INFINITY™ EVERLAST™ 33684408Recalls & reported problems
Wright Medical Technology, Inc. · Class 2 · Implant
All models of INFINITY EVERLAST (20)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 5
- Deaths reported
- 0
- Injuries reported
- 5
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 00889797074605), received between Sep 26, 2024 and Jun 15, 2026 — classified as 5 injuries.
The most frequently reported specific device problem is “Insufficient Device Problem Information” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem4
- Insufficient Device Problem Information1
- Loss of Osseointegration1
- Osseointegration Problem1
Patient problems
- Inadequate Osseointegration2
- Insufficient Information2
- Swelling/ Edema2
- Discomfort1
- Implant Pain1
- Numbness1
- Pain1
Recent adverse event reports
The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.
InjuryJun 15, 2026Loss of Osseointegration; Adverse Event Without Identified Device or Use Problem
THE DIGITAL STRYKER PROPHECY TEAM RECEIVED A CT SCAN INDICATING THAT THE PATIENT MAY REQUIRE A REVISION SURGERY. IT WAS REPORTED: THE IMPLANTS WERE INITIALLY SIZED AND PLACED CORRECTLY, SECURED CORRECTLY WITH CEMENT BUT DID LOOSEN SLOWLY OVER THE TIMEFRAME AND BECAME MUCH MORE SYMPTOMATIC RECENTLY. THE EVENT APPEARS TO BE A STRAIGHTFORWARD CASE OF MECHANICAL LOOSENING AND SUBSEQUENT SUBSIDENCE".
Patient problems: Inadequate Osseointegration; Insufficient Information
MAUDE report 25512609 ↗InjuryDec 10, 2025Insufficient Device Problem Information; Insufficient Device Problem Information
THE DIGITAL STRYKER PROPHECY TEAM RECEIVED A CT SCAN INDICATING THAT THE PATIENT MAY REQUIRE A REVISION SURGERY FOR UNKNOWN REASONS AT THE TIME OF THIS REPORT.
Patient problems: Insufficient Information
MAUDE report 23770043 ↗InjurySep 25, 2025Adverse Event Without Identified Device or Use Problem; Osseointegration Problem; Adverse Event Without Identified Device or Use Problem; Osseointegration Problem
THE DIGITAL STRYKER PROPHECY TEAM RECEIVED A CT SCAN INDICATING THAT REVISION SURGERY MAY BE REQUIRED. A TOTAL ANKLE REPLACEMENT WITH A WRIGHT MEDICAL/STRYKER DEVICE WAS PREVIOUSLY PERFORMED FOR ARTHRITIS AND ANKLE PAIN. REPORTS INCLUDE SWELLING AND MEDIAL ANKLE PAIN, WITH IMAGING SUGGESTING A POSSIBLE MEDIAL MALLEOLUS STRESS FRACTURE AND POTENTIAL TIBIAL TRAY LOOSENING. NO REVISION HAS BEEN PERFORMED TO DATE. BONE QUALITY WAS DESCRIBED AS GOOD, AND COMPLIANCE WITH MOBILIZATION INSTRUCTIONS WAS NOTED, THOUGH A PHYSICALLY ACTIVE LIFESTYLE MAY HAVE CONTRIBUTED TO THE REPORTED SYMPTOMS.
Patient problems: Inadequate Osseointegration; Pain; Swelling/ Edema
MAUDE report 23142671 ↗InjuryOct 11, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
AS REPORTED: "SUBJECT: (B)(6) INFINITY¿ WITH ADAPTIS¿ TOTAL ANKLE REPLACEMENT FOLLOW-UP (ITAR2). (B)(6) 2022: DORSAL FOOT NEURITIS. PERSISTENT NERVE PAIN. NERVE TINGLING & NUMBNESS-DORSAL FOOT. (B)(6) 2023: IMPROVING. HEALING AND PROGRESSING WELL. TREATMENT UPDATED TO INCLUDE ORTHOTICS. (B)(6) 2023: ORTHOTICS, ULTRASOUND GUIDED INJECTION. (B)(6) 2023: IMPROVING. SIGNIFICANT RELIEF AFTER ULTRASOUND INJECTION."
Patient problems: Implant Pain; Numbness
MAUDE report 20426081 ↗InjurySep 26, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
AS REPORTED: "SUBJECT: 700-001 INFINITY¿ WITH ADAPTIS¿ TECHNOLOGY TOTAL ANKLE REPLACEMENT FOLLOW-UP (ITAR2). PATIENT PRESENTED TO THE CLINIC FOR A POSTOP FOLLOW UP APPOINTMENT ON (B)(6) 2022 (3 MONTHS S/P LEFT TAR PROCEDURE), CONCERNED ABOUT CONTINUED SIGNIFICANT SWELLING AND DISCOMFORT. THE SURGEON ORDERED AN ULTRASOUND TO RULE OUT A LEFT LOWER EXTREMITY DEEP VEIN THROMBOSIS (DVT) ON (B)(6) 2022. ULTRASOUND REPORT FOUND NO EVIDENCE OF A DVT. PATIENT IS SCHEDULED FOR ANOTHER POSTOP VISIT ON (B)(6) 2022 (6 MONTHS S/P LEFT TAR PROCEDURE). UPDATE 08 DEC 2022: PATIENT REPORTED SWELLING HAS IMPROVED."
Patient problems: Discomfort; Swelling/ Edema
MAUDE report 20307751 ↗
Other PROSTHESIS, ANKLE, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER devices
- Vantage 350-22-03Exactech, Inc. · Prosthesis, Ankle, Semi-Constrained, Cemented, Metal/Polymer2 recalls2 reports
- Vantage 350-21-04Exactech, Inc. · Prosthesis, Ankle, Semi-Constrained, Cemented, Metal/Polymer2 recalls0 reports
- INFINITY 33630023Wright Medical Technology, Inc. · PROSTHESIS, ANKLE, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER60 reports
- INFINITY 33600020Wright Medical Technology, Inc. · PROSTHESIS, ANKLE, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER1 recall9 reports
- Vantage 350-22-02Exactech, Inc. · Prosthesis, Ankle, Semi-Constrained, Cemented, Metal/Polymer1 recall5 reports
- INFINITY 33650003Wright Medical Technology, Inc. · PROSTHESIS, ANKLE, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER55 reports
- INFINITY 33630022Wright Medical Technology, Inc. · PROSTHESIS, ANKLE, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER54 reports
- INFINITY 33600030Wright Medical Technology, Inc. · PROSTHESIS, ANKLE, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER1 recall4 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page