TRIGEN 71631161Recalls & reported problems

Smith & Nephew, Inc. · MULTIPURPOSE DRIVER · Class 1

All models of TRIGEN (381)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
13
Deaths reported
0
Injuries reported
4
Recalls naming it
0

Summary

The FDA's MAUDE database contains 13 adverse event reports that list this exact device (DI 03596010478450), received between Jul 16, 2020 and Sep 6, 2024 — classified as 4 injuries, 9 malfunctions.

The most frequently reported specific device problem is “Break” (6 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 1 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break6
  • Crack2
  • Fracture2
  • Connection Problem1
  • Material Deformation1
  • Material Separation1
  • Mechanical Jam1

Patient problems

  • No Clinical Signs, Symptoms or Conditions12
  • No Information1

Recent adverse event reports

The 13 most recent of 13 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 6, 2024Fracture; Fracture

    IT WAS REPORTED THAT DURING A TIBIA INTRAMEDULLARY NAIL PROCEDURE, THE MULTIPURPOSE DRIVER BROKE OUTSIDE THE PATIENT WHEN USING IT TO REMOVE THE TIBIA NAIL. THE DEVICE GOT STUCK IN THE SOFT TISSUE PROTECTOR. SURGERY WAS PERFORMED AFTER A DELAY GREATER THAN 30 MIN, WITH THE SAME DEVICE. NO PATIENT COMPLICATIONS WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20156444 ↗
  • MalfunctionMay 28, 2023Crack; Crack

    IT WAS REPORTED THAT, DURING SURGERY, WHILE SURGEON TRIED TO REMOVE SCREW, THE END OF ONE (1) MULTIPURPOSE DRIVER CRACKED RENDERING IT UNABLE TO REMOVE SCREWS. NO PIECES FELL INSIDE OF THE PATIENT. DOCTOR HAD TO USE A MIDAS REX TO BURR THE SCREW HEADS DOWN SINCE THEY WERE CAUSING THE PATIENT PAIN. THE PROCEDURE WAS RESUMED, AFTER A SIGNIFICANT DELAY, CHANGING THE SURGICAL TECHNIQUE AND WITH A COMPETITOR DEVICE. PATIENT WAS NOT HARMED AS CONSEQUENCE OF THIS PROBLEM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17016088 ↗
  • MalfunctionMay 23, 2023Fracture; Fracture

    IT WAS REPORTED THAT, DURING AN INTERNAL FIXATION SURGERY, THE END OF A MULTIPURPOSE DRIVER BROKE WHILE TRYING TO REMOVE A SCREW FROM AN INTERTAN NAIL THAT WAS CAUSING THE PATIENT PAIN. A MIDAS-REX BURR WAS USED TO CUT SCREWS FROM THE PATIENT'S BONE SO THAT THEY WERE NO LONGER PROTRUDING. PATIENT'S CURRENT HEALTH STATUS IS UNKNOWN.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16986740 ↗
  • MalfunctionFeb 14, 2022Break; Break

    IT WAS REPORTED THAT THE HEX END OF A MULTIPURPOSE DRIVER IS BROKEN. AS THIS WAS NOTICED UPON FIELD INSPECTION, THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13529613 ↗
  • InjuryOct 6, 2021Break; Break

    IT WAS REPORTED THAT, DURING INTERNAL FIXATION SURGERY, A MULTIPURPOSE DRIVER BROKE WHILE TRYING TO REMOVE SCREWS FROM TRIGEN NAIL. SURGERY WAS RESUMED, AFTER A NON-SIGNIFICANT DELAY, WITH A BACK-UP DEVICE. PATIENT WAS NOT INJURED AS CONSEQUENCE OF THIS PROBLEM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12583032 ↗
  • InjurySep 14, 2021Break; Break

    IT WAS REPORTED DURING SURGERY THAT THE TIP OF A MULTIPURPOSE DRIVER BROKE DURING RECON SCREW REMOVAL. PROCEDURE WAS COMPLETED WITH A COMPETITOR DEVICE, A DELAY OF 30 OR LESS MINUTES WAS REPORTED. NO PATIENT HARM REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12471634 ↗
  • InjurySep 14, 2021Break; Break

    IT WAS REPORTED THAT, DURING TRAUMA SURGERY, THE TIP OF A TRIGEN MULTIPURPOSE DRIVER BROKE OFF WHILE PERFORMING A RECON SCREW REMOVAL. SURGERY WAS RESUMED, AFTER A NON-SIGNIFICANT DELAY, WITH A COMPETITOR'S BACK-UP DEVICE. PATIENT WAS NOT INJURED AS CONSEQUENCE OF THIS PROBLEM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12471571 ↗
  • MalfunctionAug 6, 2021Break; Break

    IT WAS REPORTED THAT THE SHAFT OF A TRIGEN MULTIPURPOSE DRIVER WAS BROKEN. AS THIS WAS NOTICED UPON FIELD INSPECTION, THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12291828 ↗
  • MalfunctionJun 18, 2021Material Deformation; Mechanical Jam; Material Deformation; Mechanical Jam

    IT WAS REPORTED THAT, DURING A TRAUMA PROCEDURE, THE META-TAN COMP SCREW DRIVER (COVER UNDER (B)(4)) AND THE MULTIPURPOSE DRIVER RETAINING RODS WERE BOTH BENT TOGETHER AT THE TOP ENDS, AND DID NOT ALLOWED THE SCREW DRIVER RELEASE THE HANDLE TO GO INSIDE THEM. NO SURGICAL DELAY WAS REPORTED, AND THE PROCEDURE WAS FINISHED USING A SMITH AND NEPHEW BACK-UP DEVICE. NO PATIENT INJURY OR OTHER COMPLICATIONS WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12022907 ↗
  • MalfunctionApr 21, 2021Crack; Crack

    IT WAS REPORTED THAT, DURING A TRAUMA SURGERY, THE DISTAL END OF MULTIPURPOSE DRIVER CRACKED WHILE INSIDE OF THE PATIENT. A SMITH AND NEPHEW BACK UP DEVICE WAS USED. IT IS UNKNOWN IF SURGERY WAS DELAYED. PATIENT WAS NOT HARMED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11699801 ↗
  • MalfunctionMar 28, 2021Connection Problem; Connection Problem

    IT WAS REPORTED THAT DURING SURGERY, THE MULTIPURPOSE DRIVER DID NOT ATTACH TO THE T-HANDLE. THERE WAS NO SURGICAL DELAY AND NO INJURY TO THE PATIENT. IT IS UNKNOWN HOW THE PROCEDURE WAS FINISHED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11580428 ↗
  • MalfunctionMar 15, 2021Break; Break

    IT WAS REPORTED THAT, UPON INSPECTION, THE HEX END OF THE MULTIPURPOSE DRIVER WAS SEEN TO BE BROKEN. NO CASE OR PATIENT INVOLVED. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED ACCORDINGLY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11479225 ↗
  • InjuryJul 16, 2020Material Separation; Material Separation

    IT WAS REPORTED THAT, DURING A NAIL REMOTION, THE MULTIPURPOSE DRIVER BROKE WHILE TRYING TO TAKE OUT THE NAIL SCREWS. THE PROCEDURE WAS FINISHED USING A SMITH AND NEPHEW BACK UP DEVICE. SURGERY WAS NOT DELAYED. THE PATIENT WAS NOT HARMED BEYOND THE DESCRIBED EVENT.

    Patient problems: No Information

    MAUDE report 10282977 ↗