HD 120 MLC Head Assembly 8.5Recalls & reported problems
Varian Medical Systems, Inc. · Class 2
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 0
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 1
Summary
No MAUDE adverse event reports in our data reference this device identifier directly.
1 FDA recall record name this device's identifier.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Recalls naming this device
- Aug 12, 2019Terminated
After a recent upgrade to the collimator software version 8.5, the firm became aware that the multi-leaf collimator leaves did not move during an arc treatment. The issue is related to a transient carriage primary- secondary interlock due to carriage fault on the MLC that immediately preceded the initiation of the arc treatment. The issue will occur only with MLC software version 8.5 and only affects conformal arc treatments, including VMAT and RapidArc. There have been no reports of adverse health consequences due to this issue.
Z-1899-2020 · Root cause: Software Design Change · Recall details · FDA record ↗
Other Accelerator, linear, medical devices
- LANDAUER microSTAR® ii Dosimetry SystemLandauer, Inc. · Accelerator, Linear, Medical38 recalls0 reports
- MOSAIQElekta, Inc. · Accelerator, Linear, Medical3 recalls112 reports
- Elekta Medical Linear Accelerator 1552093Elekta Solutions AB · Accelerator, linear, medical5 recalls6 reports
- Elekta Medical Linear Accelerator 1552092Elekta Solutions AB · Accelerator, linear, medical5 recalls0 reports
- MOSAIQ 3.2Elekta Solutions AB · Accelerator, Linear, Medical4 recalls25 reports
- Elekta Medical Linear Accelerator 1550358Elekta Solutions AB · Accelerator, linear, medical4 recalls2 reports
- Elekta Medical Linear Accelerator 1552091Elekta Solutions AB · Accelerator, linear, medical4 recalls1 report
- Elekta Medical Linear Accelerator 1550359Elekta Solutions AB · Accelerator, linear, medical4 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page