Skip to content

LARGE CANNULATED SCREWS 71127105Recalls & reported problems

Smith & Nephew, Inc. · 7.0MM X 105MM CANNULATED SCREW 16MM THREAD STAINLESS STEEL · Class 2 · Implant

All models of LARGE CANNULATED SCREWS (21)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 03596010175816), received between Mar 9, 2021 and Mar 9, 2021 — classified as 1 injury.

The most frequently reported specific device problem is “Incorrect Measurement” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Most reported problems

Device problems

  • Incorrect Measurement1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • InjuryMar 9, 2021Incorrect Measurement; Incorrect Measurement

    IT WAS REPORTED THAT DURING SURGERY FOR NECK OF FEMUR USING SMITH AND NEPHEW CHS SYSTEM, THE SURGICAL TEAM DECIDED TO USE THE ANTI-ROTATION SCREW TREATMENT WITH SMITH AND NEPHEW 7.0MM CANNULATED SCREW SYSTEM. FOLLOWING CORRECT POSITIONING OF THE 3.2MM X 300MM THREADED GUIDE PIN, A MEASUREMENT WAS TAKEN USING THE 6.5/8.0MM DIRECT MEASURING GAUGE. REGISTRAR READ 105MM FROM END OF THE GUIDE PIN AND REQUESTED A 7.0MM X 105MM CANNULATED SCREW 16MM THREAD STAINLESS STEEL TO BE OPENED FOR USE. DURING INSERTION OF THE 105MM SCREW, IT WAS NOTED THAT THE 105MM SCREW WAS APPROX 10MM TOO LONG. A 7.0MM X 95MM CANNULATED SCREW 16MM THREAD WAS THEN REQUESTED AND INSERTED OVER THE GUIDE PIN. THE 95MM SCREW WAS NOTED TO BE OF THE CORRECT LENGTH. FINALLY, THE CHS IMPLANTATION WAS THEN COMPLETED WITH A SURGICAL DELAY OF LESS THAN 30 MINUTES AND NO INJURY TO THE PATIENT. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED ACCORDINGLY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11444215 ↗

Other PLATE, FIXATION, BONE devices

All PLATE, FIXATION, BONE devices →

Sources and updates