BIORCI 7209016Recalls & reported problems
Smith & Nephew, Inc. · BIORCI-HA INTERFERENCE SCREW 10 MM X 25 MM STERILE · Class 2 · Implant
Data updated Oct 13, 2026 from official FDA records · Editorial policy
Quick summary
This Smith & Nephew BIORCI-HA interference screw is an implant used to fix bone, typically during knee arthroscopy or ACL reconstruction. The 25 reports name no deaths and 9 injuries; most describe the screw breaking while being inserted into the tibia, with procedures finished using a backup device after short delays. No recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 25
- Deaths reported
- 0
- Injuries reported
- 9
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 25 adverse event reports that list this exact device (DI 03596010460035), received between Dec 22, 2018 and Sep 16, 2025 — classified as 9 injuries, 16 malfunctions.
The most frequently reported specific device problem is “Break” (18 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Break18
- Material Separation3
- Activation, Positioning or Separation Problem1
- Crack1
- Device Dislodged or Dislocated1
- Difficult to Insert1
- Flaked1
- Migration or Expulsion of Device1
Patient problems
- No Clinical Signs, Symptoms or Conditions18
- No Known Impact Or Consequence To Patient5
- No Information1
- Unspecified Tissue Injury1
Recent adverse event reports
The 25 most recent of 25 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 16, 2025Break; Break
IT WAS REPORTED THAT DURING AN ARTHROSCOPY, IT WAS NOTICED THAT THE TIP OF THE BIORCI STERILE SCREW BROKE OFF WHILE INSERTING IT IN THE TIBIAL DRILLED SLOT. THE ARTHROSCOPY WAS RESUMED, AFTER A NON SIGNIFICANT DELAY, USING AN S+N BACK UP. NO FURTHER COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23064437 ↗MalfunctionJan 17, 2025Break; Break
IT WAS REPORTED THAT DURING AN ARTHROSCOPY, WHILE PUTTING THE BIORCI-HA SCREW IN THE TIBIA, IT BROKE. THE PROCEDURE WAS COMPLETED USING A BACK-UP DEVICE. THERE WAS A DELAY LESS THAN 30 MINUTES AND NO FURTHER COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21185190 ↗MalfunctionOct 15, 2024Break; Break
IT WAS REPORTED THAT DURING AN ACL RECONSTRUCTION, THE SCREW BROKE WHILE FIXATION IN THE TIBIAL. ALL THE SCREW PARTICLES WERE REMOVED FROM THE SITE USING FORCEP. THE PROCEDURE WAS COMPLETED WITH A SMITH AND NEPHEW BACK UP DEVICE USING THE ORIGINAL DRILLED BONE HOLE USING A TIBIA DRILL BIT TO PREPARE THE INSERTION SITE WITH NO SURGICAL DELAY. NO VOID LEFT IN PATIENT AND NO FURTHER COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20450918 ↗MalfunctionOct 8, 2024Break; Break
IT WAS REPORTED THAT DURING AN ARTHROSCOPY, THE FRONT PART OF THE BIORCI SCREW BROKE WHEN THE SURGEON INSERTED IT IN THE TIBIA. THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITH A DELAY OF LESS THAN 30 MINUTES USING A SMITH AND NEPHEW BACK UP DEVICE. THE BROKEN PIECES WERE RETRIEVED WITH THE HELP OF TWEEZERS. THE BACK UP DEVICE WAS USED IN THE ORIGINALLY DRILLED BONE HOLE SO NO VOID WAS LEFT IN THE PATIENT. NO FURTHER COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20401912 ↗MalfunctionAug 20, 2024Break; Break
IT WAS REPORTED THAT DURING AN ARTHROSCOPY, THE TIP OF THE SCR BIORCI HA SCREW BROKE DURING THE INSERTION. THE PROCEDURE WAS COMPLETED WITH A SMITH AND NEPHEW BACK UP DEVICE. THERE WAS A DELAY LESS THAN 30 MINUTES AND NO FURTHER COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20007145 ↗MalfunctionMar 18, 2024Break; Break
IT WAS REPORTED THAT DURING AN ARTHROSCOPY, THE BIORCI-HA SCREW GOT BROKEN. THE PROCEDURE WAS FINISHED WITH A SMITH AND NEPHEW BACK UP DEVICE. THERE WAS A SURGICAL DELAY LESS THAN 30 MINUTES AND NO FURTHER COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18922109 ↗MalfunctionMar 11, 2024Device Dislodged or Dislocated; Device Dislodged or Dislocated
IT WAS REPORTED THAT DURING AN ARTHROSCOPY, THE BIORCI HELICOIL SCREW WAS PULLOUT FROM THE BONE. THE PROCEDURE WAS COMPLETED WITH A SMITH AND NEPHEW BACK UP DEVICE. THERE WAS A DELAY LESS THAN 30 MINUTES AND NO FURTHER COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18874394 ↗MalfunctionSep 11, 2023Break; Break
IT WAS REPORTED THAT, DURING AN ARTHROSCOPY, THE BIORCI SCREW GOT DAMAGED WHILE BEING INSERTED IN THE TUNNEL, THE THREADS BROKE. SURGERY WAS COMPLETED AFTER A NON-SIGNIFICANT DELAY, USING A BACK-UP DEVICE INSTEAD. NO FURTHER COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17724544 ↗MalfunctionJun 26, 2023Break; Break
IT WAS REPORTED THAT DURING AN ARTHROSCOPY, THE BIORCI-HA SCREW BROKE INTO PIECES WHILE INSERTING INTO THE TIBIA TUNNEL. ALL THE PIECES WERE REMOVED FROM THE PATIENT WITH SUCTION. THE PROCEDURE WAS COMPLETED WITH NON-SIGNIFICANT SURGICAL DELAY USING A BACK-UP DEVICE IN THE ORIGINALLY DRILLED BONE HOLE. NO FURTHER COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17199747 ↗MalfunctionMay 12, 2023Break; Break
IT WAS REPORTED THAT, DURING AN ARTHROSCOPY, THE THREADS OF THE BIORCI SCREW BROKE WHILE SCREWING IT IN THE TIBIAL TUNNEL. SURGERY WAS COMPLETED AFTER A NON-SIGNIFICANT DELAY, USING A BACK-UP DEVICE INSTEAD. NO FURTHER COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16924885 ↗MalfunctionApr 4, 2023Flaked; Flaked
IT WAS REPORTED THAT DURING AN ARTHROSCOPY, THE BIORCI-HA SCREW THREAD WAS SHREDDING DURING THE DEPLOYMENT OF THE SCREW DURING FIXATION. IT IS UNKNOWN WHETHER THE EVENT HAPPENED DURING SURGERY AND IF THERE WAS PATIENT INVOLVEMENT, AND IF THERE WAS A BACK-UP DEVICE AVAILABLE OR IF THERE WAS A DELAY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16676055 ↗InjuryFeb 1, 2023Difficult to Insert; Difficult to Insert
IT WAS REPORTED THAT DURING AN ACL PROCEDURE, THE BIORCI-HA SCREW COULD NOT ENGAGE AND HENCE WAS TAKEN OUT. THE SURGERY WAS RESUMED, AFTER A NON-SIGNIFICANT DELAY, WITH A BACK-UP DEVICE IN THE ORIGINALLY DRILLED HOLE. NO FURTHER COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16279478 ↗InjuryOct 17, 2022Break; Break
IT WAS REPORTED THAT, DURING AN ARTHROSCOPY, THE HEAD OF THE BIORCI-HA 10X25 STERILE BROKE INSIDE THE PATIENT, DURING FIXATION. THE BROKEN PIECE WAS SUCCESSFULLY RETRIEVED BY TWEEZER. NO DELAY OR FURTHER COMPLICATIONS WERE REPORTED, AND THE PROCEDURE WAS FINISHED WITH A SMITH AND NEPHEW BACK UP DEVICE USED IN THE ORIGINALLY DRILLED BONE HOLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15616305 ↗InjuryJul 22, 2022Crack; Crack
IT WAS REPORTED THAT DURING AN ARTHROSCOPY, THE BIORCI SCREW GOT CRACKED. THE PROCEDURE WAS COMPLETED WITH A S+N BACK UP DEVICE. THERE WAS NO DELAY AND NO FURTHER COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15089723 ↗InjuryMay 20, 2021Break; Break
IT WAS REPORTED THAT DURING A ACL RECONSTRUCTION PROCEDURE, THE SCR BIORCI-HA 10X25 STERILE BROKE DURING DEPLOYMENT INTO THE TIBIA. THE PROCEDURE WAS COMPLETED USING A BACK-UP DEVICE. A DELAY OF LESS THAN 30 MINUTES WAS REPORTED. NO PATIENT COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11859190 ↗InjuryApr 14, 2021Break; Break
IT WAS REPORTED THAT DURING A SURGERY, THE BIORCI SCREW BROKE WHILE IT WAS BEING DEPLOYED. THE BROKEN PIECES WERE REMOVED BY USING TWEEZERS. THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITHOUT A SIGNIFICANT DELAY USING A BACK-UP DEVICE. NO OTHER COMPLICATIONS WERE REPORTED. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED ACCORDINGLY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11668451 ↗InjuryMar 30, 2021Break; Break
IT WAS REPORTED THAT DURING SURGERY, INTERNAL TO THE PATIENT, THE TIP OF BIORCI-HA SCREW TIP BROKE IN THE TIBIA. A BACK-UP DEVICE WAS AVAILABLE. IT IS UNKNOWN IF A SIGNIFICANT DELAY OCCURRED. NO OTHER COMPLICATIONS WERE REPORTED. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED ACCORDINGLY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11593337 ↗InjuryMar 30, 2021Break; Break
IT WAS REPORTED THAT DURING SURGERY, INTERNAL TO THE PATIENT, THE TIP OF THE BIORCI-HA SCREW BROKE WHILE INSERTING IT IN THE TIBIA. THE PROCEDURE WAS COMPLETED WITHOUT DELAY USING A BACK-UP DEVICE. NO FURTHER COMPLICATIONS REPORTED. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED ACCORDINGLY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11593329 ↗InjuryMar 3, 2021Migration or Expulsion of Device; Migration or Expulsion of Device
IT WAS REPORTED THAT DURING A SURGERY, AFTER THE BIORCI-HA SCREW WAS INSERTED INTO THE BONE TUNNEL AND IT BECAME LOOSE. THE SCREW WAS REMOVED AND THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITHOUT A SIGNIFICANT DELAY USING A BACK-UP DEVICE IN THE SAME BONE HOLE. NO OTHER COMPLICATIONS WERE REPORTED. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED ACCORDINGLY.
Patient problems: Unspecified Tissue Injury
MAUDE report 11408455 ↗MalfunctionNov 19, 2020Break; Material Separation; Break; Material Separation
IT WAS REPORTED THAT THE SCREW WAS BROKEN. IT IS UNKNOWN WHETHER THE EVENT HAPPENED DURING SURGERY AND IF THERE WAS A PATIENT INVOLVEMENT. A BACK-UP DEVICE WAS AVAILABLE AND NOT SIGNIFICANT DELAY WAS REPORTED. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED ACCORDINGLY.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 10871218 ↗InjuryOct 26, 2020Material Separation; Material Separation
IT WAS REPORTED THAT, DURING AN ACL RECONSTRUCTION PROCEDURE, WHILE FIXATION THE "BIORCI SCREW-HA (10X25MM)" IN THE TIBIAL TUNNEL, THE TIP OF THE SCREW BROKE. THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITHOUT SIGNIFICANT DELAY USING A BACK-UP DEVICE. NO ADDITIONAL COMPLICATIONS WERE REPORTED. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED ACCORDINGLY.
Patient problems: No Information
MAUDE report 10738520 ↗MalfunctionMar 28, 2020Material Separation; Material Separation
IT WAS REPORTED THAT DURING AN ACL SURGERY, THE BIORCI SCREW FAILED. THE SCREW WAS NOT FORWARDING AND WHEN TRIED, THE TIP OF THE SCREW BROKE. ALL PIECES WERE REMOVED FROM THE PATIENT. THE PROCEDURE WAS COMPLETED WITHOUT DELAY USING A BACK-UP DEVICE IN THE SAME BONE HOLE. NO PATIENT INJURY OR OTHER COMPLICATIONS WERE REPORTED. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED ACCORDINGLY.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 9894759 ↗MalfunctionMay 21, 2019Break; Activation, Positioning or Separation Problem
IT WAS REPORTED THAT DURING SURGERY, SCREW BROKE WHILE DEPLOYING. THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITH NO SIGNIFICANT DELAY USING A BACK-UP DEVICE. NO PATIENT INJURY OR OTHER COMPLICATIONS WERE REPORTED.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8628895 ↗MalfunctionMar 27, 2019Break
IT WAS REPORTED THAT DURING AN ACL SURGERY THE TIP OF THE SCREW BROKE. THE PROCEDURE WAS COMPLETED WITH A BACKUP DEVICE WITH NO DELAY OR PATIENT INJURY REPORTED. ALL THE PIECES WERE REMOVED, THE BACK UP DEVICE WAS USED IN THE SAME BONE HOLE.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8458297 ↗MalfunctionDec 22, 2018Break
IT WAS REPORTED THAT THE HE BIORCI HA SCREW GOT DAMAGED DURING THE SURGERY. NO PATIENT INJURY REPORTED. BACK-UP DEVICE WAS AVAILABLE.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8192893 ↗
Other Screw, fixation, bone devices
- NA 03.043.029Synthes GmbH · Screw, fixation, bone503 reports
- OSTEORAPTOR 72201995Smith & Nephew, Inc. · Screw, fixation, bone253 reports
- Control 360 SK12Treace Medical Concepts, Inc. · Screw, Fixation, Bone195 reports
- HEALIX PEEK ORTHOCORD 222205Medos International Sàrl · Screw, fixation, bone186 reports
- PRECICE® Bone Transport System STP3.5-375Nuvasive Specialized Orthopedics, Inc. · Screw, Fixation, Bone3 recalls0 reports
- PRECICE® Bone Transport System STP3.5-225Nuvasive Specialized Orthopedics, Inc. · Screw, Fixation, Bone3 recalls0 reports
- PRECICE® Bone Transport System STP3.5-200Nuvasive Specialized Orthopedics, Inc. · Screw, Fixation, Bone3 recalls0 reports
- PRECICE® Bone Transport System STP3.5-300Nuvasive Specialized Orthopedics, Inc. · Screw, Fixation, Bone3 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 13, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 13, 2026
- Page last rebuilt: Oct 13, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.