Skip to content

APEX 3D Total Ankle Replacement System P10-251-TLR2-SRecalls & reported problems

Paragon 28, Inc. · APEX 3D™ Talus, Flat, Right, Size 2, TPS · Class 2 · Implant

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
3
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00889795060280), received between Oct 15, 2025 and Mar 26, 2026 — classified as 3 injuries.

The most frequently reported specific device problem is “Implant subsidence” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 1 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 2 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem2
  • Implant subsidence1
  • Insufficient Device Problem Information1
  • Loosening of Implant Not Related to Bone-Ingrowth1
  • Patient Device Interaction Problem1

Patient problems

  • Pain2
  • Aseptic loosening1
  • Bone Fracture(s)1
  • Cellulitis1
  • Erythema1
  • Exostosis (bone)1
  • Fluid Discharge1
  • Joint Impingement1

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • InjuryMar 26, 2026Adverse Event Without Identified Device or Use Problem; Loosening of Implant Not Related to Bone-Ingrowth; Implant subsidence

    IT WAS FURTHER REPORTED THAT APPROXIMATELY 17 MONTHS POST IMPLANTATION OF A RIGHT TOTAL ANKLE ARTHROPLASTY, THE PATIENT PRESENTED WITH PAIN AND SWELLING AND UNDERWENT A REVISION WHERE SIGNIFICANT SCAR TISSUE WAS RESECTED FROM THE ANTERIOR JOINT, SIGNIFICANT OSTEOPHYTES WERE DEBRIDED FROM THE MEDIAL AND LATERAL GUTTERS, AND THE POLY WAS EXCHANGED. A COMPETITOR SCREW WAS PLACED TO HELP SUPPORT THE MEDIAL MALLEOLUS. DURING THE REVISION, IT WAS NOTED THAT THE TIBIAL AND TALAR COMPONENTS WERE WELL-FIXED. UPON EXAMINATION OF THE INTRAOPERATIVE X-RAYS, IT WAS NOTED THAT THE STRESS REACTION OF THE MEDIAL TIBIA WAS SUBSTANTIAL COMPARED TO PREOPERATIVE X-RAYS, AND IT WAS BELIEVED THAT THE MEDIAL PORTION OF THE TALUS AND THE MEDIAL PORTION OF THE TIBIA HAD SIGNIFICANT IMPINGEMENT AND THIS WAS CAUSING STRESS CHANGE OF THE DISTAL TIBIA. ATTEMPTS HAVE BEEN MADE AND ALL AVAILABLE INFORMATION HAS BEEN PROVIDED.

    Patient problems: Aseptic loosening; Exostosis (bone); Joint Impingement; Osteopenia/ Osteoporosis; Pain; Scar Tissue; Swelling/ Edema

    MAUDE report 24705359 ↗
  • InjuryMar 25, 2026Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT APPROXIMATELY 7 MONTHS POST IMPLANTATION OF A RIGHT ANKLE PROSTHESIS SYSTEM, THE PATIENT PRESENTED WITH A NONDISPLACED POSTERIOR TALUS FRACTURE. PATIENT WAS TREATED WITH CAM BOOT, ELEVATION, ICE AND ORAL ANALGESICS. ATTEMPTS HAVE BEEN MADE AND ALL AVAILABLE INFORMATION HAS BEEN PROVIDED.

    Patient problems: Periprosthetic fracture

    MAUDE report 24688733 ↗
  • InjuryOct 15, 2025Insufficient Device Problem Information; Patient Device Interaction Problem; Insufficient Device Problem Information; Patient Device Interaction Problem

    IT WAS FURTHER REPORTED THAT APPROXIMATELY 5 MONTHS POST IMPLANTATION OF A RIGHT TOTAL ANKLE ARTHROPLASTY, THE PATIENT DEVELOPED WOUND DEHISCENCE, ERYTHEMA, AND DRAINAGE, AND WAS ADMITTED FOR OSTEOMYELITIS WITH SECONDARY CELLULITIS, EXPOSED TENDON WITH ODOR, AND PAIN. PATIENT UNDERWENT AN I&D TO BONE AND NECROTIC TISSUE, AND COMPLETED A TENDON TRANSFER AND WOUND CLOSURE. INTRA-OP CULTURES WERE POSITIVE FOR PREVOTELLA AND THE POLY WAS EXCHANGED WITHOUT COMPLICATION. ATTEMPTS HAVE BEEN MADE AND ALL AVAILABLE INFORMATION HAS BEEN PROVIDED.

    Patient problems: Bone Fracture(s); Cellulitis; Erythema; Fluid Discharge; Medical device site infection; Necrosis; Osteomyelitis; Pain; Skin Infection; Wound Dehiscence; Wound Odor

    MAUDE report 23300702 ↗

Other PROSTHESIS, ANKLE, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER devices

All PROSTHESIS, ANKLE, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER devices →