Skip to content

GENESIS 114968Recalls & reported problems

Smith & Nephew, Inc. · GENESIS PIN AND DRILL SET · Class 2

All models of GENESIS (14)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 03596010051332), received between Jul 29, 2022 and Dec 8, 2023 — classified as 2 injuries.

The most frequently reported specific device problem is “Fracture” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 1 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Fracture2

Patient problems

  • Foreign Body In Patient2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • InjuryDec 8, 2023Fracture; Fracture

    IT WAS REPORTED THAT, AFTER A TKA SURGERY, IT WAS NOTICED WITH AN X-RAY THAT THE DRILL PIN TIP HAD BROKEN OFF DURING SURGERY AND WAS BURIED INSIDE OF THE PATIENT. THE BROKEN PIECE HAS NOT BEEN RETRIEVED FROM PATIENT. CURRENT HEALTH STATUS OF PATIENT IS UNKNOWN.

    Patient problems: Foreign Body In Patient

    MAUDE report 18291549 ↗
  • InjuryJul 29, 2022Fracture; Fracture

    IT WAS REPORTED THAT, DURING A TKA SURGERY, A GENESIS PIN AND DRILL SET BROKE WHEN THE SURGEON WAS DRILLING THE HOLES, PRIOR TO PINNING TO SECURE AND FINISH THE TIBIAL REAM. THE BROKEN PART COULD NOT BE REMOVED AND REMAINED IN THE PATIENT'S TIBIA. SURGERY WAS FINISHED WITHOUT FURTHER USAGE OF THE DRILL. NO SIGNIFICANT DELAYS REPORTED. CURRENT HEALTH STATUS OF PATIENT IS UNKNOWN.

    Patient problems: Foreign Body In Patient

    MAUDE report 15136635 ↗

Other PLATE, FIXATION, BONE devices

All PLATE, FIXATION, BONE devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page