Skip to content

NA 32-8105-27-01 (REF 32-8105-027-01)Recalls & reported problems

Zimmer, Inc. · Class 2 · Implant

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00889024273993), received between Apr 20, 2020 and Feb 8, 2023 — classified as 2 injuries.

The most frequently reported specific device problem is “Fracture” (1 report).

10 other recalls cover devices cleared under the same FDA submission (K001989, K181307); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Most reported problems

Device problems

  • Fracture1
  • Naturally Worn1
  • Unstable1

Patient problems

  • Failure of Implant2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • InjuryFeb 8, 2023Naturally Worn; Naturally Worn

    IT WAS REPORTED THAT THE PATIENT UNDERWENT REVISION DUE TO A BUSHING EXCHANGE FOR POLYETHYLENE WEAR. NO ADDITIONAL PATIENT CONSEQUENCES WERE REPORTED.

    Patient problems: Failure of Implant

    MAUDE report 16329392 ↗
  • InjuryApr 20, 2020Fracture; Unstable; Fracture; Unstable

    IT WAS FURTHER REPORTED THAT THE PATIENT HAD AN INITIAL LEFT TOTAL ELBOW ARTHROPLASTY PERFORMED APPROXIMATELY FIVE (5) YEARS AGO. SUBSEQUENTLY, THE PATIENT WAS REVISED TWO (2) YEARS LATER DUE TO PAIN AND METALLOSIS FROM POLY WEAR (COVERED UNDER ANOTHER COMPLAINT). THE PATIENT THEN UNDERWENT A SECOND REVISION ON ABOUT THREE (3) MONTHS AGO DUE TO FAILURE OF THE COUPLING MECHANISM LEADING TO SEPARATION OF THE ULNAR AND HUMERAL COMPONENTS. METALLOSIS WAS DEBRIDED FROM THE JOINT, AND A NEW BUSHING WAS PLACED WITHOUT COMPLICATION.

    Patient problems: Failure of Implant

    MAUDE report 9977099 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K001989, K181307. One submission often covers many models, so check the product and lot details in each record.

  • Z-1000-2020 · Dec 12, 2019 · Terminated
    Coonrad/Morrey Total Elbow Humeral Assembly, Extra Small, 4-inch Length. Model No. 32-8105-027-04. UDI (01)00889024274006 (17)241031(10)64481139 The device is a Humeral component;
  • Z-0301-2020 · Oct 10, 2019 · Terminated
    Coorad Morrey Total Elbow, Item Nos. 32810502501 32810502701 32810502900 32810503000 32810600012 32810600013 32855510796 32855510797 32855510799 32855510800 32855510801
  • Z-1239-2018 · Jan 24, 2018 · Terminated
    Coonrad/Morrey Elbow Implant System humeral assembly 4IN XSML, Item Number: 32-8105-027-04 (UDI: (01)0088902427 4006(17)220731(10)63694912)
  • Z-2752-2016 · Aug 16, 2016 · Terminated
    Coonrad/ Morrey Elbow Pin/ Bushing Replacement size Kit Extra Small. This product is sold sterile. The parts are placed into an inner cavity and are sealed using a nitrogen process
  • Z-0538-2013 · Nov 19, 2012 · Terminated
    Coonrad/Morrey New Ulnar Revision Kit for Use with Non-Interchangeable Regular Humerals STERILE Rx, Tivanium QTY/PKG:1 Product Usage: The ulnar and humeral components are connec
  • Z-1058-2009 · Jan 30, 2009 · Terminated
    Zimmer Coonrad/Morrey Total Elbow, Interchangeable Ulnar Assembly, plasma sprayed, extra small, right, long, sterile, Zimmer, Warsaw, IN; REF 32-8105-093-02. The device is utilize
  • Z-1057-2009 · Jan 30, 2009 · Terminated
    Zimmer Coonrad/Morrey Total Elbow, Interchangeable Ulnar Assembly, plasma sprayed, extra small, left, long, sterile, Zimmer, Warsaw, IN; REF 32-8105-093-01. The device is utilized
  • Z-1054-2009 · Jan 30, 2009 · Terminated
    Zimmer Coonrad/Morrey Total Elbow, Interchangeable Humeral Assembly, extra small, 6 inch length, sterile, Zimmer Inc, Warsaw, IN; REF 32-8105-027-06. The device is utilized for a
  • Z-1053-2009 · Jan 30, 2009 · Terminated
    Zimmer Coonrad/Morrey Total Elbow, Interchangeable Humeral Assembly, extra small, 4 inch length, sterile, Zimmer Inc., Warsaw, IN; REF 32-8105-027-04. The device is utilized for a
  • Z-0767-04 · Mar 16, 2004 · Terminated
    Zimmer brand Coonrad/Morrey total elbow interchangeable ulnar assembly, small, left, 3 in. length. tivanium TI-6AL-4V alloy/UHMWPE, sterile, for use with bone cement. Cat. no. 32-

Other PROSTHESIS, ELBOW, CONSTRAINED, CEMENTED devices

All PROSTHESIS, ELBOW, CONSTRAINED, CEMENTED devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page