIOR MICRO FRACTURE TO THE PATIENT'S FEMUR. ATTEMPTS HAVE BEEN MADE, HOWEVER, NO ADDITIONAL INFORMATION IS AVAILABLE AT THIS TIME.

    Patient problems: Bone Fracture(s)

    MAUDE report 11278117 ↗

Other ORTHOPEDIC MANUAL SURGICAL INSTRUMENT devices

All ORTHOPEDIC MANUAL SURGICAL INSTRUMENT devices →