A.T.S.® Personalized Pressure® 60-5000-101-00Recalls & reported problems
Zimmer Orthopaedic Surgical Products, Inc. · Class 1
All models of A.T.S. Personalized Pressure (10)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 5
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 00889024626454), received between Aug 15, 2022 and Oct 8, 2024 — classified as 5 malfunctions.
The most frequently reported specific device problem is “Measurement System Incompatibility” (3 reports).
1 FDA recall record name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Recalls naming this device
- Dec 24, 2025Class IIOpen, Classified
Device user interface (screen) may freeze or become unresponsive to touch when operating in certain non-English language settings. Issue may cause delay or extension while alternate device is located.
Z-1201-2026 · Root cause: Under Investigation by firm · Recall details · FDA record ↗
Most reported problems
Device problems
- Measurement System Incompatibility3
- Deflation Problem2
- Detachment of Device or Device Component1
- Display or Visual Feedback Problem1
- Gas/Air Leak1
- Pressure Problem1
- Stretched1
Patient problems
- No Clinical Signs, Symptoms or Conditions5
Recent adverse event reports
The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.
MalfunctionOct 8, 2024Measurement System Incompatibility; Pressure Problem; Measurement System Incompatibility; Pressure Problem
THERE IS NO ADDITIONAL INFORMATION REGARDING THE EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20396384 ↗MalfunctionMay 28, 2024Measurement System Incompatibility; Measurement System Incompatibility
IT WAS REPORTED THAT DURING SURGERY THAT THE UNIT WAS NOT MAINTAINING PRESSURE. THERE WAS NO HARM OR INJURY AND NO REPORTED DELAY. DUE DILIGENCE IS IN PROGRESS. NO ADDITIONAL INFORMATION IS AVAILABLE AT THIS TIME.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19406226 ↗MalfunctionMay 28, 2024Measurement System Incompatibility; Measurement System Incompatibility
IT WAS REPORTED THAT DURING SURGERY THAT THE UNIT WAS NOT MAINTAINING PRESSURE. THERE WAS NO HARM OR INJURY TO PATIENT AND NO REPORTED DELAY. DUE DILIGENCE IS IN PROGRESS. NO ADDITIONAL INFORMATION IS AVAILABLE AT THIS TIME.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19405975 ↗MalfunctionSep 30, 2022Deflation Problem; Display or Visual Feedback Problem; Deflation Problem; Display or Visual Feedback Problem
IT WAS REPORTED THAT DURING SURGERY THE UNIT EXPERIENCED SPONTANEOUS DEFLATION OF TOURNIQUET CUFF AND DISPLAY KEPT FLUCTUATING. NO HARM OR DELAY WERE REPORTED AS IT RELATES TO THIS EVENT. NO ADVERSE EVENTS WERE REPORTED AS IT RELATES TO THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15509891 ↗MalfunctionAug 15, 2022Deflation Problem; Stretched; Detachment of Device or Device Component; Gas/Air Leak; Deflation Problem; Stretched; Detachment of Device or Device Component; Gas/Air Leak
IT WAS FURTHER REPORTED THAT THE DEVICE WAS A DEMO UNIT THAT WAS IN NEED OF REPAIR AS THE VALVE ASSEMBLY, MAIN CUFF, AND THE DEFLATE VALVE FOR THE SECONDARY CUFF WERE LEAKING AIR. THE MAIN TUBING WAS FALLING APART AND THE TUBING WAS LEAKING AIR AS WELL DUE TO BEING STRETCHED OUT. THE BARB FITTING FOR THE MAIN CUFF WAS ALSO ADJUSTED DUE TO AN AIR LEAK. THERE WAS ALSO SPONTANEOUS DEFLATION. NO ADVERSE EVENTS WERE REPORTED AS A RESULT OF THIS MALFUNCTION. ATTEMPTS HAVE BEEN MADE AND NO FURTHER INFORMATION HAS BEEN PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15223621 ↗
Other TOURNIQUET, PNEUMATIC devices
- N/AStryker Sustainability Solutions, Inc. · TOURNIQUET, PNEUMATIC1 recall82 reports
- Color Cuff 5921018135Stryker Corporation · TOURNIQUET, PNEUMATIC2 recalls31 reports
- N/AStryker Sustainability Solutions, Inc. · TOURNIQUET, PNEUMATIC1 recall72 reports
- Color Cuff 5921018235Stryker Corporation · TOURNIQUET, PNEUMATIC2 recalls20 reports
- N/AStryker Sustainability Solutions, Inc. · TOURNIQUET, PNEUMATIC1 recall38 reports
- SmartPump 5920011000Stryker Corporation · TOURNIQUET, PNEUMATIC72 reports
- A.T.S.® Personalized Pressure® 60320010100Zimmer Orthopaedic Surgical Products, Inc. · TOURNIQUET, PNEUMATIC1 recall20 reports
- Color Cuff 5921-034-135NSStryker Corporation · TOURNIQUET, PNEUMATIC1 recall20 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page