BIORCI 7207561Recalls & reported problems
Smith & Nephew, Inc. · 8 X 30 MM BIORCI SCREW · Class 2 · Implant
All models of BIORCI (32)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 15
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 15 adverse event reports that list this exact device (DI 03596010426239), received between Apr 21, 2018 and Dec 12, 2024 — classified as 1 injury, 14 malfunctions.
The most frequently reported specific device problem is “Break” (11 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Break11
- Material Separation4
Patient problems
- No Known Impact Or Consequence To Patient12
- No Clinical Signs, Symptoms or Conditions2
- Foreign Body In Patient1
- No Information1
Recent adverse event reports
The 15 most recent of 15 reports, in the reporter's own words as published by the FDA.
MalfunctionDec 12, 2024Break; Break
IT WAS REPORTED THAT, DURING AN ARTHROSCOPY, THE TIP OF THE BIORCI SCREW GOT BROKEN WHEN IT WAS BEING PLACED INTO THE TIBIAL BONE. SURGERY WAS COMPLETED WITH A SMITH AND NEPHEW BACK-UP DEVICE INSTEAD. THERE WAS A DELAY OF LESS THAN 30 MINUTES, AND NO FURTHER COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20916163 ↗InjuryNov 1, 2020Break; Break
IT WAS REPORTED THAT, DURING A PROCEDURE, THE BIORCI ANCHOR BROKE WHILE INSERTING IT INTO TIBIA. THERE WAS A BACKUP DEVICE AVAILABLE. NO DELAY OR FURTHER COMPLICATIONS WERE REPORTED. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED ACCORDINGLY.
Patient problems: No Clinical Signs, Symptoms or Conditions; No Information
MAUDE report 10767885 ↗MalfunctionApr 5, 2020Material Separation; Material Separation
IT WAS REPORTED THAT DURING A TIBIAL FIXATION THE BIORCI SCREW BROKE WHILE IT WAS SCREWING INTO THE BONE. THE BROKEN PIECE WAS RETRIEVED USING A GRASPER. THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITHOUT A SIGNIFICANT DELAY USING A BACK-UP DEVICE. NO PATIENT INJURY OR OTHER COMPLICATIONS WERE REPORTED. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED ACCORDINGLY.
Patient problems: Foreign Body In Patient
MAUDE report 9926978 ↗MalfunctionAug 26, 2019Material Separation; Material Separation
IT WAS REPORTED THAT DURING AN ACL RECONSTRUCTION SURGERY, WHILE THE SCREW WAS BEING IMPLANTED THE FRONT END BROKE OFF. NO PATIENT INJURIES OR SIGNIFICANT DELAY REPORTED. BACK-UP DEVICE WAS AVAILABLE TO COMPLETE THE SURGERY. ALL THE BROKEN PIECES WERE REMOVED FROM THE PATIENT'S ANATOMY.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8932825 ↗MalfunctionAug 26, 2019Material Separation; Material Separation
IT WAS REPORTED THAT DURING AN ACL RECONSTRUCTION SURGERY, WHEN THE DEVICE WAS BEING IMPLANTED, THE FRONT END BROKE OFF. NO PATIENT INJURIES OR SIGNIFICANT DELAY REPORTED. BACK-UP DEVICE WAS AVAILABLE. ALL THE BROKEN PIECES WERE REMOVED FROM THE PATIENT'S ANATOMY.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8932793 ↗MalfunctionAug 5, 2019Material Separation; Material Separation
IT WAS REPORTED THAT DURING AN ACL RECONSTRUCTION SURGERY, THE SCREW BROKE, BROKEN PIECES WERE REMOVED USING GRASPERS THE PROCEDURE WAS COMPLETED WITH A BACK UP DEVICE WITH NO DELAY OR PATIENT INJURY REPORTED.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8859417 ↗MalfunctionDec 5, 2018Break
IT WAS REPORTED THAT THE SCREW BROKE WHILE TIGHTENING. THERE WAS NOT DELAY ON THE PROCEDURE AND A BACK-UP WAS AVAILABLE TO COMPLETE THE PROCEDURE. UNKNOWN IF THE PIECES WERE RETRIEVED FROM THE PATIENT. NO PATIENT INJURY WAS REPORTED.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8135882 ↗MalfunctionOct 4, 2018Break
IT WAS REPORTED THAT THE DEVICE WAS BROKEN AT THE SURGERY. NO PATIENT INJURY WAS REPORTED.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7935325 ↗MalfunctionSep 13, 2018Break
IT WAS REPORTED THAT THE SCREW BROKE WHEN ENTERING THE TUNNEL. A BACKUP DEVICE WAS USED TO COMPLETE THE SURGERY. THERE WAS A SURGICAL DELAY GREATER THAN 30 MINUTES.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7873896 ↗MalfunctionSep 13, 2018Break
IT WAS REPORTED THAT THE SCREW BROKE WHEN ENTERING THE TUNNEL. A BACKUP DEVICE WAS USED TO COMPLETE THE SURGERY. THERE WAS A SURGICAL DELAY GREATER THAN 30 MINUTES.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7873882 ↗MalfunctionSep 12, 2018Break
IT WAS REPORTED THAT THE SCREW BROKE WHEN ENTERING THE TUNNEL. A BACKUP DEVICE WAS USED TO COMPLETE THE SURGERY. THERE WAS A SURGICAL DELAY GREATER THAN 30 MINUTES.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7870606 ↗MalfunctionJun 29, 2018Break
IT WAS REPORTED THAT THE SCREW BROKE. ANOTHER SCREW WAS USED TO COMPLETE THE PROCEDURE. NO SIGNIFICANT DELAY OR PATIENT INJURY REPORTED.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7649418 ↗MalfunctionJun 29, 2018Break
IT WAS REPORTED THAT THE SCREW BROKE. ANOTHER SCREW WAS USED TO COMPLETE THE PROCEDURE. NO SIGNIFICANT DELAY OR PATIENT INJURY REPORTED.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7649416 ↗MalfunctionApr 25, 2018Break
IT WAS REPORTED THAT THE INTERFERENCE SCREW BROKE IN THE FEMORAL TUNNEL AND WAS REMOVED. NO PATIENT INJURY REPORTED.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7458603 ↗MalfunctionApr 21, 2018Break
IT WAS REPORTED THAT WHILE ENGAGING THE SCREW IN THE TIBIAL SIDE DURING INSERTION, THE SCREW BROKE AT THE TIP. NO PATIENT INJURY WAS REPORTED.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7449509 ↗
Other Screw, fixation, bone devices
- NA 03.043.029Synthes GmbH · Screw, fixation, bone503 reports
- OSTEORAPTOR 72201995Smith & Nephew, Inc. · Screw, fixation, bone253 reports
- Control 360 SK12Treace Medical Concepts, Inc. · Screw, Fixation, Bone195 reports
- HEALIX PEEK ORTHOCORD 222205Medos International Sàrl · Screw, fixation, bone186 reports
- PRECICE® Bone Transport System STP3.5-350Nuvasive Specialized Orthopedics, Inc. · Screw, Fixation, Bone3 recalls0 reports
- PRECICE® Bone Transport System STP3.5-400Nuvasive Specialized Orthopedics, Inc. · Screw, Fixation, Bone3 recalls0 reports
- PRECICE® Bone Transport System STP3.5-300Nuvasive Specialized Orthopedics, Inc. · Screw, Fixation, Bone3 recalls0 reports
- PRECICE® Bone Transport System STP3.5-200Nuvasive Specialized Orthopedics, Inc. · Screw, Fixation, Bone3 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page