Skip to content

NCB® 02.02263.018Recalls & reported problems

Zimmer GmbH · NCB PP PROX FEM PLATE, R,18 H, L. 363MM · Class 2

All models of NCB (78)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
9
Deaths reported
0
Injuries reported
9
Recalls naming it
0

Summary

The FDA's MAUDE database contains 9 adverse event reports that list this exact device (DI 00889024292710), received between Dec 18, 2018 and Feb 10, 2026 — classified as 9 injuries.

The most frequently reported specific device problem is “Fracture” (6 reports).

2 other recalls cover devices cleared under the same FDA submission (K100111); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 1 reportsNov 2024: 1 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 1 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Fracture6
  • Adverse Event Without Identified Device or Use Problem2
  • Material Twisted/Bent1

Patient problems

  • Failure of Implant2
  • No Code Available2
  • No Information2
  • Pain2
  • Bone Fracture(s)1
  • Nonunion/Delayed-union Bone Fracture1
  • Periprosthetic fracture1
  • Skin Inflammation/ Irritation1

Recent adverse event reports

The 9 most recent of 9 reports, in the reporter's own words as published by the FDA.

  • InjuryFeb 10, 2026Fracture

    IT WAS REPORTED THE PATIENT UNDERWENT PRIMARY SURGERY, AND SUBSEQUENTLY WAS REVISED DUE TO A NCB PLATE FRACTURE APPROXIMATELY 8 WEEKS POST PRIMARY SURGERY. A NEW PLATE WAS IMPLANTED. ATTEMPTS HAVE BEEN MADE AND NO FURTHER INFORMATION HAS BEEN PROVIDED.

    Patient problems: Nonunion/Delayed-union Bone Fracture; Periprosthetic fracture

    MAUDE report 24301946 ↗
  • InjuryNov 26, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    NO FURTHER EVENT INFORMATION AVAILABLE AT THE TIME OF THIS REPORT.

    Patient problems: Bone Fracture(s)

    MAUDE report 20785659 ↗
  • InjuryOct 23, 2024Fracture; Fracture

    IT WAS REPORTED THAT THE PATIENT UNDERWENT A REVISION PROCEDURE DUE TO A BONE FRACTURE. THE PATIENT FELL AND SUSTAINED A DIAPHYSEAL FEMUR FRACTURE OF HER RIGHT HIP WHICH WAS INITIALLY TREATED WITH THE PLACEMENT OF AN 18-HOLE PROXIMAL NCB PLATE. APPROXIMATELY A MONTH LATER, THE PATIENT EXPERIENCED ACUTE PAIN WHEN TURNING OVER AND FELT IT CRACK RESULTING IN HER NOT BEING ABLE TO SIT AFTER. ON X-RAY IT WAS SEEN THAT THE PLATE HAS COME OFF JUST ABOVE THE FRACTURE. THERE IS ALSO A SUSPICION OF INFECTION AS THE WOUND WAS RED AND WARM. THE PATIENT WAS REVISED APPROXIMATELY 1 WEEK LATER. DUE DILIGENCE HAS BEEN COMPLETED FOR THIS COMPLAINT; TO DATE NO ADDITIONAL INFORMATION HAS BEEN RECEIVED.

    Patient problems: Failure of Implant; Pain; Skin Inflammation/ Irritation; Unspecified Infection

    MAUDE report 20517272 ↗
  • InjuryOct 5, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    NO ADDITIONAL EVENT INFORMATION AT THIS TIME.

    Patient problems: Pain

    MAUDE report 17874638 ↗
  • InjuryOct 14, 2020Fracture; Fracture

    INVESTIGATION RESULTS ARE NOW AVAILABLE.

    Patient problems: No Code Available

    MAUDE report 10676542 ↗
  • InjuryOct 12, 2020Fracture; Fracture

    INVESTIGATION RESULTS ARE NOW AVAILABLE.

    Patient problems: No Code Available

    MAUDE report 10661505 ↗
  • InjuryApr 7, 2020Fracture; Fracture

    INVESTIGATION RESULTS ARE NOW AVAILABLE.

    Patient problems: Failure of Implant

    MAUDE report 9933239 ↗
  • InjuryApr 6, 2020Material Twisted/Bent; Material Twisted/Bent

    INVESTIGATION RESULTS ARE NOW AVAILABLE.

    Patient problems: No Information

    MAUDE report 9928360 ↗
  • InjuryDec 18, 2018Fracture; Fracture

    PLEASE REFER TO REPORT 0009613350 - 2018 - 01245.

    Patient problems: No Information

    MAUDE report 8173582 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K100111. One submission often covers many models, so check the product and lot details in each record.

  • Z-0329-2020 · Oct 10, 2019 · Terminated
    Various temporary devices used during the implantation of the device system: bone fixation plates and screws, drill bits, manual surgical instruments for general use. Item Nos. 020
  • Z-0304-2020 · Oct 10, 2019 · Terminated
    NCB PERIPROSTHETIC, "NCB PP PROX FEM PLATE", Various sizes, Item Nos. 0202263009 0202263012 0202263015 0202263018 0202263021 0202263109 0202263112 0202263115 0202263118 0202263121

Other Screw, fixation, bone devices

All Screw, fixation, bone devices →