Skip to content

Avenir® Müller 01.06010.201Recalls & reported problems

Zimmer GmbH · Class 2 · Implant

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00889024590731), received between Nov 5, 2024 and Nov 5, 2024 — classified as 1 injury.

The most frequently reported specific device problem is “Loosening of Implant Not Related to Bone-Ingrowth” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 1 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Loosening of Implant Not Related to Bone-Ingrowth1
  • Separation Problem1

Patient problems

  • Failure of Implant1
  • Joint Dislocation1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • InjuryNov 5, 2024Loosening of Implant Not Related to Bone-Ingrowth; Separation Problem; Loosening of Implant Not Related to Bone-Ingrowth; Separation Problem

    IT WAS REPORTED PATIENT UNDERWENT HIP REVISION APPROXIMATELY 21 DAYS POST IMPLANTATION DUE TO A DISLOCATION CAUSED BY A FALL IN THE DOMESTIC ENVIRONMENT. DURING THE INTRAOPERATIVE ATTEMPT TO DETACH THE HEAD FROM THE STEM, THE HEAD AND STEM CAME OUT OF THE MEDULLARY CAVITY OF THE FEMUR AFTER THE THIRD STROKE. AS A RESULT, THE MEDULLARY CANAL HAD TO BE PARTIALLY FREED FROM THE CEMENT IN ORDER TO BE ABLE TO PLACE A NEW STEM INCLUDING A NEW THIN CEMENT MANTLE AND A HEAD WITH MORE OFFSET HAD TO BE IMPLANTED. DELAY 60 MINUTES FOR COMPLEX PARTIAL REMOVAL OF THE CEMENT IN THE FEMUR TO MAKE ROOM FOR RE-IMPLANTATION AND PARTIAL NEW CEMENT MANTLE, TRIAL REDUCTION, RE-IMPLANTATION INCLUDING HARDENING OF THE CEMENT, REDUCTION. ATTEMPTS HAVE BEEN MADE AND ADDITIONAL INFORMATION ON THE REPORTED EVENT IS UNAVAILABLE AT THIS TIME.

    Patient problems: Failure of Implant; Joint Dislocation

    MAUDE report 20612029 ↗

Other Prosthesis, hip, hemi-, femoral, metal/polymer, cemented or uncemented devices

All Prosthesis, hip, hemi-, femoral, metal/polymer, cemented or uncemented devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page