TRUCLEARTM Control Unit 7209808Recalls & reported problems
Smith & Nephew, Inc. · TRUCLEAR CONTROL UNIT · Class 2
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 3
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 03596010501929), received between Feb 10, 2022 and Jul 24, 2023 — classified as 3 injuries.
The most frequently reported specific device problem is “Intermittent Energy Output” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Most reported problems
Device problems
- Appropriate Device Problem Term/Code Not Available2
- Intermittent Energy Output2
- Adverse Event Without Identified Device or Use Problem1
- Connection Problem1
- Fail-Safe Problem1
- Mechanical Jam1
Patient problems
- No Clinical Signs, Symptoms or Conditions2
- Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available1
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
InjuryJul 24, 2023Appropriate Device Problem Term/Code Not Available; Intermittent Energy Output; Appropriate Device Problem Term/Code Not Available; Intermittent Energy Output
ACCORDING TO THE REPORTER, DURING A HYSTEROSCOPY MYOMECTOMY, THE HANDPIECE AND BLADE JUST STOPPED WORKING AND CONTROL UNIT SAID "STA LLED MOTOR, TRY NEW HANDPIECE". THE SURGICAL TIME WAS EXTENDED 30 MINUTES OR MORE DUE TO THE PRODUCT PROBLEM. HAND TO SWITCH OUT HANDPIECES AND BLADES AND CASE WAS ABORTED AFTER TWO BLADES AND TWO HANDPIECES WERE ATTEMPTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17386366 ↗InjuryApr 13, 2023Connection Problem; Appropriate Device Problem Term/Code Not Available; Connection Problem; Appropriate Device Problem Term/Code Not Available
ACCORDING TO THE REPORTER, DURING A HYSTEROSCOPY, THE HANDPIECE PORT ON THE CONTROL UNIT WAS NOT ACCEPTING THE HANDPIECE PLUG. THE HANDPIECE WAS UNABLE TO BE PLUGGED IN TO PERFORM THE PROCEDURE. CASE WAS ABORTED, RESCHEDULE AND THERE BE AN ADDITIONAL OPERATION TO BE PERFORMED TO CORRECT THE ISSUE. THE PATIENT WAS UNDER ANESTHESIA BEFORE DISCOVERING THAT THE CONTROL UNIT WAS DEFECTIVE. ANOTHER CONTROL UNIT WAS TESTED AND IT WORKED FINE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16737314 ↗InjuryFeb 10, 2022Fail-Safe Problem; Mechanical Jam; Adverse Event Without Identified Device or Use Problem; Intermittent Energy Output; Fail-Safe Problem; Mechanical Jam; Adverse Event Without Identified Device or Use Problem; Intermittent Energy Output
ACCORDING TO THE REPORTER, DURING A PROCEDURE, THE BLADE OF THE DEVICE WOULD NOT MOVE WHEN WINDOW LOCK PRESSED ON FOOTSWITCH. WINDOW LOCK DID NOT FUNCTION AND TEST BLADE RAN AT EXCEEDINGLY FAST SPEEDS. THE PATIENT WAS UNABLE TO BE TREATED AND WAS RESCHEDULED.
Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available
MAUDE report 13510454 ↗
Other HYSTEROSCOPE (AND ACCESSORIES) devices
- Olympus WA2T430AOLYMPUS Winter & Ibe GmbH · Hysteroscope (And Accessories)783 reports
- Olympus WA2T412AOLYMPUS Winter & Ibe GmbH · Hysteroscope (And Accessories)453 reports
- Olympus A22001AOLYMPUS Winter & Ibe GmbH · Hysteroscope (And Accessories)366 reports
- Olympus WA2T470AOLYMPUS Winter & Ibe GmbH · Hysteroscope (And Accessories)323 reports
- Olympus A22002AOLYMPUS Winter & Ibe GmbH · Hysteroscope (And Accessories)297 reports
- Olympus A22003AOLYMPUS Winter & Ibe GmbH · Hysteroscope (And Accessories)198 reports
- n.a. 27040XAKarl Storz GmbH & Co. KG · HYSTEROSCOPE (AND ACCESSORIES)102 reports
- Myosure Reach Tissue Removal Device 10-401FCHologic, Inc. · HYSTEROSCOPE (AND ACCESSORIES)4 deaths99 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.