Hysteroscopic Fluid Management System Stand 7210165Recalls & reported problems
Smith & Nephew, Inc. · HYST. FLUID MGMT STAND · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 9
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 9 adverse event reports that list this exact device (DI 03596010529541), received between Apr 16, 2019 and Sep 3, 2020 — classified as 9 malfunctions.
The most frequently reported specific device problem is “Incorrect Measurement” (7 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Incorrect Measurement7
- Display or Visual Feedback Problem2
Patient problems
- No Patient Involvement5
- No Information2
- Blood Loss1
- Bowel Perforation1
- Fever1
- No Code Available1
- No Consequences Or Impact To Patient1
- Septic Shock1
Recent adverse event reports
The 9 most recent of 9 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 3, 2020Incorrect Measurement; Incorrect Measurement
ACCORDING TO THE REPORTER, DURING HYSTEROSCOPIC MYOMECTOMY, THE SURGEON USED A COMBINATION OF INSTRUMENTS TO REMOVE MYOMA'S. BOVIE PENCIL AND TRUCLEAR DENSE SHAVER (WHICH WAS NOT DEFECTIVE. USED FOR RESECTION ONLY). LARGE PIECES WERE REMOVED. PATIENT HAD BLEEDING FROM WHERE THESE MYOMA PORTIONS WERE JOINED TO THE CERVIX. SURGEON SUTURED THE CERVIX TO STOP THE BLEEDING. THE BLOOD LOSS WAS NOTED TO BE 1 LITER. ALSO NOTED WAS THERE WAS MORE FLUID OUTPUT THAN THERE WAS FLUID INPUT. FLUID SCALE NUMBERS WERE NEGATIVE THROUGHOUT PROCEDURE. MANUAL COUNTS OF FLUID WERE PERFORMED. HEMOGLOBIN AND HEMATOCRICT COUNT WAS ORDERED, HEMOGLOBIN WAS ~ 13 (PRE-OPERATIVE WAS ~ 14). THE BOWEL OF THE PATIENT PERFORATED AND CURRENTLY IN COLOSTOMY BAG. THE PATIENT HAD 102 DEGREES OF FEVER DAY AFTER THE PROCEDURE, INSTRUCTED TO GO TO EMERGENCY ROOM BUT DID NOT COME IMMEDIATELY. THE FOLLOWING DAY, PATIENT WAS TAKEN TO OTHER HOSPITAL EMERGENCY ROOM AND FOUND TO BE IN SEPSIS SHOCK AND NOT PRODUCING URINE. THEY FREE THE FLUID IN THE ABDOMEN AND A PERFORATION SEEN IN THE SIGMOID. THE PATIENT WAS RELEASED TO EMERGENCY ROOM 1-2 DAYS LATER.
Patient problems: Blood Loss; Bowel Perforation; Fever; No Code Available; Septic Shock
MAUDE report 10490859 ↗MalfunctionJan 29, 2020Incorrect Measurement; Incorrect Measurement
ACCORDING TO THE REPORTER, PRIOR TO USE, THE DEVICE SCALE OFF HAD INACCURATE READINGS. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Patient Involvement
MAUDE report 9641599 ↗MalfunctionJan 8, 2020Incorrect Measurement; Incorrect Measurement
ACCORDING TO THE REPORTER, PRIOR TO USE, THE DEVICE'S SCALE WAS NOT REGISTERING CORRECT VOLUME. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Patient Involvement
MAUDE report 9562198 ↗MalfunctionDec 12, 2019Incorrect Measurement; Incorrect Measurement
ACCORDING TO THE REPORTER, DURING TESTING, THE UNIT WAS NOT WORKING AND SHOWED LARGER DEFICIT WHEREIN WAS CONFIRMED BY DOING MANUAL COUNT. THE DEFICIT ON FIRST CASE WAS APPROXIMATELY 1100 CUBIC CENTIMETERS (CC) AND ON THE SECOND CASE WAS APPROXIMATELY 500 CC. THERE WAS NO PATIENT INVOLVED.
Patient problems: No Patient Involvement
MAUDE report 9459237 ↗MalfunctionOct 21, 2019Incorrect Measurement; Incorrect Measurement
ACCORDING TO THE REPORTER, DURING HYSTEROSCOPY MYOMECTOMY, THE LEGACY FLUID MANAGEMENT SCALE WENT INTO AN EXTENDED ¿BAG MODE¿ FOR AP PROXIMATELY 7-8 MINUTES WHICH RESET TO AN INCORRECT DEFICIT READING. A MANUAL COUNT WAS PERFORMED AND THE FLUID DEFICIT LIMIT WAS REACHED. THE CASE WAS STOPPED DUE TO REACHING THE MAXIMUM FLUID DEFICIT, WHICH WAS CALCULATED BY PERFORMING A MANUAL COUNT. THERE WAS NO PATIENT INJURY. NO ADDITIONAL DETAILS ARE AVAILABLE.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9215904 ↗MalfunctionSep 30, 2019Incorrect Measurement; Incorrect Measurement
ACCORDING TO THE REPORTER, DURING TESTING, THE DEVICE'S OUTFLOW ON THE SCALE WAS NOT MEASURING CORRECTLY. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Patient Involvement
MAUDE report 9135984 ↗MalfunctionAug 20, 2019Incorrect Measurement; Incorrect Measurement
ACCORDING TO THE REPORTER, DURING TESTING, THE DEVICE WAS NOT READING FLUID DEFICITS PROPERLY. THERE WAS NO PATIENT INVOLVED.
Patient problems: No Information
MAUDE report 8911486 ↗MalfunctionJun 5, 2019Display or Visual Feedback Problem; Display or Visual Feedback Problem
ACCORDING TO THE REPORTER, THE BOTTOM SCALE WAS NOT READING CORRECTLY.
Patient problems: No Information
MAUDE report 8671337 ↗MalfunctionApr 16, 2019Display or Visual Feedback Problem; Display or Visual Feedback Problem
ACCORDING TO THE REPORTER, DURING ANNUAL INSPECTION, THE UNIT WAS NOT MEASURING PROPERLY. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Patient Involvement
MAUDE report 8521009 ↗
Other INSUFFLATOR, HYSTEROSCOPIC devices
- Fluent Fluid Management System FLT-112SHologic, Inc. · INSUFFLATOR, HYSTEROSCOPIC140 reports
- Aquilex® Fluid Control System AQL-100CBSW.O.M. World of Medicine GmbH · Insufflator, Hysteroscopic1 recall18 reports
- Aquilex Fluid Control System AQL-100PBSW.O.M. World of Medicine GmbH · Insufflator, Hysteroscopic1 recall8 reports
- Aquilex® Fluid Control System AQL-100CSW.O.M. World of Medicine GmbH · Insufflator, Hysteroscopic15 reports
- ENDOMAT UP210Karl Storz GmbH & Co. KG · Insufflator, hysteroscopic4 deaths15 reports
- Fluent Console FLT-100Hologic, Inc. · INSUFFLATOR, HYSTEROSCOPIC1 death13 reports
- Hysteroscopic Fluid Management 7210164Smith & Nephew, Inc. · INSUFFLATOR, HYSTEROSCOPIC6 reports
- HysteroLux™ Fluid Management System Control Unit PH304W.O.M. World of Medicine GmbH · Insufflator, Hysteroscopic6 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page