T-FIX 014771Recalls & reported problems
Smith & Nephew, Inc. · T-FIX RCG DRILL PAC, STERILE, WITH 1 DRILL AND 2 SLEEVES · Class 2
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 9
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 9 adverse event reports that list this exact device (DI 03596010446114), received between Mar 27, 2018 and Nov 17, 2025 — classified as 2 injuries, 7 malfunctions.
The most frequently reported specific device problem is “Break” (7 reports).
1 other recall cover devices cleared under the same FDA submission (K925573); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Break7
- Detachment of Device or Device Component2
- Delivered as Unsterile Product1
- Fracture1
- Material Deformation1
Patient problems
- No Clinical Signs, Symptoms or Conditions5
- No Known Impact Or Consequence To Patient2
- Foreign Body In Patient1
- Perforation1
Recent adverse event reports
The 9 most recent of 9 reports, in the reporter's own words as published by the FDA.
MalfunctionNov 17, 2025Break; Material Deformation; Break; Material Deformation
IT WAS REPORTED THAT DURING A PROCEDURE, WHILE DRILLING INTO THE GLENOID THROUGH THE GUIDE, THE INNER DRILL OF THE T-FIX RCG STERILE PAC WAS BROKEN. ALL BROKEN PIECES WERE REMOVED FROM THE PATIENT. THE PROCEDURE WAS RESUMED, AFTER A NON-SIGNIFICANT DELAY, USING AN EQUIVALENT SN BACK-UP DEVICE. NO FURTHER COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23569519 ↗MalfunctionOct 22, 2025Break; Break
IT WAS REPORTED THAT, DURING A SURGERY, THE WING AT THE TIP OF ONE (1) T-FIX DRILL TIP DISINTEGRATED, BROKE OFF IN THE PATIENT UPON OPENING. THE PROCEDURE WAS RESUMED, AFTER A 10 TO 15-MINUTE DELAY, USING AN EQUIVALENT S+N BACK-UP. NO FURTHER COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23356131 ↗MalfunctionJan 28, 2025Delivered as Unsterile Product; Delivered as Unsterile Product
IT WAS REPORTED THAT THE T-FIX RCG PACKING WAS NOT SEALED WELL. IT IS UNKNOWN WHETHER THE EVENT HAPPENED DURING SURGERY, IF THERE WAS PATIENT INVOLVEMENT. HOWEVER, THERE WAS NO SURGICAL DELAY AND A SMITH AND NEPHEW BACK UP DEVICE WAS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21260906 ↗MalfunctionJun 14, 2024Fracture; Detachment of Device or Device Component; Fracture; Detachment of Device or Device Component
IT WAS REPORTED THAT DURING AN ARTHROSCOPIC LATARJET SURGERY, WHEN DRILLING A BONE HOLE BY THE GLENOID FOSSA, THE PART OF THE T-FIX THAT CONNECTS TO THE OUTER DRILL CAME OFF AND THE DEVICE COULD NO LONGER BE CONNECTED TO THE OUTER DRILL. THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITH A DELAY OF 10 MINUTES USING A SMITH AND NEPHEW BACK UP DEVICE. NO FURTHER COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19542026 ↗InjuryApr 11, 2024Break; Break
IT WAS REPORTED THAT DURING A MINI-OPEN LATERJET PROCEDURE, THE SLEEVE OF THE 2.0MM T-FIX DRILL BROKE WHEN DRILLING THE INFERIOR HOLE USING WITH A GLENOID GUIDE. THE METAL SLEEVE BROKE OFF ABOUT 1 INCH INSIDE THE GLENOID. THE PIECES WERE REMOVED FROM THE PATIENT. THERE WAS A DELAY IN THE LESS THAN 30 MINUTES AND IT WAS COMPLETED WITH A COMPETITOR DEVICE ( SMALL AND LARGE FRAG SCREW SET, SYNTHES) USING AN ADDITIONAL BONE HOLE. NO FURTHER COMPLICATIONS WERE REPORTED.
Patient problems: Perforation
MAUDE report 19094127 ↗InjuryJun 16, 2021Break; Break
IT WAS REPORTED THAT, DURING AN ARTHROSCOPIC BONE BLOCK SURGERY, WHEN TRYING TO DRILL THE ¿RIGHT¿ BULLET FROM THE JIG, RESISTANCE WAS FELT WHEN DRILLING. IT WAS NOTICED THAT A SMALL FRAGMENT OF THE DRILL SLEEVE WAS STUCK IN THE BULLET. THE BULLET WAS REMOVED AND REMNANT OF THE SLEEVE REMOVED FROM THE BULLET. A SECOND PACK WAS OPENED AND USED TO CHECK IF THE NEW SLEEVE WOULD PASS SAFELY THROUGH THE RIGHT BULLET, THE RIGHT TUNNEL WAS SUCCESSFULLY DRILLED WITH THE BACKUP PACK. A DELAY OF 3 MINUTES WAS REPORTED AND NO FURTHER COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12010336 ↗MalfunctionFeb 13, 2019Break
IT WAS REPORTED THAT DURING AN UNKNOWN PROCEDURE WHILE USING THE DRILL AND SLEEVE TO DRILL THROUGH THE GLENOID BONE IN THE SHOULDER VIA THE GLENOID DRILLING GUIDE THE DRILL OF THE PRODUCT BROKE OF IN THE BONE. THE FLUTED ENDS WERE EXPOSED BUT THE DRILL WAS STILL PARTLY IN THE GLENOID. A GRASPER WAS USED TO PULL THE DRILL OUT OF THE BONE AND REMOVED FROM THE PATIENT. A NEW DRILL AND SLEEVE WAS OPENED AND THE PROCEDURE CONTINUED WITH ONLY A FEW MINUTES DELAY (3MINS). ALL THE DRILL PARTS WERE REMOVED. A BACK UP DEVICE WAS AVAILABLE AND NO PATIENT INJURIES OR SIGNIFICANT DELAY WERE REPORTED.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8336732 ↗MalfunctionJan 24, 2019Break; Detachment of Device or Device Component
IT WAS REPORTED THAT DURING AN ARTHROSCOPIC LATERJET, WHILE USING THE DRILL AND SLEEVE TO DRILL THROUGH THE GLENOID BONE IN THE SHOULDER VIA THE GLENOID DRILLING GUIDE THE DRILL OF THE PRODUCT BROKE OF IN THE BONE. THE FLUTED ENDS WERE EXPOSED BUT THE DRILL WAS STILL PARTLY IN THE GLENOID. A GRASPER WAS USED TO PULL THE DRILL OUT OF THE BONE AND REMOVED FROM THE PATIENT. A NEW DRILL AND SLEEVE WAS OPENED AND THE PROCEDURE CONTINUED WITH ONLY A FEW MINUTES DELAY (3MINS). I RETRIEVED THE DRILL PARTS TO BE ANALYSED.
Patient problems: Foreign Body In Patient
MAUDE report 8274634 ↗MalfunctionMar 27, 2018Break
IT WAS REPORTED THAT THE T-FIX WAS READY TO BE USED, PLUGGED ON THE POWER TOOL. THE T-FIX BROKE AT FULL SPEED, ENTERING THE CORACOID PROCESS. THE TIP OF THE MALE PART BROKE, JUST UNDER THE CONNECTION OF MALE/FEMALE PARTS. BROKEN PIECE WAS REMOVED FROM THE PATIENT. PROCEDURE WAS COMPLETED WITH A BACK-UP DEVICE. NO PATIENT INJURY REPORTED.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7376039 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K925573. One submission often covers many models, so check the product and lot details in each record.
- Z-2702-2020 · Jun 24, 2020 · Open, ClassifiedSmith & Nephew T-F1X0 RCG Drill Pac, 2 mm Drill & (2) Sleeves - Product Usage: indicated for use in surgical arthroscopic rotator cuff repair procedures. REF/ Catalog Number : 0147
Other ARTHROSCOPE devices
- FMS 283512Medos International Sàrl · ARTHROSCOPE1,800 reports
- Arthrex® AR-6480Arthrex, Inc. · ARTHROSCOPE949 reports
- FMS VUE 284002Depuy Mitek, LLC · ARTHROSCOPE829 reports
- MITEK 242018Medos International Sàrl · Arthroscope731 reports
- METRx® System 9560100Medtronic Sofamor Danek, Inc. · Arthroscope667 reports
- FMS VUE II 284004Medos International Sàrl · Arthroscope562 reports
- VAPR COOLPULSE 228146Depuy Mitek, LLC · ARTHROSCOPE474 reports
- FMS DUO 284580Depuy Mitek, LLC · ARTHROSCOPE458 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page