TRUCLEARTM Handpiece 7209807Recalls & reported problems
Smith & Nephew, Inc. · TRUCLEAR HANDPIECE · Class 2
Data updated Nov 27, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 9
- Deaths reported
- 0
- Injuries reported
- 7
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 9 adverse event reports that list this exact device (DI 03596010501943), received between Jul 20, 2020 and Jun 19, 2024 — classified as 7 injuries, 2 malfunctions.
The most frequently reported specific device problem is “Fail-Safe Problem” (6 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem9
- Fail-Safe Problem6
- Mechanical Jam4
- Intermittent Energy Output3
- Break2
- Insufficient Device Problem Information2
Patient problems
- Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available6
- No Clinical Signs, Symptoms or Conditions2
- No Code Available2
Recent adverse event reports
The 9 most recent of 9 reports, in the reporter's own words as published by the FDA.
InjuryJun 19, 2024Break; Adverse Event Without Identified Device or Use Problem; Break; Adverse Event Without Identified Device or Use Problem
ACCORDING TO THE REPORTER, DURING A POLYPECTOMY PROCEDURE, THE NURSE TRIED TO ATTACHED THE FIRST HANDPIECE TO THE CONTROL UNIT, BUT IT WOULD NOT ATTACH. TRIED WITH ANOTHER ONE BUT IT ALSO WOULD NOT ATTACH. THE OUTER RING FROM THE SHAFT OF THE HANDPIECES HAS COME OFF AND ONE OF THESE WAS STILL IN THE CONTROL UNIT. THERE WAS NO PATIENT INJURY, HOWEVER, PROCEDURE WAS NOT COMPLETED AND HAD TO BE ABANDONED. AND AN ADDITIONAL OPERATION WAS PERFORMED TO CORRECT THE ISSUE.
Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available; No Clinical Signs, Symptoms or Conditions
MAUDE report 19571027 ↗InjuryFeb 22, 2022Fail-Safe Problem; Mechanical Jam; Adverse Event Without Identified Device or Use Problem; Insufficient Device Problem Information; Fail-Safe Problem; Mechanical Jam; Adverse Event Without Identified Device or Use Problem; Insufficient Device Problem Information
ACCORDING TO THE REPORTER, DURING AN OPERATIVE HYSTEROSCOPY, HYSTEROSCOPIC MYOMECTOMY PROCEDURE, THE BLADE OF THE DEVICE WOULD NOT MOVE WHEN WINDOW LOCK PRESSED ON FOOTSWITCH. WINDOW LOCK DID NOT FUNCTION AND TEST BLADE RAN AT EXCEEDINGLY FAST SPEEDS. THE PATIENT WAS UNABLE TO BE TREATED AND WAS RESCHEDULED.
Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available
MAUDE report 13578466 ↗InjuryFeb 10, 2022Fail-Safe Problem; Adverse Event Without Identified Device or Use Problem; Intermittent Energy Output; Fail-Safe Problem; Adverse Event Without Identified Device or Use Problem; Intermittent Energy Output
ACCORDING TO THE REPORTER, DURING A PROCEDURE INSIDE THE UTERINE CAVITY PRIOR TO MYOMECTOMY, THEY WENT TO PERFORM WINDOW LOCK, BUT DID NOT FUNCTION. THE DEVICE WAS THEN REMOVED FROM THE UTERINE CAVITY AND FURTHER COMPLICATIONS IN RELATION TO DEVICES NOT WORKING WERE EXPERIENCED. THEY HAVE TESTED ONE HANDPIECE AND IT WAS RUNNING EXCEEDINGLY FAST, SO FAST THAT IT SEIZES THE TEST BLADE. IT RAN AT 3500 RPM THEN STOPPED. THEY'VE TRIED A SECOND HANDPIECE AND AGAIN IT WAS RUNNING AT SPEEDS IN EXCESS OF WHAT IS DISPLAYED ON THE CONTROL UNIT, ABOUT 1000 RPM OVER. THEY CHANGED HANDPIECE AND BLADE AND EXPERIENCED THE SAME ISSUE. THEY WERE USING A DENSE TISSUE SHAVER PLUS. IT WAS NOTED THAT THE DENSE TISSUE SHAVER PLUS DEVICE JAMMED ON ALL HANDPIECES WHEN THE DEVICE HAD BEEN RUNNING FOR OVER 1MIN AT 2500 RPM. THE CASE WAS CANCELLED. THEY WERE UNABLE TO TREAT THE PATIENT AND WAS RESCHEDULED.
Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available
MAUDE report 13510484 ↗InjuryFeb 10, 2022Fail-Safe Problem; Adverse Event Without Identified Device or Use Problem; Intermittent Energy Output; Fail-Safe Problem; Adverse Event Without Identified Device or Use Problem; Intermittent Energy Output
ACCORDING TO THE REPORTER, DURING A PROCEDURE INSIDE THE UTERINE CAVITY PRIOR TO MYOMECTOMY, THEY WENT TO PERFORM WINDOW LOCK, BUT DID NOT FUNCTION. THE DEVICE WAS THEN REMOVED FROM THE UTERINE CAVITY AND FURTHER COMPLICATIONS IN RELATION TO DEVICES NOT WORKING WERE EXPERIENCED. THEY HAVE TESTED ONE HANDPIECE AND IT WAS RUNNING EXCEEDINGLY FAST, SO FAST THAT IT SEIZES THE TEST BLADE. IT RAN AT 3500 RPM THEN STOPPED. THEY'VE TRIED A SECOND HANDPIECE AND AGAIN IT WAS RUNNING AT SPEEDS IN EXCESS OF WHAT IS DISPLAYED ON THE CONTROL UNIT, ABOUT 1000 RPM OVER. THEY CHANGED HANDPIECE AND BLADE AND EXPERIENCED THE SAME ISSUE. THEY WERE USING A DENSE TISSUE SHAVER PLUS. IT WAS NOTED THAT THE DENSE TISSUE SHAVER PLUS DEVICE JAMMED ON ALL HANDPIECES WHEN THE DEVICE HAD BEEN RUNNING FOR OVER 1MIN AT 2500 RPM. THE CASE WAS CANCELLED. THEY WERE UNABLE TO TREAT THE PATIENT AND WAS RESCHEDULED.
Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available
MAUDE report 13510478 ↗InjuryFeb 10, 2022Fail-Safe Problem; Adverse Event Without Identified Device or Use Problem; Intermittent Energy Output; Fail-Safe Problem; Adverse Event Without Identified Device or Use Problem; Intermittent Energy Output
ACCORDING TO THE REPORTER, DURING A PROCEDURE INSIDE THE UTERINE CAVITY PRIOR TO MYOMECTOMY, THEY WENT TO PERFORM WINDOW LOCK, BUT DID NOT FUNCTION. THE DEVICE WAS THEN REMOVED FROM THE UTERINE CAVITY AND FURTHER COMPLICATIONS IN RELATION TO DEVICES NOT WORKING WERE EXPERIENCED. THEY HAVE TESTED ONE HANDPIECE AND IT WAS RUNNING EXCEEDINGLY FAST, SO FAST THAT IT SEIZES THE TEST BLADE. IT RAN AT 3500 RPM THEN STOPPED. THEY'VE TRIED A SECOND HANDPIECE AND AGAIN IT WAS RUNNING AT SPEEDS IN EXCESS OF WHAT IS DISPLAYED ON THE CONTROL UNIT, ABOUT 1000 RPM OVER. THEY CHANGED HANDPIECE AND BLADE AND EXPERIENCED THE SAME ISSUE. THEY WERE USING A DENSE TISSUE SHAVER PLUS. IT WAS NOTED THAT THE DENSE TISSUE SHAVER PLUS DEVICE JAMMED ON ALL HANDPIECES WHEN THE DEVICE HAD BEEN RUNNING FOR OVER 1MIN AT 2500 RPM. THE CASE WAS CANCELLED. THEY WERE UNABLE TO TREAT THE PATIENT AND WAS RESCHEDULED.
Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available
MAUDE report 13510467 ↗InjuryFeb 10, 2022Fail-Safe Problem; Mechanical Jam; Adverse Event Without Identified Device or Use Problem; Fail-Safe Problem; Mechanical Jam; Adverse Event Without Identified Device or Use Problem
ACCORDING TO THE REPORTER, DURING A PROCEDURE, THE BLADE OF THE DEVICE WOULD NOT MOVE WHEN THE WINDOW LOCK PRESSED ON FOOTSWITCH. THE WINDOW LOCK DID NOT FUNCTION AND THE TEST BLADE RAN AT EXCEEDINGLY FAST SPEEDS. THE PATIENT WAS UNABLE TO BE TREATED AND WAS RESCHEDULED.
Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available
MAUDE report 13510451 ↗InjuryFeb 8, 2022Mechanical Jam; Adverse Event Without Identified Device or Use Problem; Insufficient Device Problem Information; Mechanical Jam; Adverse Event Without Identified Device or Use Problem; Insufficient Device Problem Information
ACCORDING TO THE REPORTER, DURING A PROCEDURE, AT THE START OF THE PROCEDURE, WHEN RUNNING A DENSE TISSUE SHAVER PLUS DEVICE ON THE HANDPIECE WITH A SCOPE AND A TISSUE SHAVER, IT JAMMED AND FAILED TO ROTATE. SWITCHED HANDPIECES AND BLADES AND AGAIN EXPERIENCED THE SAME ISSUE. THE PROCEDURE WAS CANCELLED, THEN RESCHEDULED AND COMPLETED USING ANOTHER EQUIPMENT. PATIENT WAS UNDER ANAESTHESIA. THERE WAS NO PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13489177 ↗MalfunctionJul 20, 2020Break; Mechanical Jam; Adverse Event Without Identified Device or Use Problem; Break; Mechanical Jam; Adverse Event Without Identified Device or Use Problem
ACCORDING TO THE REPORTER, DURING POLYPECTOMY, THE DEVICE HAD A MALFUNCTION THAT LEAD TO THE PROCEDURE BEING ABORTED. THE PATIENT WAS NOT ANESTHETIZED AND THE RE-SCHEDULED PROCEDURE WAS SUCCESSFUL.
Patient problems: No Code Available
MAUDE report 10299469 ↗MalfunctionJul 20, 2020Fail-Safe Problem; Adverse Event Without Identified Device or Use Problem; Fail-Safe Problem; Adverse Event Without Identified Device or Use Problem
ACCORDING TO THE REPORTER, DURING POLYPECTOMY, THE DEVICE HAD A MALFUNCTION THAT LEAD TO THE PROCEDURE BEING ABORTED. THE PATIENT WAS NOT ANESTHETIZED AND THE RE-SCHEDULED PROCEDURE WAS SUCCESSFUL.
Patient problems: No Code Available
MAUDE report 10299462 ↗
FDA clearance and classification
The FDA regulates product code HIH (hysteroscope (and accessories)) under 21 CFR 884.1690 as a Class II device. 21 CFR 884.1690 ↗
Devices of this type normally need 510(k) clearance before sale.
Common questions about the TRUCLEARTM Handpiece 7209807
Has the TRUCLEARTM Handpiece 7209807 been recalled?
No FDA recall record in our data names this exact device identifier (DI 03596010501943), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.
What problems have been reported with the TRUCLEARTM Handpiece 7209807?
The FDA's MAUDE database lists 9 adverse event reports for this device, received between Jul 20, 2020 and Jun 19, 2024. In those reports, the most common specific device problems are “Fail-Safe Problem” (6), “Mechanical Jam” (4) and “Intermittent Energy Output” (3). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the TRUCLEARTM Handpiece 7209807?
None of the reports are classified as deaths, 7 as injuries and 2 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Who makes the TRUCLEARTM Handpiece 7209807, and is it still sold?
The FDA's device database lists Smith & Nephew, Inc. as the company responsible for it (brand TRUCLEARTM Handpiece, model 7209807). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other HYSTEROSCOPE (AND ACCESSORIES) devices
- Olympus WA2T430AOLYMPUS Winter & Ibe GmbH · Hysteroscope (And Accessories)783 reports
- Olympus WA2T412AOLYMPUS Winter & Ibe GmbH · Hysteroscope (And Accessories)452 reports
- Olympus A22001AOLYMPUS Winter & Ibe GmbH · Hysteroscope (And Accessories)366 reports
- Olympus WA2T470AOLYMPUS Winter & Ibe GmbH · Hysteroscope (And Accessories)323 reports
- Olympus A22002AOLYMPUS Winter & Ibe GmbH · Hysteroscope (And Accessories)297 reports
- Olympus A22003AOLYMPUS Winter & Ibe GmbH · Hysteroscope (And Accessories)198 reports
- n.a. 27040XAKarl Storz GmbH & Co. KG · HYSTEROSCOPE (AND ACCESSORIES)102 reports
- VISERA HYSTEROVIDEOSCOPE OLYMPUS HYF TYPE VOlympus Medical Systems Corp. · HYSTEROSCOPE (AND ACCESSORIES)1 recall48 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page