TANDEM 71367900Recalls & reported problems
Smith & Nephew, Inc. · TANDEM UNIPOLAR CONVERSION TRIAL 0MM · Class 1
Data updated Oct 13, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 9
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 9 adverse event reports that list this exact device (DI 03596010481733), received between Jul 18, 2020 and Feb 3, 2022 — classified as 1 injury, 8 malfunctions.
The most frequently reported specific device problem is “Break” (4 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Most reported problems
Device problems
- Break4
- Connection Problem1
- Crack1
- Device Damaged Prior to Use1
- Flaked1
- Incomplete or Inadequate Connection1
- Naturally Worn1
Patient problems
- No Clinical Signs, Symptoms or Conditions8
- No Known Impact Or Consequence To Patient1
Recent adverse event reports
The 9 most recent of 9 reports, in the reporter's own words as published by the FDA.
MalfunctionFeb 3, 2022Break; Break
IT WAS REPORTED THAT, DURING THR SURGERY, A TANDEM UNI CONV TRL 0MM BROKE. NO PIECES FELL INTO PATIENT. DEVICE OUTSIDE THE PATIENT. THE PROCEDURE WAS COMPLETED WITHOUT DELAY USING A S+N BACK-UP DEVICE. PATIENT WAS NOT HARMED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13446701 ↗MalfunctionDec 28, 2021Break; Break
IT WAS REPORTED THAT, DURING A THR SURGERY, THE INSIDE SNAP RING OF A TANDEM UNI CONV TRL 0MM WAS BROKEN AND WOULD NOT STAY ON THE TRIAL NECK. INCIDENT OCCURRED OUTSIDE THE PATIENT. SURGERY WAS RESUMED, WITH A DELAY OF LESS THAN OR EQUAL TO 30 MINUTES, WITH THE SAME DEVICE. PATIENT WAS NOT INJURED AS CONSEQUENCE OF THIS PROBLEM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13101513 ↗MalfunctionNov 22, 2021Break; Break
IT WAS REPORTED THAT, DURING A THR SURGERY, THE INSIDE SNAP RING OF A TANDEM UNI CONV TRL 0MM WAS BROKEN AND WOULD NOT STAY ON THE TRIAL NECK. INCIDENT OCCURRED OUTSIDE THE PATIENT. SURGERY WAS RESUMED, WITH A DELAY OF LESS THAN OR EQUAL TO 30 MINUTES, WITH THE SAME DEVICE. PATIENT WAS INJURED AS CONSEQUENCE OF THIS PROBLEM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12852041 ↗MalfunctionNov 1, 2021Break; Break
IT WAS REPORTED THAT THR SURGERY A WHITE RING INSIDE OF TANDEM UNI CONV TRL 0MM WAS NOTICED BROKEN AND MISSING, THE PIECE COULDN'T BE FOUND. THE PATIENT WAS NOT HARMED AS A CONSEQUENCE OF THIS PROBLEM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12727631 ↗MalfunctionJul 7, 2021Crack; Naturally Worn; Crack; Naturally Worn
IT WAS REPORTED THAT A TANDEM UNIPOLAR CONVERSION TRIAL 0MM WAS OBSERVED TO BE WORN AND CRACKED. AS THIS WAS NOTICED UPON INSPECTION, THERE WAS NOT PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12132797 ↗MalfunctionMay 19, 2021Incomplete or Inadequate Connection; Incomplete or Inadequate Connection
IT WAS REPORTED THAT THE TANDEM UNI CONV TRL 0 MM WAS NOT LONGER ENGAGING TAPERS. IT IS UNKNOWN IF THIS WAS NOTICED DURING SURGERY OR NOT; THEREFORE, PATIENT INVOLVEMENT IS NOT CONFIRMED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11850236 ↗MalfunctionJan 6, 2021Device Damaged Prior to Use; Device Damaged Prior to Use
IT WAS REPORTED BEFORE THE THR PROCEDURE THAT THE TRIALS WERE FLAKING AND IMPACTED. THE PROCEDURE WAS COMPLETED WITHOUT DELAY, USING A BACKUP FROM SMITH & NEPHEW. NO PATIENT INJURY OR OTHER COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11124100 ↗MalfunctionJan 5, 2021Flaked; Flaked
IT WAS REPORTED THAT, DURING SET UP OR INSPECTION BEFORE A TKA IT WAS NOTICED THAT THE TANDEM UNI CONV TRIAL 0MM, +45MM, +47MM, +53MM AND +57MM WERE FLAKING AND IMPACTING STERILIZATION PROCESS. THE PROCEDURE WAS COMPLETED WITHOUT DELAY USING A SMITH-NEPHEW BACK-UP DEVICE. NO PATIENT INJURY OR OTHER COMPLICATIONS WERE REPORTED. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED ACCORDINGLY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11119824 ↗InjuryJul 18, 2020Connection Problem; Connection Problem
IT WAS REPORTED THAT DURING TOTAL HIP REPLACEMENT THE UNIPOLAR TRIAL HEADS DO NOT STAY ON THE STEM OR BROACH TRIALS. IT WAS TRIED ALL 5 DEVICES AND THE POP OFF NON STOP. A DELAY WAS REPORTED OF 10 MINUTES TO 15 MINUTES. THE PROCEDURE WAS FINISHED USING A CHANGE IN THE SURGICAL TECHNIQUE. A DIFFERENT DEVICE HAS TO BE USED TO FINISHED THE PROCEDURE. THE PROCEDURE WAS FINISHED USING 28 MM BIPOLAR HEAD.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 10293443 ↗
Other ORTHOPEDIC MANUAL SURGICAL INSTRUMENT devices
- AML EXCEL 2001-42-000Depuy (Ireland) · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT2,009 reports
- NA 03.010.523Synthes GmbH · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT1,208 reports
- EXPRESSEW 214141Medos International Sàrl · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT1,166 reports
- N/A 9560524Medtronic Sofamor Danek, Inc. · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT866 reports
- INHANCE 6500-11-104Depuy (Ireland) · Orthopedic manual surgical instrument831 reports
- INHANCE 6500-11-102Depuy (Ireland) · Orthopedic manual surgical instrument567 reports
- R3 71364450Smith & Nephew, Inc. · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT520 reports
- N/A 9561524Medtronic Sofamor Danek, Inc. · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT494 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 13, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 13, 2026
- Page last rebuilt: Oct 13, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page