Skip to content

CREO 6120.2010Recalls & reported problems

Globus Medical, Inc. · Geared Reducer, Overhead, Threaded · Class 1

All models of CREO (100)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
1
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00889095173376), received between Jan 6, 2026 and Jan 6, 2026 — classified as 1 death.

The most frequently reported specific device problem is “Device-Device Incompatibility” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 1 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Device-Device Incompatibility1

Patient problems

  • Multiple Organ Dysfunction Syndrome1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • DeathJan 6, 2026Device-Device Incompatibility

    T WAS REPORTED THAT THESE PARTS WOULD ENGAGE WITH TULIP, BUT THE SET SCREW WOULD NOT THREAD INTO THE TULIP. WHEN THE REDUCERS WERE ON THE TULIP & FULLY REDUCED ROD. THIS ADDED AN EXTRA 45 MIN TO SURGERY AS THEY TROUBLESHOOTED THE ISSUES, ULTIMATELY RESOLVED BY SWITCHING TO NEW INSTRUMENTATION. THE PATIENT PASSED AWAY AT THE HOSPITAL 5 DAYS LATER, THE SURGEON BELIEVES THE PATIENT HAD PROPOSAL INFUSION SYNDROME.

    Patient problems: Multiple Organ Dysfunction Syndrome

    MAUDE report 23975742 ↗

Other ORTHOPEDIC MANUAL SURGICAL INSTRUMENT devices

All ORTHOPEDIC MANUAL SURGICAL INSTRUMENT devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page