VITEK 2 - AST Test kits (in vitro diagnostic test cards for… recallZ-1464-2024 · Biomerieux Inc · initiated Mar 14, 2024

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Biomerieux Inc or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Ceftriaxone (cro02n) concentration errors associated with specific recently manufactured test kits which would have the potential for for false susceptible results. The risk is present for isolates that have a result of MIC=0.5, 1, 2.

FDA root cause category: Employee error

What the company is doing

BioMerieux issued an URGENT FIELD SAFETY NOTICE to its consignees on 03/14/2024 via email. The notice explained the issue, impact to patient, and provided the consignee (user) with alternate testing methods. BioMerieux issued another URGENT FIELD SAFETY NOTICE dated 07MAY2024 to its consignees on May 9, 2024, via email. This letter includes the statement, "however, no results for the Ceftriaxone (cro02n) should be reported outside the lab."

Recall details

Recall number
Z-1464-2024
Status
Open, Classified
Initiated by the company
Mar 14, 2024
Posted by the FDA
Apr 25, 2024
Recalling firm
Biomerieux Inc
Quantity
49215 units
Product code
LON
FDA submissions
K161217
Distribution
Worldwide

Product description

VITEK 2 - AST Test kits (in vitro diagnostic test cards for antimicrobial susceptibility): a) AST-N334, REF 418984; b) AST-GN95, REF 421982; c) AST-GN99, REF 423102; d) AST-N390, REF 423340; e) AST-N391, REF 423341; f) AST-N392, REF 423342; g) AST-N395, REF 423491; h) AST-N401, REF 423643; i) AST-N402, REF 423644; j) AST-N404, REF 423664; k) AST-N802, REF 423706; l) AST-N405, REF 423864; m) AST-N417, REF 423880; n) AST-N408, REF 423924; o) AST-N409, REF 423925; p) AST-N420, REF 424039; q) AST-N423, REF 424042; r) AST-N422, REF 424056; s) AST-N436, REF 424440; t) AST-N809, REF 424703; u) AST-N806, REF 424709; v) AST-N807, REF 424710; w) AST-N808, REF 424711; x) AST-N810, REF 424712; y) AST-N812, REF 424721

Code information (models, lots, serial numbers)

Show the full code information (3k characters)

a) AST-N334, REF 418984, UDI/DI 3573026490720, Lot/Serial Numbers: 7742790503, 7742805403, 7742828503, 7742830203; b) AST-GN95, REF 421982, UDI/DI 3573026565923, Lot/Serial Numbers: 6852740103, 6852751503, 6852778503, 6852811503; c) AST-GN99, REF 423102, UDI/DI 3573026608132, Lot/Serial Numbers: 6892721103, 6892727403, 6892740503, 6892740403, 6892743503, 6892761503, 6892770403, 6892790103, 6892799403, 6892799103, 6892810103, 6892818103, 6892831503; d) AST-N390, REF 423340, UDI/DI 3573026612177, Lot/Serial Numbers: 1402832104, 1402761504; e) AST-N391, REF 423341, UDI/DI 3573026612269, Lot/Serial Numbers: 1412831104, 1412754504; f) AST-N392, REF 423342, UDI/DI 3573026612283, Lot/Serial Numbers: 1422819204; g) AST-N395, REF 423491, UDI/DI 3573026615802, Lot/Serial Numbers: 1452725204; h) AST-N401, REF 423643, UDI/DI 3573026616816, Lot/Serial Numbers: 1512776404, 1512789204, 1512811104, 1512831104, 1512720404, 1512729404, 1512729104, 1512748504, 1512754104, 1512785404; i) AST-N402, REF 423644, UDI/DI 3573026616830, Lot/Serial Numbers: 1522720504, 1522733404; j) AST-N404, REF 423664, UDI/DI 3573026617356, Lot/Serial Numbers: 1542819204; k) AST-N802, REF 423706, UDI/DI 3573026617752, Lot/Serial Numbers: 0422750104, 0422763104, 0422764504, 0422782404, 0422820504, 0422820404; l) AST-N405, REF 423864, UDI/DI 3573026621360, Lot/Serial Numbers: 1552757504, 1552792404, 1552800404, 1552820104, 1552733104, 1552741504, 1552757404; m) AST-N417, REF 423880, UDI/DI 3573026621704, Lot/Serial Numbers: 0192763504, 0192819404; n) AST-N408, REF 423924, UDI/DI 3573026622725, Lot/Serial Numbers: 1582810504, 1582820104, 1582734404, 1582742504, 1582764404, 1582782504; o) AST-N409, REF 423925, UDI/DI 3573026622749, Lot/Serial Numbers: 1592811104, 1592746404; p) AST-N420, REF 424039, UDI/DI 3573026625283, Lot/Serial Numbers: 0462810404; q) AST-N423, REF 424042, UDI/DI 3573026625337, Lot/Serial Numbers: 0502750404, 0502825504; r) AST-N422, REF 424056, UDI/DI 3573026625610, Lot/Serial Numbers: 0482819504; s) AST-N436, REF 424440, UDI/DI 3573026631666, Lot/Serial Numbers: 0672734104, 0672754104; t) AST-N809, REF 424703, UDI/DI 3573026635978, Lot/Serial Numbers: 0962799404; u) AST-N806, REF 424709, UDI/DI 3573026636067, Lot/Serial Numbers: 0932729404; v) AST-N807, REF 424710, UDI/DI 3573026636081, Lot/Serial Numbers: 0942723404; w) AST-N808, REF 424711, UDI/DI 3573026636104, Lot/Serial Numbers: 0952758104; x) AST-N810, REF 424712, UDI/DI 3573026636128, Lot/Serial Numbers: 0972729504; y) AST-N812, REF 424721, UDI/DI 3573026636296, Lot/Serial Numbers: 0992777104, 0992776504, 0992821404