The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00889797059152), received between Apr 17, 2026 and Apr 17, 2026 — classified as 1 malfunction.

The most frequently reported specific device problem is “Component Missing” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 1 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Component Missing1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionApr 17, 2026Component Missing

    AS REPORTED: "THIS CASE WAS MISSING THE TALAR PEG DRILL (IB200020) THIS CAUSED SOME ISSUES AS WE NEEDED TO SOURCE ANOTHER 4MM DRILL THAT WE COULD USE TO DRILL THE PEG HOLES. THE DRILL WE USED CAME FROM ANOTHER SET AND DIDN¿T QUITE FIT CAUSING HEAT BUILD UP ON INSPECTION AFTER THE CASE I NOTICED HEAT STRESS MARKS ON THE ADAPTIS TALAR TRIAL, AND THE DRILL BIT HAD BEEN RENDERED BLUNT. THERE WAS NO METAL DEBRIS LEFT IN THE PATIENT. THIS ISSUE CAUSED A SMALL DELAY BUT NEEDED EXTRA IMPACTION OF THE TALUS COMPONENT AS THE DRILL PROBABLY DIDN¿T DO AS GOOD A JOB AS THE REAL DEVICE. THIS PATIENT HAD A 3 TIBIA AND A 3 TALUS. IF WE HAD OPENED THE PREVIOUS SET, WE WOULD HAVE HAD SOME ISSUES WITH THE MISSING POLY IT WAS MORE LUCK THAN JUDGEMENT. I NORMALLY GET A NO STOCK FOR THESE ITEMS BUT ON THIS OCCASION, I HAD NO EMAIL WARNING OF THE STATE OF THE SETS. THE SETS ARE DUE TO BE PICKED UP TODAY CAN THEY BE QUARANTINED AND RECTIFIED BEFORE REDEPLOYMENT."

    Patient problems: No Clinical Signs, Symptoms or Conditions