A.T.S.® Personalized Pressure® 60-2200-101-00Recalls & reported problems
Zimmer Orthopaedic Surgical Products, Inc. · Class 1
All models of A.T.S. Personalized Pressure (10)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 6
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 00889024376540), received between Oct 27, 2022 and Aug 28, 2025 — classified as 6 malfunctions.
The most frequently reported specific device problem is “Measurement System Incompatibility” (3 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Measurement System Incompatibility3
- Deflation Problem1
- Noise, Audible1
- Output Problem1
- Use of Incorrect Control/Treatment Settings1
Patient problems
- No Clinical Signs, Symptoms or Conditions4
- Bruise/Contusion1
- Insufficient Information1
Recent adverse event reports
The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.
MalfunctionAug 28, 2025Measurement System Incompatibility; Measurement System Incompatibility
IT WAS REPORTED THAT DURING KIT INSPECTION THE DEVICE DID NOT HOLD PRESSURE. DEVICE WAS SET TO 250MMHG; HOWEVER, THE DEVICE IS READING BETWEEN 230 TO 320MMHG. THERE WAS NO PATIENT INVOLVEMENT. DUE DILIGENCE IS COMPLETE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22916485 ↗MalfunctionAug 12, 2025Measurement System Incompatibility; Measurement System Incompatibility
IT WAS REPORTED THAT DURING AN UNKNOWN TIME, THE UNIT STRUGGLED TO MAINTAIN PRESSURE AND WAS CONSTANTLY MAKING THE INFLATING NOISE. IT WAS SET TO 250MMHG, INFLATE AND THE MACHINE WOULD HOVER BETWEEN 240-249MMHG BUT WITH A CONTINUOUS SOUND OF INFLATING. IT APPEARED TO BE WORKING HARD TO MAINTAIN THE PRESSURE. THERE WAS NO INFORMATION AVAILABLE REGARDING THE PATIENT IMPACT. DUE DILIGENCE IS COMPLETE; NO FURTHER INFORMATION IS AVAILABLE.
Patient problems: Insufficient Information
MAUDE report 22782321 ↗MalfunctionApr 4, 2024Measurement System Incompatibility; Measurement System Incompatibility
IT WAS REPORTED THAT DURING SURGERY, THE PRESSURE ON THE DEVICE WAS NOT MAINTAINING THE PRESSURE IT WAS SET TO AT THE BEGINNING OF THE CASE. THERE WAS NO SURGICAL DELAY. DOCTORS WERE SEEING PATIENTS COMING BACK WITH SEVERE BRUISING FROM THE TOURNIQUET. THE PA STARTED NOTICING THAT THIS PROBLEM WAS ONLY OCCURRING WITH THIS SPECIFIC MACHINE. DUE DILIGENCE IS IN PROGRESS, NO FURTHER INFORMATION IS AVAILABLE.
Patient problems: Bruise/Contusion
MAUDE report 19045253 ↗MalfunctionDec 4, 2023Use of Incorrect Control/Treatment Settings; Noise, Audible; Use of Incorrect Control/Treatment Settings; Noise, Audible
IT WAS REPORTED THAT DURING SURGERY THE DEVICE WAS MAKING A PECULIAR NOISE AND THE PRESSURE SENSOR IS CONSTANTLY ACTIVE CALCULATING THE PRESSURE OCCURRING FROM BOTH PORTS. THERE WAS NO PATIENT IMPACT OR DELAY. DUE DILIGENCE IS IN PROCESS AND THERE IS NO ADDITIONAL INFORMATION AVAILABLE. THERE WAS NO ADVERSE EVENT ASSOCIATED WITH THE MALFUNCTION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18255448 ↗MalfunctionDec 19, 2022Deflation Problem; Deflation Problem
IT WAS REPORTED DURING INSPECTION/TESTING, THE DEVICE SPONTANEOUSLY DEFLATED AND REQUIRED UNKNOWN REPAIRS. NO ADVERSE EVENT WAS REPORTED AS IT RELATES TO THE EVENT. DUE DILIGENCE IS COMPLETE AND NO ADDITIONAL INFORMATION IS AVAILABLE. IF ADDITIONAL INFORMATION IS PROVIDED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16010069 ↗MalfunctionOct 27, 2022Output Problem; Output Problem
IT WAS REPORTED THAT DURING PM, THE PRESSURE GIVEN OUT IS 10-13 MMHG MORE THAN WHAT IT WAS SET TO. THERE WAS NO PATIENT INVOLVEMENT. NO ADVERSE EVENTS WERE REPORTED AS A RESULT OF THIS MALFUNCTION. DUE DILIGENCE IS COMPLETE AND NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15682588 ↗
Other Tourniquet, Pneumatic devices
- N/AStryker Sustainability Solutions, Inc. · TOURNIQUET, PNEUMATIC1 recall82 reports
- Color Cuff 5921018135Stryker Corporation · TOURNIQUET, PNEUMATIC2 recalls31 reports
- N/AStryker Sustainability Solutions, Inc. · TOURNIQUET, PNEUMATIC1 recall72 reports
- Color Cuff 5921018235Stryker Corporation · TOURNIQUET, PNEUMATIC2 recalls20 reports
- N/AStryker Sustainability Solutions, Inc. · TOURNIQUET, PNEUMATIC1 recall38 reports
- SmartPump 5920011000Stryker Corporation · TOURNIQUET, PNEUMATIC72 reports
- A.T.S.® Personalized Pressure® 60320010100Zimmer Orthopaedic Surgical Products, Inc. · TOURNIQUET, PNEUMATIC1 recall20 reports
- Color Cuff 5921-034-135NSStryker Corporation · TOURNIQUET, PNEUMATIC1 recall20 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page