Skip to content

INFINITY™ 33680022Recalls & reported problems

Wright Medical Technology, Inc. · Class 2 · Implant

Adverse event reports
20
Deaths reported
0
Injuries reported
20
Recalls naming it
0

Summary

The FDA's MAUDE database contains 20 adverse event reports that list this exact device (DI 00889797101523), received between Nov 1, 2021 and Jan 26, 2026 — classified as 20 injuries.

The most frequently reported specific device problem is “Osseointegration Problem” (6 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 1 reportsDec 2023: 0 reportsJan 2024: 1 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 3 reportsNov 2024: 0 reportsDec 2024: 2 reportsJan 2025: 1 reportsFeb 2025: 1 reportsMar 2025: 1 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 1 reportsAug 2025: 0 reportsSep 2025: 4 reportsOct 2025: 0 reportsNov 2025: 1 reportsDec 2025: 0 reportsJan 2026: 1 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520264
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem14
  • Osseointegration Problem6
  • Migration3
  • Appropriate Device Problem Term/Code Not Available1
  • Inadequacy of Device Shape and/or Size1
  • Insufficient Device Problem Information1

Patient problems

  • Inadequate Osseointegration6
  • Pain6
  • Implant Pain5
  • Bone Fracture(s)3
  • Joint Impingement3
  • Swelling/ Edema2
  • Wound Dehiscence2
  • Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available1

Recent adverse event reports

The 20 most recent of 20 reports, in the reporter's own words as published by the FDA.

  • InjuryJan 26, 2026Adverse Event Without Identified Device or Use Problem

    THE DIGITAL STRYKER PROPHECY TEAM RECEIVED A CT SCAN INDICATING THAT THE PATIENT MAY REQUIRE A REVISION SURGERY DUE TO PERSISTENT PAIN.

    Patient problems: Pain

    MAUDE report 24184410 ↗
  • InjuryNov 12, 2025Adverse Event Without Identified Device or Use Problem; Osseointegration Problem; Adverse Event Without Identified Device or Use Problem; Osseointegration Problem

    SUBJECT: (B)(6). INFINITY¿ WITH ADAPTIS¿ TECHNOLOGY TOTAL ANKLE REPLACEMENT FOLLOW-UP (ITAR2). (B)(6) 2022: PATIENT REPORTS PAIN IN LEFT ANKLE. PATIENT HAS SOME TENDERNESS SUB FIBULAR IMPINGEMENT AREA. CT PROPHECY PROTOCOL DONE TO ENSURE THAT IMPLANT IS NOT LOOSE. (B)(6) 2023: SURGICAL INTERVENTION INVOLVING EXPLANT OF DEVICE DUE TO ASEPTIC LOOSENING. PATIENT DISCHARGED ON (B)(6) 2023.

    Patient problems: Inadequate Osseointegration; Pain

    MAUDE report 23536171 ↗
  • InjurySep 9, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    SUBJECT: (B)(6) STUDY: STRYKER ITAR-2 - INFINITY WITH ADAPTIS. SUBJECT HAS PERSISTENT PAIN IN HIS LEFT ANKLE.

    Patient problems: Pain

    MAUDE report 23001788 ↗
  • InjurySep 4, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    STUDY: STRYKER ITAR-2. SUBJECT: (B)(6). (B)(6) 2024: PI REVIEWED HER SPECT/CT SCAN WITH HER TODAY, AND PI FEELS THAT HER IMAGES ARE MOST CONSISTENT WITH LATERAL GUTTER IMPINGEMENT. THIS IS ALSO CONCORDANT WITH HER EXAM AND FOCAL PAIN IN THE LATERAL GUTTER. PATIENT ELECTED TO PROCEED WITH A GUTTER DEBRIDEMENT. TREATMENT FOR THIS CONDITION WOULD CONSIST OF A LATERAL GUTTER DEBRIDEMENT. (B)(6) 2024: PATIENT HAD SURGERY ON THIS DAY FOR A RIGHT ANKLE ARTHROTOMY AND BONY DEBRIDEMENT OF LATERAL GUTTER.

    Patient problems: Joint Impingement; Pain

    MAUDE report 22963654 ↗
  • InjurySep 4, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    SUBJECT: (B)(6). STUDY: (B)(4). SUBJECT HAS INTERMITTENT SWELLING STILL PRESENT.

    Patient problems: Swelling/ Edema

    MAUDE report 22961240 ↗
  • InjurySep 4, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    SUBJECT: (B)(6). STUDY: STRYKER ITAR-2 - INFINITY WITH ADAPTIS. SUBJECT HAS SOME NEW NUMBNESS ALONG THE PLANTAR DISTAL TOES.

    Patient problems: Numbness

    MAUDE report 22961232 ↗
  • InjuryJul 23, 2025Osseointegration Problem; Osseointegration Problem

    THE DIGITAL STRYKER PROPHECY TEAM RECEIVED A CT SCAN INDICATING THAT THE PATIENT MAY REQUIRE A REVISION SURGERY DUE TO GUTTER IMPINGEMENT AND POSSIBLE LOOSENING OF PROSTHESES.

    Patient problems: Inadequate Osseointegration; Joint Impingement; Pain

    MAUDE report 22580193 ↗
  • InjuryMar 18, 2025Inadequacy of Device Shape and/or Size; Adverse Event Without Identified Device or Use Problem; Inadequacy of Device Shape and/or Size; Adverse Event Without Identified Device or Use Problem

    THE DIGITAL STRYKER PROPHECY TEAM RECEIVED A CT SCAN INDICATING THAT THE PATIENT MAY REQUIRE A REVISION SURGERY DUE TO PAIN DUE TO IMPLANT SIZE.

    Patient problems: Implant Pain; Pain

    MAUDE report 21636843 ↗
  • InjuryFeb 26, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    AS REPORTED: "SUBJECT: (B)(6) - INFINITY¿ WITH ADAPTIS¿ TOTAL ANKLE REPLACEMENT FOLLOW-UP (ITAR2) (B)(6) 2021: WORSENING REDNESS AROUND INCISION. DELAYED WOUND HEALING. NO EVIDENCE OF DEEP INFECTION, HOLD MOTION, CONTINUE SOAP AND WATER WASH, BACTRIM ANTIBIOTIC PRESCRIBED. (B)(6) 2021: WORSENING. SMALL AREA OF WOUND DEHISCENCE WITH MILD EXUDATE MILD ERYTHEMA. NO PURULENT DRAINAGE, NO SIGNS OF DEEP INFECTION. CONTINUE WET-TO-DRY DRESSINGS OVER SMALL AREA OF WOUND DEHISCENCE. (B)(6) 2021: NO CHANGE. WOUND DEBRIDED IN OFFICE. HOLD MOTION FOR ANOTHER WEEK. (B)(6) 2021: IMPROVING. WOUND APPEARS LESS IRRITATED, CONTINUE WET-TO-DRY DRESSINGS, REFERRAL TO WOUND CARE. (B)(6) 2021: IMPROVING. GRANULATION TISSUE BED LOOKS EXCEPTIONAL. WOUND APPEARS TO BE RESPONDING EXCELLENT. BACTRIM TO BE CAUTIOUS. (B)(6) 2021: IMPROVING. INCISION HEALING. MILD WOUND DEHISCENCE. MUCH IMPROVED. CONTINUE WITH WOUND CARE. (B)(6) 2021: IMPROVING. INCISION HEALING WELL. MUCH IMPROVED. (B)(6) 2021: RESOLVED. INCISION HEALED. NO DRAINAGE. NO SIGNS OF INFECTION."

    Patient problems: Erythema; Wound Dehiscence

    MAUDE report 21468382 ↗
  • InjuryJan 3, 2025Osseointegration Problem; Migration; Osseointegration Problem; Migration

    THE DIGITAL STRYKER PROPHECY TEAM RECEIVED A CT SCAN INDICATING THAT THE PATIENT MAY REQUIRE A REVISION SURGERY OF BOTH THE TIBIAL AND TALAR COMPONENTS DUE TO FAILURE OF TIBIAL INGROWTH WITH SECONDARY STRESS FRACTURE MEDIAL MALLEOLUS.

    Patient problems: Bone Fracture(s); Cyst(s); Inadequate Osseointegration

    MAUDE report 21071406 ↗
  • InjuryDec 10, 2024Osseointegration Problem; Osseointegration Problem

    THE DIGITAL STRYKER PROPHECY TEAM RECEIVED A CT SCAN INDICATING THAT THE PATIENT MAY REQUIRE A REVISION SURGERY OF TALAR COMPONENT FOR REASONS OF LOOSE COMPONENT. PLAN TO REVISE TO INBONE TALUS.

    Patient problems: Inadequate Osseointegration

    MAUDE report 20891207 ↗
  • InjuryDec 5, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    AS REPORTED: "SUBJECT: (B)(6), STUDY: STRYKER ITAR-2 - INFINITY WITH ADAPTIS. PERSISTENT NERVE PAIN. SUBJECT CONTINUES TO HAVE PAIN IN HER ANKLE AND HAS SIGNIFICANT SWELLING, WHICH IS WORSE AT THE END OF THE DAY. SHE DESCRIBES A BURNING PAIN OVER THE ANTERIOR ANKLE."

    Patient problems: Extraskeletal ossification; Joint Impingement; Neuralgia; Swelling/ Edema

    MAUDE report 20854444 ↗
  • InjuryOct 11, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    AS REPORTED: "SUBJECT: (B)(6) INFINITY¿ WITH ADAPTIS¿ TOTAL ANKLE REPLACEMENT FOLLOW-UP (ITAR2) ON (B)(6) 2021: INTRA-OPERATIVE FRACTURE (R LATERAL DISTAL TIBIAL FX). AFTER MAKING TIBIAL CUT, A FRACTURE WAS NOTED IN R LAT DISTAL TIBIA-NON-DISPLACED. NO ADDITIONAL PROCEDURES OR FIXATION WAS NEEDED. COMPONENTS WERE STABLE, ANKLE WAS STABLE AT END OF PROCEDURE."

    Patient problems: Bone Fracture(s)

    MAUDE report 20426123 ↗
  • InjuryOct 11, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    AS REPORTED: "SUBJECT: (B)(6) INFINITY¿ WITH ADAPTIS¿ TOTAL ANKLE REPLACEMENT FOLLOW-UP (ITAR2). (B)(6) 2021: SUPERFICIAL WOUND DEHISCENCE, NO SIGN OF DEEP INFECTION. HEALED WITH LOCAL WOUND CARE AND PO ANTIBIOTIC. NO FUTURE SURGERY REQUIRED.".

    Patient problems: Wound Dehiscence

    MAUDE report 20426076 ↗
  • InjuryOct 11, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    AS REPORTED: "SUBJECT: 200-003INFINITY¿ WITH ADAPTIS¿ TOTAL ANKLE REPLACEMENT FOLLOW-UP (ITAR2) (B)(6) 2021: PERSISTENT PAIN. PATIENT STATES SHE IS HAVING SIGNIFICANT PAIN ON THE MEDIAL ASPECT OF HER ANKLE. (B)(6) 2022: RESOLVED. DENIES PAIN AT OFFICE VISIT.".

    Patient problems: Implant Pain

    MAUDE report 20426067 ↗
  • InjuryJan 9, 2024Adverse Event Without Identified Device or Use Problem; Osseointegration Problem; Adverse Event Without Identified Device or Use Problem; Osseointegration Problem

    THE DIGITAL STRYKER PROPHECY TEAM RECEIVED A CT SCAN INDICATING THAT THE PATIENT MAY REQUIRE A REVISION SURGERY OF BOTH THE TIBIAL AND TALAR COMPONENTS DUE TO PAIN AND LOOSE TIBIAL TRAY.

    Patient problems: Implant Pain; Inadequate Osseointegration

    MAUDE report 18479643 ↗
  • InjuryNov 22, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    THE DIGITAL STRYKER PROPHECY TEAM RECEIVED A CT SCAN INDICATING THAT THE PATIENT MAY REQUIRE A REVISION SURGERY DUE TO PAIN AND TIBIAL TRAY LOOSENING.

    Patient problems: Implant Pain

    MAUDE report 18196249 ↗
  • InjurySep 15, 2023Osseointegration Problem; Insufficient Device Problem Information; Osseointegration Problem; Insufficient Device Problem Information

    THE DIGITAL STRYKER PROPHECY TEAM RECEIVED A CT SCAN INDICATING THAT THE PATIENT MAY REQUIRE A REVISION SURGERY DUE TO CONTINUED PAIN FOR TWO YEARS SINCE THE IMPLANT PROCEDURE. THE PHYSICIAN PRESUMES LOOSENING WITH ASSOCIATED PAIN.

    Patient problems: Implant Pain; Inadequate Osseointegration

    MAUDE report 17757701 ↗
  • InjuryJul 6, 2023Appropriate Device Problem Term/Code Not Available; Migration; Appropriate Device Problem Term/Code Not Available; Migration

    THE DIGITAL STRYKER PROPHECY TEAM RECEIVED A CT SCAN INDICATING THAT THE PATIENT GETS A STRESS FRACTURE AND IMPLANT LOOSENING FOLLOWING STRESS FRACTURE. THE PATIENT WILL BE UNDERGOING A REVISION PROCEDURE.

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available; Bone Fracture(s)

    MAUDE report 17264121 ↗
  • InjuryNov 1, 2021Migration; Migration

    IT WAS REPORTED THAT THE PATIENT UNDERWENT A TOTAL ANKLE REPLACEMENT SURGERY. ALLEGEDLY, THE PATIENT WILL NEED TO UNDERGO A REVISION SURGERY DUE TO TALOR COMPONENT COLLAPSE. WITHIN PAST COUPLE OF MONTHS, TALUS/IMPLANT STARTING TO COLLAPSE AND FORCING IMPLANT ANTERIOR AND STRUCTURE NO LONGER SUPPORTIVE AS SEEN ON CURRENT LATERAL WEIGHT BEARING X-RAY.

    Patient problems: Failure of Implant

    MAUDE report 12733221 ↗

Other PROSTHESIS, ANKLE, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER devices

All PROSTHESIS, ANKLE, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.