Skip to content

TrueBeam 2.5Recalls & reported problems

Varian Medical Systems, Inc. · Class 2

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00899475002004), received between Jun 22, 2018 and Jun 22, 2018 — classified as 1 malfunction.

The most frequently reported specific device problem is “No Display/Image” (1 report).

1 other recall cover devices cleared under the same FDA submission (K143224); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Most reported problems

Device problems

  • No Display/Image1

Patient problems

  • No Known Impact Or Consequence To Patient1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJun 22, 2018No Display/Image

    DURING TREATMENT, WE EXPERIENCED A COLLIMATOR FAULT. WE ENTERED HOME SCREEN FOR THE LINEAR ACCELERATOR AND REINITIALIZED THE MLC'S. WHEN WE RETURNED TO THE TREATMENT SCREEN, IT DID NOT DISPLAY WHAT WAS TREATED. THE PLAN WAS SENT BACK TO MOSAIQ EMR AND NO DOSE OR TREATMENT PARAMETERS WERE RECORDED. PHYSICS NOTIFIED. PATIENT WAS REMOVED FROM THE TREATMENT ROOM. PHYSICIST WAS ABLE TO REVIEW THE LOG FILE ON THE LINEAR ACCELERATOR AND RETRIEVE THE DATA. PHYSICIST MANUALLY ENTERED THE TREATMENT DATA. PATIENT WAS ABLE TO COMPLETE THE REMAINDER OF HIS TREATMENT. VARIAN NOTIFIED.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 7629652 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K143224. One submission often covers many models, so check the product and lot details in each record.

  • Z-0004-2019 · Aug 20, 2018 · Terminated
    TrueBeam Radiotherapy Delivery System version 2.0 and 2.5 EDGE Radiotherapy Delivery System version 2.0 and 2.5 Product Usage: The TrueBeam delivery systems are intended to provi

Other Accelerator, linear, medical devices

All Accelerator, linear, medical devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.