Skip to content

TANDEM 71322046Recalls & reported problems

Smith & Nephew, Inc. · TANDEM BIPOLAR COBALT CHROME 46MM OUTER DIAMETER 28MM INNER DIAMETER · Class 2 · Implant

All models of TANDEM (58)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
8
Deaths reported
0
Injuries reported
8
Recalls naming it
0

Summary

The FDA's MAUDE database contains 8 adverse event reports that list this exact device (DI 03596010492456), received between May 17, 2017 and Aug 12, 2025 — classified as 8 injuries.

The most frequently reported specific device problem is “Device Dislodged or Dislocated” (4 reports).

3 other recalls cover devices cleared under the same FDA submission (K023743); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 1 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Device Dislodged or Dislocated4
  • Adverse Event Without Identified Device or Use Problem3
  • Defective Component1
  • Detachment Of Device Component1
  • Material Erosion1

Patient problems

  • Joint Dislocation4
  • Fall2
  • Injury2
  • Pain2
  • Failure of Implant1
  • Unspecified Infection1

Recent adverse event reports

The 8 most recent of 8 reports, in the reporter's own words as published by the FDA.

  • InjuryAug 12, 2025Device Dislodged or Dislocated; Device Dislodged or Dislocated

    IT WAS REPORTED THAT, AFTER THE SURGERY HAD BEEN PERFORMED ON (B)(6) 2025, THE PATIENT EXPERIENCED DISLOCATION AFTER FALLING FROM BED AND A CLOSED REDUCTION WAS ATTEMPTED. THIS ADVERSE EVENT WAS TREATED BY A REVISION SURGERY ON (B)(6) 2025, IN WHICH THE TANDEM BIPOLAR COCR 46OD 28ID WAS REMOVED AND IT WAS NOTICED THAT THE 28MM HEAD WAS STILL ATTACHED TO THE STEM AND THE TANDEM BIPOLAR LOCKING RING WAS INTACT. THE CURRENT HEALTH STATUS OF THE PATIENT IS UNKNOWN. NO FURTHER COMPLICATIONS WERE REPORTED.

    Patient problems: Fall; Joint Dislocation

    MAUDE report 22784853 ↗
  • InjurySep 25, 2023Defective Component; Adverse Event Without Identified Device or Use Problem; Defective Component; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT, AFTER A THR SURGERY HAD BEEN PERFORMED ON 28-AUG-2023, THE PATIENT EXPERIENCED A DISLOCATION, AS THE ENTIRE CONSTRUCT DISLOCATED FROM THE ACETABULUM. WHEN THIS WAS BEING REDUCED IN THE ER, ON (B)(6) 2023, THE INNER FEMORAL HEAD POPPED OUT OF THE BIPOLAR COMPONENT. THE CONSTRUCT WAS REPLACED WITH A NEW 46 OD COMPONENT AND A 28 +8 FEMORAL HEAD. CURRENT HEALTH STATUS OF PATIENT IS UNKNOWN.

    Patient problems: Joint Dislocation

    MAUDE report 17814028 ↗
  • InjuryDec 26, 2022Material Erosion; Device Dislodged or Dislocated; Material Erosion; Device Dislodged or Dislocated

    IT WAS REPORTED THAT, AFTER A THR SURGERY, HAD BEEN PERFORMED ON (B)(6) 2022, THE PATIENT SUSTAINED A FALL AND THE COCR 12/14 FEM HEAD 28 + 4 OF THE TANDEM BIPOLAR BECAME DISASSOCIATED. THE SHELL WITH THE LOCKING RING REMAINED INTACT. HOWEVER THERE WAS WEAR AROUND THE LOCKING RING OF THE TANDEM BIPOLAR COCR 46OD 28ID WHERE THE BALL CAME OUT. THE LOCKING RING WAS REMOVED AND THE INNER BALL REPLACED ON (B)(6) 2022. PATIENT CURRENT HEALTH STATUS IS UNKNOWN.

    Patient problems: Failure of Implant; Fall; Joint Dislocation

    MAUDE report 16048280 ↗
  • InjuryDec 8, 2020Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT, AFTER A HIP HEMIARTHROPLASTY, PATIENT DEVELOPED AN INFECTION. A REVISION SURGERY WAS PERFORMED TO EXPLANT THE BIPOLAR LINER THE PATIENT OUTCOME IS UNKNOWN.

    Patient problems: Unspecified Infection

    MAUDE report 10968583 ↗
  • InjurySep 18, 2020Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT REVISION WAS PERFORMED DUE TO PAINFUL RIGHT HIP BIPOLAR ARTHROPLASTY WITH EROSIVE CHANGES OF THE ACETABULUM. TAPER FEMORAL HEAD, SHELL/ UHMWPE LINER AND FEMORAL COMPONENT WERE EXCHANGED AND PROCEDURE WAS CONVERTED TO A TOTAL HIP REPLACEMENT. STEM WAS RETAINED.

    Patient problems: Injury; Pain

    MAUDE report 10551840 ↗
  • InjuryApr 28, 2019Device Dislodged or Dislocated; Device Dislodged or Dislocated

    IT WAS REPORTED THAT PATIENT HAD A DISLOCATED BIPOLAR HIP, ON (B)(6) 2019 SURGEON TRIED TO REDUCE THE HIP AND HEAD DISSOCIATED FROM THE TANDEM BIPOLAR SHELL. ON (B)(6) 2019 A REVISION SURGERY WAS PERFORMED, IMPLANTS WERE REMOVED AND REPLACED.

    Patient problems: Injury

    MAUDE report 8559682 ↗
  • InjuryJun 26, 2017Detachment Of Device Component; Detachment Of Device Component

    IT WAS REPORTED A REVISION SURGERY WAS REQUIRED DUE TO HIP PAIN. IT WAS DISCOVERED THE POLY LINER AND FEMORAL HEAD HAD DISASSOCIATED.

    Patient problems: Pain

    MAUDE report 6668513 ↗
  • InjuryMay 17, 2017Device Dislodged or Dislocated; Device Dislodged or Dislocated

    IT WAS REPORTED THE PATIENT PRESENTED WITH A DISLOCATED BIPOLAR SHELL. THE SURGEON ATTEMPTED TO REDUCE AND THE SHELL DISLOCATED FROM THE 28MM HEAD. A REVISION SURGERY WAS PERFORMED TO REMOVE THE CONSTRUCT.

    Patient problems: Joint Dislocation

    MAUDE report 6571873 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K023743. One submission often covers many models, so check the product and lot details in each record.

  • Z-0326-2025 · Oct 15, 2024 · Open, Classified
    Tandem Intl Bipolar Cocr Shell/UHMWPE Liner, TI-6AL-4V Retaining Ring, 28MM I.D. 46 MM O.D., REF 71324046, sterile, femoral head bipolar component of the Tandem Bipolar Hip System.
  • Z-0351-2017 · Aug 29, 2016 · Terminated
    TANDEM (TM) INTL COCR SHELL UHMWPE LINER BIPOLAR, REF NUMBERS: 71324038, 71324039, 71324040, 71324041, 71324042, 71324043, 71324044, 71324045, 71324046, 71324047, 71324048, 7132
  • Z-0350-2017 · Aug 29, 2016 · Terminated
    TANDEM (TM) COCR SHELL UHMWPE LINER BIPOLAR, REF NUMBERS: 71322040, 71322041, 71322042, 71322043, 71322044, 71322045, 71322046, 71322047, 71322048, 71322049, 71322050, 71322051,

Other Prosthesis, hip, hemi-, femoral, metal/polymer, cemented or uncemented devices

All Prosthesis, hip, hemi-, femoral, metal/polymer, cemented or uncemented devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page