GENESIS II 71430447Recalls & reported problems
Smith & Nephew, Inc. · GENESIS II CRUCIATE RETAINING DEEP FLEXION INSERT TRIAL SIZE 5-6 11M · Class 1
All models of GENESIS II (462)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 6
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 03596010512567), received between Sep 21, 2020 and Feb 6, 2024 — classified as 6 malfunctions.
The most frequently reported specific device problem is “Break” (4 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Break4
- Crack1
- Fracture1
- Material Deformation1
- Naturally Worn1
Patient problems
- No Clinical Signs, Symptoms or Conditions5
- No Consequences Or Impact To Patient2
Recent adverse event reports
The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.
MalfunctionFeb 6, 2024Fracture; Fracture
IT WAS REPORTED THAT DURING A TKA SURGERY, THE GII CR DEEP FLX ISR TR S5-6 11 GOT FRACTURED. ALL BROKEN PIECES WERE RETRIEVED USING PICKUPS. THE PROCEDURE WAS RESUMED, AFTER A NON-SIGNIFICANT DELAY, WITH A S+N BACK-UP DEVICE. PATIENT WAS NOT INJURED AS CONSEQUENCE OF THIS PROBLEM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18655496 ↗MalfunctionJun 8, 2022Break; Break
IT WAS REPORTED THAT, DURING TKA SURGERY, A GII CR DEEP FLX ISR TR S5-6 11 BROKE IN HALF WHILE TRIALING. DEVICE INSIDE THE PATIENT. THE PROCEDURE WAS COMPLETED WITHOUT DELAY USING A S+N BACK-UP DEVICE. PATIENT WAS NOT HARMED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14629571 ↗MalfunctionDec 3, 2021Crack; Crack
IT WAS REPORTED THAT THESE INSERT TRIALS AND SPACERS ARE CRACKED: GII PS HI FLEX ISRT TR S5-6 11, UNIV ISRT SPACER SZ5-6 13M, UNIV ISRT SPACER SZ 5-6 15MM, GII CR DEEP FLX ISR TR S5-6 9, AND GII CR DEEP FLX ISR TR S5-6 11. IT IS UNKNOWN WHETHER THE EVENT HAPPENED DURING SURGERY AND IF THERE WAS PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12924858 ↗MalfunctionNov 25, 2021Break; Material Deformation; Naturally Worn; Break; Material Deformation; Naturally Worn
IT WAS REPORTED THAT A GENESIS II CRUCIATE RETAINING DEEP FLEXION INSERT TRIAL SIZE 5-6 9MM (CASE: (B)(4)), A GENESIS II CRUCIATE RETAINING DEEP FLEXION INSERT TRIAL SIZE 5-6 11M (CASE: (B)(4)) AND A GENESIS II CRUCIATE RETAINING DEEP FLEXION INSERT TRIAL SIZE7-8 11M (CASE: (B)(4)) NEED REPLACING DUE TO WEAR AND TEAR, PLASTIC BURS, GOUGES, CRACKED ENDS, AMONG OTHERS. AS THIS WAS NOTICED UPON FIELD INSPECTION, THERE WAS NOT PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12874407 ↗MalfunctionSep 28, 2020Break; Break
IT WAS REPORTED THAT DURING SET UP THE GENESIS II CR DEEP FLX ISR TR S5-6 11 WAS FOUND BROKEN IN HALF. PROCEDURE FINISHED WITH S&N BACKUP DEVICE. NO HARM TO PATIENT. NO SURGERY DELAY.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10591658 ↗MalfunctionSep 21, 2020Break; Break
IT WAS REPORTED THAT DURING SURGERY THE INSERT TRIAL BROKE IN HALF OUTSIDE THE PATIENT. THE PROCEDURE WAS COMPLETED WITHOUT DELAY WITH THE SAME DEVICE. NO INJURY TO PATIENT WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions; No Consequences Or Impact To Patient
MAUDE report 10564205 ↗
Other ORTHOPEDIC MANUAL SURGICAL INSTRUMENT devices
- AML EXCEL 2001-42-000Depuy (Ireland) · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT2,009 reports
- NA 03.010.523Synthes GmbH · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT1,208 reports
- EXPRESSEW 214141Medos International Sàrl · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT1,166 reports
- N/A 9560524Medtronic Sofamor Danek, Inc. · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT866 reports
- INHANCE 6500-11-104Depuy (Ireland) · Orthopedic manual surgical instrument831 reports
- INHANCE 6500-11-102Depuy (Ireland) · Orthopedic manual surgical instrument567 reports
- R3 71364450Smith & Nephew, Inc. · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT520 reports
- N/A 9561524Medtronic Sofamor Danek, Inc. · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT494 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.