Skip to content

PathFinder NXT® 3505-6550Recalls & reported problems

Zimmer Spine, Inc. · Class 2

All models of PathFinder NXT (13)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00889024335714), received between Jul 19, 2021 and Nov 20, 2024 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Detachment of Device or Device Component” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 1 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Detachment of Device or Device Component2

Patient problems

  • No Clinical Signs, Symptoms or Conditions2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionNov 20, 2024Detachment of Device or Device Component; Detachment of Device or Device Component

    IT WAS REPORTED THAT INTRA-OPERATIVELY, A PATHFINDER TORQUE RATCHET HANDLE WAS NOT ABLE TO TIGHTEN CLOSURE TOPS PROPERLY, STRIPPING THE THREADS ON FIVE CLOSURE TOPS AND STRIPPING THE INNER DRIVE HEXAGON ON THREE MORE. A PEDICLE SCREW WAS ALSO REPORTED TO HAVE DISASSEMBLED DURING THE SURGERY. THERE WAS NO PATIENT HARM, HOWEVER THERE WAS A 30 MINUTE PROCEDURAL DELAY AS A RESULT OF THIS. THIS IS REPORT ONE OF TEN FOR THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20736754 ↗
  • MalfunctionJul 19, 2021Detachment of Device or Device Component; Detachment of Device or Device Component

    IT WAS REPORTED THAT DURING SURGERY, A CLOSURE TOP CROSS-THREADED INTO ITS MATING SCREW AND COULD NOT BE ADVANCED OR REMOVED. ADDITIONALLY, THREE SCREWS DISASSEMBLED. THE SCREWS WERE ALL REMOVED AND REPLACED, AND THE CLOSURE TOP WAS EVENTUALLY REMOVED AND REPLACED AFTER A DELAY TO THE PROCEDURE. THE PATIENT'S BONE WAS NOT DAMAGED AND THE PROCEDURE WAS SUCCESSFULLY COMPLETED. THIS IS REPORT FOUR OF FOUR FOR THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12188478 ↗

Other Orthosis, spondylolisthesis spinal fixation devices

All Orthosis, spondylolisthesis spinal fixation devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page