TRUCLEARTM Handpiece 7209807Recalls & reported problems

Smith & Nephew, Inc. · TRUCLEAR HANDPIECE · Class 2

Adverse event reports
9
Deaths reported
0
Injuries reported
7
Recalls naming it
0

Summary

The FDA's MAUDE database contains 9 adverse event reports that list this exact device (DI 03596010501943), received between Jul 20, 2020 and Jun 19, 2024 — classified as 7 injuries, 2 malfunctions.

The most frequently reported specific device problem is “Fail-Safe Problem” (6 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 1 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem9
  • Fail-Safe Problem6
  • Mechanical Jam4
  • Intermittent Energy Output3
  • Break2
  • Insufficient Device Problem Information2

Patient problems

  • Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available6
  • No Clinical Signs, Symptoms or Conditions2
  • No Code Available2

Recent adverse event reports

The 9 most recent of 9 reports, in the reporter's own words as published by the FDA.

  • InjuryJun 19, 2024Break; Adverse Event Without Identified Device or Use Problem; Break; Adverse Event Without Identified Device or Use Problem

    ACCORDING TO THE REPORTER, DURING A POLYPECTOMY PROCEDURE, THE NURSE TRIED TO ATTACHED THE FIRST HANDPIECE TO THE CONTROL UNIT, BUT IT WOULD NOT ATTACH. TRIED WITH ANOTHER ONE BUT IT ALSO WOULD NOT ATTACH. THE OUTER RING FROM THE SHAFT OF THE HANDPIECES HAS COME OFF AND ONE OF THESE WAS STILL IN THE CONTROL UNIT. THERE WAS NO PATIENT INJURY, HOWEVER, PROCEDURE WAS NOT COMPLETED AND HAD TO BE ABANDONED. AND AN ADDITIONAL OPERATION WAS PERFORMED TO CORRECT THE ISSUE.

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available; No Clinical Signs, Symptoms or Conditions

    MAUDE report 19571027 ↗
  • InjuryFeb 22, 2022Fail-Safe Problem; Mechanical Jam; Adverse Event Without Identified Device or Use Problem; Insufficient Device Problem Information; Fail-Safe Problem; Mechanical Jam; Adverse Event Without Identified Device or Use Problem; Insufficient Device Problem Information

    ACCORDING TO THE REPORTER, DURING AN OPERATIVE HYSTEROSCOPY, HYSTEROSCOPIC MYOMECTOMY PROCEDURE, THE BLADE OF THE DEVICE WOULD NOT MOVE WHEN WINDOW LOCK PRESSED ON FOOTSWITCH. WINDOW LOCK DID NOT FUNCTION AND TEST BLADE RAN AT EXCEEDINGLY FAST SPEEDS. THE PATIENT WAS UNABLE TO BE TREATED AND WAS RESCHEDULED.

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available

    MAUDE report 13578466 ↗
  • InjuryFeb 10, 2022Fail-Safe Problem; Adverse Event Without Identified Device or Use Problem; Intermittent Energy Output; Fail-Safe Problem; Adverse Event Without Identified Device or Use Problem; Intermittent Energy Output

    ACCORDING TO THE REPORTER, DURING A PROCEDURE INSIDE THE UTERINE CAVITY PRIOR TO MYOMECTOMY, THEY WENT TO PERFORM WINDOW LOCK, BUT DID NOT FUNCTION. THE DEVICE WAS THEN REMOVED FROM THE UTERINE CAVITY AND FURTHER COMPLICATIONS IN RELATION TO DEVICES NOT WORKING WERE EXPERIENCED. THEY HAVE TESTED ONE HANDPIECE AND IT WAS RUNNING EXCEEDINGLY FAST, SO FAST THAT IT SEIZES THE TEST BLADE. IT RAN AT 3500 RPM THEN STOPPED. THEY'VE TRIED A SECOND HANDPIECE AND AGAIN IT WAS RUNNING AT SPEEDS IN EXCESS OF WHAT IS DISPLAYED ON THE CONTROL UNIT, ABOUT 1000 RPM OVER. THEY CHANGED HANDPIECE AND BLADE AND EXPERIENCED THE SAME ISSUE. THEY WERE USING A DENSE TISSUE SHAVER PLUS. IT WAS NOTED THAT THE DENSE TISSUE SHAVER PLUS DEVICE JAMMED ON ALL HANDPIECES WHEN THE DEVICE HAD BEEN RUNNING FOR OVER 1MIN AT 2500 RPM. THE CASE WAS CANCELLED. THEY WERE UNABLE TO TREAT THE PATIENT AND WAS RESCHEDULED.

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available

    MAUDE report 13510484 ↗
  • InjuryFeb 10, 2022Fail-Safe Problem; Adverse Event Without Identified Device or Use Problem; Intermittent Energy Output; Fail-Safe Problem; Adverse Event Without Identified Device or Use Problem; Intermittent Energy Output

    ACCORDING TO THE REPORTER, DURING A PROCEDURE INSIDE THE UTERINE CAVITY PRIOR TO MYOMECTOMY, THEY WENT TO PERFORM WINDOW LOCK, BUT DID NOT FUNCTION. THE DEVICE WAS THEN REMOVED FROM THE UTERINE CAVITY AND FURTHER COMPLICATIONS IN RELATION TO DEVICES NOT WORKING WERE EXPERIENCED. THEY HAVE TESTED ONE HANDPIECE AND IT WAS RUNNING EXCEEDINGLY FAST, SO FAST THAT IT SEIZES THE TEST BLADE. IT RAN AT 3500 RPM THEN STOPPED. THEY'VE TRIED A SECOND HANDPIECE AND AGAIN IT WAS RUNNING AT SPEEDS IN EXCESS OF WHAT IS DISPLAYED ON THE CONTROL UNIT, ABOUT 1000 RPM OVER. THEY CHANGED HANDPIECE AND BLADE AND EXPERIENCED THE SAME ISSUE. THEY WERE USING A DENSE TISSUE SHAVER PLUS. IT WAS NOTED THAT THE DENSE TISSUE SHAVER PLUS DEVICE JAMMED ON ALL HANDPIECES WHEN THE DEVICE HAD BEEN RUNNING FOR OVER 1MIN AT 2500 RPM. THE CASE WAS CANCELLED. THEY WERE UNABLE TO TREAT THE PATIENT AND WAS RESCHEDULED.

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available

    MAUDE report 13510478 ↗
  • InjuryFeb 10, 2022Fail-Safe Problem; Adverse Event Without Identified Device or Use Problem; Intermittent Energy Output; Fail-Safe Problem; Adverse Event Without Identified Device or Use Problem; Intermittent Energy Output

    ACCORDING TO THE REPORTER, DURING A PROCEDURE INSIDE THE UTERINE CAVITY PRIOR TO MYOMECTOMY, THEY WENT TO PERFORM WINDOW LOCK, BUT DID NOT FUNCTION. THE DEVICE WAS THEN REMOVED FROM THE UTERINE CAVITY AND FURTHER COMPLICATIONS IN RELATION TO DEVICES NOT WORKING WERE EXPERIENCED. THEY HAVE TESTED ONE HANDPIECE AND IT WAS RUNNING EXCEEDINGLY FAST, SO FAST THAT IT SEIZES THE TEST BLADE. IT RAN AT 3500 RPM THEN STOPPED. THEY'VE TRIED A SECOND HANDPIECE AND AGAIN IT WAS RUNNING AT SPEEDS IN EXCESS OF WHAT IS DISPLAYED ON THE CONTROL UNIT, ABOUT 1000 RPM OVER. THEY CHANGED HANDPIECE AND BLADE AND EXPERIENCED THE SAME ISSUE. THEY WERE USING A DENSE TISSUE SHAVER PLUS. IT WAS NOTED THAT THE DENSE TISSUE SHAVER PLUS DEVICE JAMMED ON ALL HANDPIECES WHEN THE DEVICE HAD BEEN RUNNING FOR OVER 1MIN AT 2500 RPM. THE CASE WAS CANCELLED. THEY WERE UNABLE TO TREAT THE PATIENT AND WAS RESCHEDULED.

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available

    MAUDE report 13510467 ↗
  • InjuryFeb 10, 2022Fail-Safe Problem; Mechanical Jam; Adverse Event Without Identified Device or Use Problem; Fail-Safe Problem; Mechanical Jam; Adverse Event Without Identified Device or Use Problem

    ACCORDING TO THE REPORTER, DURING A PROCEDURE, THE BLADE OF THE DEVICE WOULD NOT MOVE WHEN THE WINDOW LOCK PRESSED ON FOOTSWITCH. THE WINDOW LOCK DID NOT FUNCTION AND THE TEST BLADE RAN AT EXCEEDINGLY FAST SPEEDS. THE PATIENT WAS UNABLE TO BE TREATED AND WAS RESCHEDULED.

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available

    MAUDE report 13510451 ↗
  • InjuryFeb 8, 2022Mechanical Jam; Adverse Event Without Identified Device or Use Problem; Insufficient Device Problem Information; Mechanical Jam; Adverse Event Without Identified Device or Use Problem; Insufficient Device Problem Information

    ACCORDING TO THE REPORTER, DURING A PROCEDURE, AT THE START OF THE PROCEDURE, WHEN RUNNING A DENSE TISSUE SHAVER PLUS DEVICE ON THE HANDPIECE WITH A SCOPE AND A TISSUE SHAVER, IT JAMMED AND FAILED TO ROTATE. SWITCHED HANDPIECES AND BLADES AND AGAIN EXPERIENCED THE SAME ISSUE. THE PROCEDURE WAS CANCELLED, THEN RESCHEDULED AND COMPLETED USING ANOTHER EQUIPMENT. PATIENT WAS UNDER ANAESTHESIA. THERE WAS NO PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13489177 ↗
  • MalfunctionJul 20, 2020Break; Mechanical Jam; Adverse Event Without Identified Device or Use Problem; Break; Mechanical Jam; Adverse Event Without Identified Device or Use Problem

    ACCORDING TO THE REPORTER, DURING POLYPECTOMY, THE DEVICE HAD A MALFUNCTION THAT LEAD TO THE PROCEDURE BEING ABORTED. THE PATIENT WAS NOT ANESTHETIZED AND THE RE-SCHEDULED PROCEDURE WAS SUCCESSFUL.

    Patient problems: No Code Available

    MAUDE report 10299469 ↗
  • MalfunctionJul 20, 2020Fail-Safe Problem; Adverse Event Without Identified Device or Use Problem; Fail-Safe Problem; Adverse Event Without Identified Device or Use Problem

    ACCORDING TO THE REPORTER, DURING POLYPECTOMY, THE DEVICE HAD A MALFUNCTION THAT LEAD TO THE PROCEDURE BEING ABORTED. THE PATIENT WAS NOT ANESTHETIZED AND THE RE-SCHEDULED PROCEDURE WAS SUCCESSFUL.

    Patient problems: No Code Available

    MAUDE report 10299462 ↗