Skip to content

A.T.S.® Personalized Pressure® 60-2200-301-06Recalls & reported problems

Zimmer Orthopaedic Surgical Products, Inc. · Class 1

All models of A.T.S. Personalized Pressure (10)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00889024419889), received between Apr 19, 2022 and May 23, 2025 — classified as 3 malfunctions.

The most frequently reported specific device problem is “Measurement System Incompatibility” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 2 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Measurement System Incompatibility2
  • Deflation Problem1
  • Inflation Problem1

Patient problems

  • Insufficient Information2
  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMay 23, 2025Measurement System Incompatibility; Measurement System Incompatibility

    IT WAS REPORTED THAT DURING AN UNKNOWN TIME THE PRESSURE DOES NOT REACH THE DESIRED VALUE AND IS UNSTABLE. WHEN THE PRESSURE IS SET TO 300, THE VALUES DISPLAYED OSCILLATE WITHOUT EVER REACHING THE REQUIRED LEVEL. THIS MALFUNCTION PERSISTS REGARDLESS OF THE CUFF OR CABLE USED. IT IS UNKNOWN IF THERE WAS PATIENT IMPACT OR DELAY. DUE DILIGENCE IS COMPLETE AND THERE IS NO ADDITIONAL INFORMATION AVAILABLE.

    Patient problems: Insufficient Information

    MAUDE report 22078386 ↗
  • MalfunctionMay 23, 2025Measurement System Incompatibility; Measurement System Incompatibility

    IT WAS REPORTED DURING AN UNKNOWN EVENT TIMING THAT THE PRESSURE DOES NOT REACH DESIRED VALUE AND IS UNSTABLE. WHEN THE PRESSURE WAS SET TO 300, THE VALUES DISPLAYED OSCILLATE WITHOUT EVER REACHING THE REQUIRED LEVEL. DUE DILIGENCE IS COMPLETE AND THERE IS NO ADDITIONAL INFORMATION AVAILABLE.

    Patient problems: Insufficient Information

    MAUDE report 22078081 ↗
  • MalfunctionApr 19, 2022Deflation Problem; Inflation Problem; Deflation Problem; Inflation Problem

    IT WAS REPORTED THAT BEFORE THE PROCEDURE THE DEVICE WAS INFLATING AND DEFLATING FOR NO REASON. NO HARM OR DELAY WERE NOTED AND ANOTHER DEVICE WAS NOT NOTED TO BE USED. NO OTHER ADVERSE EVENTS WERE REPORTED ALONG WITH THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14140582 ↗

Other TOURNIQUET, PNEUMATIC devices

All TOURNIQUET, PNEUMATIC devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page