VIDAS® HCG 30405-01Recalls & reported problems
Biomerieux SA · VIDAS ® HCG (HCG) assay is a quantitative immunoassay intended for use on VIDAS family instrument in the determination of hCG concentration in human serum or plasma. · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 03573026156923), received between Apr 15, 2022 and Aug 25, 2022 — classified as 3 malfunctions.
The most frequently reported specific device problem is “High Test Results” (1 report).
1 FDA recall record name this device's identifier, 1 still open.
3 other recalls cover devices cleared under the same FDA submission (K141133, K921302); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Recalls naming this device
- Nov 9, 2022Class IIOpen, Classified
As the temperature and time out of range was exceeded, product performances cannot be guaranteed
Z-0864-2023 · Root cause: Storage · Recall details · FDA record ↗
Most reported problems
Device problems
- High Test Results1
- Low Test Results1
- Non Reproducible Results1
Patient problems
- No Clinical Signs, Symptoms or Conditions3
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
MalfunctionAug 25, 2022Non Reproducible Results; Non Reproducible Results
A CUSTOMER IN THE UNITED STATES NOTIFIED BIOMÉRIEUX THAT THEY OBSERVED NON REPEATABLE UNDERESTIMATED RESULTS WHEN TESTING A PATIENT¿S SAMPLES USING VIDAS HCG 60 TESTS (LOT 1009059670, EXPIRY DATE 29-OCT-2022). ON (B)(6) 2022. THE PATIENT¿S FIRST SAMPLE WAS DRAWN ON (B)(6) 2022 AND TESTED WITH THE RESULT : 133.81 MUI/ML. THIS SAMPLE WAS RETESTED ON (B)(6) 2022 AND THE RESULT WAS 33.33 MUI/ML. THE SAME SAMPLE WAS STORED IN THE FRIDGE AND TESTED ON (B)(6) 2022: 4050 MUI/ML AND 4000 MUI/ML ON SECTION B. A NEW SAMPLE WAS DRAWN AND TESTED ON (B)(6) 2022. THE RESULTS FOR THIS SAME SAMPLE AFTER DILUTION B 1/20 WERE THE FOLLOWING: - RESULT 10:43 AM IN A2 POSITION: 5974.59 MUI/ML - RESULT AT 11:15 AM IN B2 POSITION: 6494.57 MUI/ML THE CUSTOMER STATED THAT THEY ARE CONCERNED THERE IS TOO MUCH DIFFERENCE BETWEEN THE RESULTS OBTAINED WITH THE FIRST SAMPLE AND THE SECOND AND NOT SURE IF THIS PATIENT IS CONSIDERED PREGNANT. THESE RESULTS WERE PRECEDED BY A VALID CALIBRATION PERFORMED ON (B)(6) 2022. THE LAST QCV RESULTS ON (B)(6) 2022 ON SECTIONS A AND B, SHOW THAT THERE IS NO PROBLEM WITH THE MINI VIDAS INSTRUMENT. AT THE TIME OF THIS ASSESSMENT, THERE WAS NO INDICATION THAT ANY PATIENT HARM OCC
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15296956 ↗MalfunctionJun 9, 2022High Test Results; High Test Results
A CUSTOMER FROM UNITED STATES NOTIFIED BIOMÉRIEUX OF OBTAINING OVERESTIMATED RESULT WHEN TESTING EXTERNAL QUALITY CONTROL ON VIDAS HCG 60 TESTS (REF. 30405-01, BATCH NUMBER: 1009059670, EXPIRY DATE: 29-OCT 2022). THE CUSTOMER REPORTED THAT THE EXTERNAL CONTROLS LEVEL 3 RESULT WAS TOO HIGH WHEN ANALYZED WITH VIDAS HCG 60 TESTS (REF. 30405-01, BATCH NUMBER: 1009059670). THE CUSTOMER DID NOT PROVIDE THE RESULT OBTAINED AND EXPECTED WITH THE EXTERNAL CONTROLS LEVEL 3. THE CUSTOMER ALSO MENTIONED THAT THE CONTROLS PROVIDED IN THE VIDAS KIT WERE CONFORM. THE CUSTOMER ALSO COMPARED THESE EXTERNAL RESULTS WITH ANOTHER INSTRUMENT THAT'S NOT A MINIVIDAS. FOLLOWING THIS ISSUE, THE CUSTOMER REQUESTED AN FSE (FIELD SERVICE ENGINEER) TO BE DISPATCHED ON SITE. AS IT WAS AN EXTERNAL QUALITY CONTROL, NO PATIENT SAMPLE WAS INVOLVED. A BIOMÉRIEUX INTERNAL INVESTIGATION WILL BE INITIATED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14644522 ↗MalfunctionApr 15, 2022Low Test Results; Low Test Results
A CUSTOMER IN THE UNITED STATES REPORTED TO BIOMÉRIEUX THAT THEY OBSERVED NON REPEATABLE UNDERESTIMATED RESULTS WHEN TESTING PATIENT SAMPLES USING VIDAS® HCG 60 TESTS (REF. 30405-01, LOT 1009059670, EXPIRY DATE = 29-OCT-2022) COMPARED TO ANOTHER BATCH. THE RESULTS WERE: RESULT ON (B)(6) 2022 AT 04:02 PM FOR A NEW SAMPLE OF THE SAME PATIENT USING VIDAS HCG 60 TESTS LOT 1009059670: (B)(6). RESULT ON (B)(6) 2022 FOR A NEW SAMPLE OF THE SAME PATIENT USING VIDAS HCG 60 TESTS LOT 1009059670: (B)(6). RESULT ON (B)(6) 2022 FOR A NEW SAMPLE OF THE SAME PATIENT USING VIDAS HCG 60 TESTS LOT 1009059670 : (B)(6). THESE RESULTS WERE PRECEDED BY A VALID CALIBRATION PERFORMED ON THE (B)(6) 2022. THE ONLY INFORMATION AVAILABLE AT THE TIME OF THIS ASSESSMENT REGARDING THE PATIENT IS THAT SHE WAS NOT TRYING TO GET PREGNANT. THE CUSTOMER REPORTED THAT THEY PERFORM QCV WEEKLY AND THAT IT WAS CONFORM. AT THE TIME OF THIS ASSESSMENT, THERE WAS NO INDICATION THAT ANY PATIENT HARM OCCURRED AS A RESULT FO THIS ISSUE. A BIOMÉRIEUX INTERNAL INVESTIGATION WILL BE INITIATED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14121086 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K141133, K921302. One submission often covers many models, so check the product and lot details in each record.
- Z-0797-2023 · Nov 9, 2022 · Open, ClassifiedVIDAS CLINICAL VIDAS RUB IGG 60 TESTS, CATALOG 30226
- Z-0427-2022 · Sep 22, 2021 · Open, ClassifiedVIDAS DEX@, Dimer Exclusion II, REF 30455-01
- Z-0928-2020 · Oct 22, 2019 · TerminatedBIOMERIEUX VIDAS 3, Compact immunoanalyzer with full traceability and automation
Other Visual, Pregnancy Hcg, Prescription Use devices
- McKesson 5001Mckesson Medical-Surgical Inc. · Visual, Pregnancy Hcg, Prescription Use2 recalls165 reports
- CARDINAL HEALTH B1077-23Cardinal Health 200, LLC · Visual, pregnancy hCG, prescription use1 recall149 reports
- McKesson 5000Mckesson Medical-Surgical Inc. · Visual, Pregnancy Hcg, Prescription Use2 recalls90 reports
- McKesson 5002Mckesson Medical-Surgical Inc. · Visual, Pregnancy Hcg, Prescription Use2 recalls33 reports
- Sure-Vue 4581405021Fisher Scientific Company L.L.C. · Visual, pregnancy hCG, prescription use1 recall83 reports
- Henry Schein 9004072Henry Schein, Inc. · Visual, pregnancy hCG, prescription use1 recall79 reports
- AlereTM 92210Alere San Diego, Inc. · Visual, Pregnancy Hcg, Prescription Use2 recalls23 reports
- MEDLINE MPH12025HCGMedline Industries, Inc. · Visual, pregnancy hCG, prescription use1 recall69 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page