GENESIS II 71430435Recalls & reported problems

Smith & Nephew, Inc. · GENESIS II OVAL RESURFACING PATELLAR TRIAL 35MM · Class 1

All models of GENESIS II (462)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
4
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 03596010484932), received between Jul 6, 2020 and Aug 27, 2024 — classified as 2 injuries, 2 malfunctions.

The most frequently reported specific device problem is “Break” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 1 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break2
  • Crack1
  • Fracture1
  • Scratched Material1

Patient problems

  • No Clinical Signs, Symptoms or Conditions3
  • No Known Impact Or Consequence To Patient1

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • MalfunctionAug 27, 2024Fracture; Fracture

    IT WAS REPORTED THAT, DURING A TKA SURGERY, THE BUTTON OF A GII OVAL RESURF PAT TRIAL 35MM BROKE. THE PROCEDURE WAS COMPLETED, WITHOUT ANY DELAY, USING THE SAME DEVICE. PATIENT WAS NOT HARMED AS CONSEQUENCE OF THIS PROBLEM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20075000 ↗
  • InjuryOct 28, 2021Break; Break

    IT WAS REPORTED, THAT DURING A TKA A GENESIS II OVAL RESURF PAT TRIAL 35MM BROKE INSIDE THE PATIENT. ALL PIECES WERE RETRIEVED. SURGERY WAS COMPLETED, WITHOUT DELAY, BY USING A BACK-UP DEVICE. THE PATIENT WAS NOT HARMED BEYOND WHAT HAS BEEN DESCRIBED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12715831 ↗
  • MalfunctionJul 13, 2021Break; Scratched Material; Break; Scratched Material

    IT WAS REPORTED THAT, THREE (3) PATELLA TRIALS GII OVAL RESURF PAT TRIAL 35MM, GII OVAL RESURF PAT TRIAL 29MM AND GII OVAL RESURF PAT TRIAL 32MM, HAVE DEEP SCRATCHES AND SMALL CHUNKS MISSING. AS THIS WAS NOTICED UPON FIELD INSPECTION, THERE WAS NOT PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12164304 ↗
  • InjuryJul 6, 2020Crack; Crack

    IT WAS REPORTED THAT DURING TKA PROCEDURE THE PATELLA TRIAL MADE CONTACT WITH A NAVIO BURR. DEVICE GOT CRACKED/CHIPPED. THE DEVICE WAS BEING USED INSIDE THE PATIENT BUT NO PIECES FELL INSIDE. DEVICE DID NOT BREAK. NO DELAY TO PROCEDURE. SURGERY WAS FINISHED USING THE SAME DEVICE. NO INJURIES REPORTED. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED ACCORDINGLY.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 10237518 ↗