Fluent Console FLT-100Recalls & reported problems
Hologic, Inc. · The Fluent®Fluid Management System is intended to provide liquid distension of the uterus during diagnostic and operative hysteroscopy, and to monitor the volume differential between the irrigation fluid flowing into and out of the uterus while providing drive, control and suction for hysteroscopic morcellators · Class 2
Data updated Nov 27, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 13
- Deaths reported
- 1
- Injuries reported
- 11
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 13 adverse event reports that list this exact device (DI 15420045508552), received between Oct 3, 2019 and Feb 16, 2026 — classified as 1 death, 11 injuries, 1 malfunction.
The most frequently reported specific device problem is “Use of Device Problem” (4 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem9
- Use of Device Problem4
Patient problems
- Hypervolemia7
- No Clinical Signs, Symptoms or Conditions2
- Pulmonary Edema2
- Abdominal Pain1
- Bronchospasm1
- Cardiogenic Shock1
- Incomplete Induced Abortion1
- Low Blood Pressure/ Hypotension1
Recent adverse event reports
The 13 most recent of 13 reports, in the reporter's own words as published by the FDA.
InjuryFeb 16, 2026Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT A PATIENT RECEIVED A NOVASURE PROCEDURE ON (B)(6) 2026, THE PHYSICIAN REPORTED THAT THE PATIENT HAD ¨DIFFERENT¨ ANATOMY AND CALLED A PARTNER AS WELL AS AN ULTRASOUND TECH TO ASSIST. ABLATION WAS COMPLETED AND A HYSTEROSCOPY USING A FLUENT AND OMNI SCOPE WAS PERFORMED PRE AND POST ABLATION WITH NO ISSUES TO NOTE NOR INCREASE IN DEFICIT WITH THE FLUENT MACHINE. PATIENT DID NOT HAD AN ENDOMETRIAL BIOPSY PRIOR PROCEDURE SO THE PHYSICIAN PERFORMED A DILATION AND CURETTAGE IN THE OPERATING ROOM. PATHOLOGY CAME BACK WITH RESULTS AS ACUTE CHRONIC INFLAMMATION. ON (B)(6) THE PATIENT COMPLAINED OF ABDOMINAL PAIN FOR WHICH DIFFERENT MEDICATIONS WERE GIVEN AT THE TIME. ON (B)(6) 2026 THE PATIENT PRESENTED TO THE EMERGENCY DEPARTMENT WITH PAIN AND CONFUSION. LACTATE WAS 6.7 AND BLOOD PRESSURE PER HOSPITAL WAS ¨38/15¨. PATIENT RECEIVED AN EXPLORATORY LAPAROSCOPY AND WHEN THE TROCARS WERE INSERTED PUSS WAS OOZING AND THE ABDOMEN OF THE PATIENT WAS FULL OF PUSS. NO THERMAL INJURY WAS OBSERVED AND A SPOT IN THE CORNUAL AREA WAS OBSERVED TO HAVE FAT PRESSED AGAINST THE UTERUS. PATIENT WAS TOO DELICATE TO PERFORM A HYSTERECTOMY AND WAS INTUBATED AND BROUGHT TO THE ICU ON ANTIBIOT
Patient problems: Abdominal Pain; Cardiogenic Shock; Sepsis; Septic Shock
MAUDE report 24348927 ↗InjuryMay 5, 2025Use of Device Problem; Use of Device Problem
IT WAS REPORTED THAT DURING A PROCEDURE USING A FLUENT SYSTEM ON (B)(6) 2025, A MAXIMUM DEFICIT OF 2500ML WAS REACHED. THE NURSE, UNKNOWN TO THE SURGEON, ZEROED THE DEFICIT AND THE PROCEDURE CONTINUED. WHEN THE PROCEDURE WAS COMPLETED, A MANUAL DEFICIT SHOWED THAT THE DEFICIT WAS 5000ML. PATIENT WAS SENT TO THE PACU, WHO MADE SURE SHE COULD EMPTY HER BLADDER. PATIENT WAS DISCHARGED WITH NO FOLLOW UP TREATMENT OR MEDICATION GIVEN.
Patient problems: Hypervolemia
MAUDE report 21964128 ↗InjuryDec 26, 2024Use of Device Problem; Use of Device Problem
IT WAS REPORTED THAT DURING A FLUENT PROCEDURE ON (B)(6) 2024, DURING A FLUENT PROCEDURE, IT WAS REPORTED THAT DURING THE PROCEDURE THE FLUENT ALARM ACTIVATED SEVERAL TIMES THAT THE FLUID DEFICIT LIMIT WAS REACHED BUT WAS OVER-RIDDEN SEVERAL TIMES AFTER THE 2500 ML LIMIT. NO ISSUES WERE REPORTED WITH THE MACHINED AND THE PHYSICIAN WAS THE ONE WHO CHOOSE TO CANCEL THE ALARMS AND TO CONTINUE THE PROCEDURE. THE DEFICIT WAS SHOWING AN APPROXIMATE 5L DEFICIT. PATIENT EXHIBITED SYMPTOMS RELATED TO FLUID OVERLOAD AND THE PATIENT HAD TO BE TRANSFERRED TO THE INTENSIVE CARE UNIT FOR 24 HOURS.
Patient problems: Hypervolemia
MAUDE report 21019685 ↗InjuryJan 25, 2024Use of Device Problem; Use of Device Problem
IT WAS REPORTED THAT DURING A FLUENT PROCEDURE ON (B)(6) 2024, THE PHYSICIAN RESECTED A FIBROID AND EXCEEDED THE DEFICIT LIMIT OF 2500 ML 3 TIMES. IT WAS REPORTED THAT THE DEFICIT WAS MET 3 TIMES AND, EACH TIME, A MANUAL COUNT WAS PERFORMED. IT WAS REPORTED THAT THE PATIENT HAD TO BE TAKEN TO THE INTENSIVE CARE UNIT. THE FINAL MANUAL COUNT WAS REPORTED AT 5000 ML. NO OTHER INFORMATION IS AVAILABLE.
Patient problems: Hypervolemia
MAUDE report 18579384 ↗InjuryNov 8, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT DURING A FLUENT PROCEDURE ON (B)(6) 2023, THE FLUID DEFICIT STAYED AT A CERTAIN NUMBER DURING THE PROCEDURE AND THE STAFF DID NOT REALIZE IT , THUS THE PATIENT RECEIVED A FLUID OVERLOAD ESTIMATED AT AROUND 2 LITERS. PATIENT BECAME UNSTABLE DUE TO BRONCHOSPASM AND AFTER THE SURGERY PRESENTED WITH COUGH. DUE THE PATIENT EXHIBITING SYMPTOMS OF FLUID OVERLOAD IT WAS ADMITTED TO THE INTENSIVE CARE UNIT AND RECEIVED A CHEST X-RAY AND TREATMENT WITH STEROIDS AND DIURETICS. THE PATIENT WAS DISCHARGED AFTER SEVERAL DAYS AND IS RECOVERING WELL. THE CONSOLE WAS EXAMINED AND TESTED AND IT WAS FOUND WITHIN TOLERANCE LIMIT. NO OTHER INFORMATION IS AVAILABLE.
Patient problems: Bronchospasm; Hypervolemia
MAUDE report 18092013 ↗InjuryJul 18, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT DURING A FLUENT PROCEDURE ON (B)(6) 2022 WHILE USING THE FLUID MANAGEMENT SYSTEM (FLUENT), THE PHYSICIAN NOTICED THAT THE FLOW RATE WAS HIGH WHEN AT 80 MMHG; HE THEN DECIDED TO DECREASE THE INTRA UTERINE PRESSURE TO 60 MMHG. SUDDENLY, THE TISSUE TRAP OF THE FLUENT SYSTEM EXPLODED, DUE TO THE PRESSURE OF THE SALINE SOLUTION. AT THE SAME TIME, THE PATIENT'S BLOOD PRESSURE DROPPED; THE PH DID DROP AS WELL (AROUND 6 ¿ ACIDOSIS) AND HER BODY TEMPERATURE AT 34C. THE ANESTHESIOLOGIST SUSPECTED A PULMONARY EMBOLISM BUT WAS DISCARDED. NO INJURIES RELATED TO HYPERVOLEMIA AND FLUID OVERLOAD WERE REPORTED AND THE PATIENT WAS KEPT IN OBSERVATION , THE PATIENT DID RECOVER BY THAT AFTERNOON, AND NO OTHER ADVERSE EVENTS WERE NOTED: NO NEUROLOGICAL DISORDER OBSERVED. THE PATIENT WAS HOSPITALIZED FOR ONE NIGHT UNDER OBSERVATION. THE NEXT DAY, SHE WAS BETTER AND RETURNED BACK HOME. THE PATIENT DID NOT REQUIRE ADDITIONAL TREATMENT. NO OTHER INFORMATION IS AVAILABLE.
Patient problems: Low Blood Pressure/ Hypotension; No Clinical Signs, Symptoms or Conditions
MAUDE report 15042476 ↗InjuryApr 21, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT ON (B)(6), AFTER A FLUENT PROCEDURE THE PATIENT HAD TO BE KEPT IN OBSERVATION DUE TO POSSIBLE FLUID OVERLOAD. THE PATIENT EVOLVED WELL AND WAS DISCHARGED. NO OTHER INFORMATION IS AVAILABLE.
Patient problems: Hypervolemia
MAUDE report 14175685 ↗InjuryAug 11, 2021Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT DURING A MYOMECTOMY THAT INVOLVED A FLUENT ON JULY 16TH, DURING THE PROCEDURE THE DEFICIT JUMPED TO 5000 AND THE STAFF HAD TO PERFORM A MANUAL COUNT. THE PROCEDURE WAS COMPLETED AND AT THE END OF THE PROCEDURE THE MANUAL COUNT WAS 2500 ML. THE PATIENT HAD TO BE READMITTED TO THE HOSPITAL AFTER THE SURGERY DUE TO A PULMONARY EDEMA. NO OTHER INFORMATION IS AVAILABLE
Patient problems: Pulmonary Edema
MAUDE report 12308626 ↗InjuryJul 6, 2021Use of Device Problem; Use of Device Problem
IT WAS REPORTED THAT ON JUNE 17TH DURING A FLUENT PROCEDURE, THE FLUID DEFICIT QUICKLY AUGMENTED REACHING A DEFICIT OF 2500 ML, THE FLUENT CONSOLE STOPPED THE INFLOW AND ALERTED THE PHYSICIAN ABOUT REACHING THE LIMIT FOR 2500 ML, THEN THE PHYSICIAN DECIDED TO CONTINUE WITH THE PROCEDURE BY RESETTING THE FLUID DEFICIT TO ZERO. AFTER THE PROCEDURE, THE PATIENT WAS LOW ON OXYGEN AND PRESENTED WITH FLUID IN HER LUNGS AND BLOOD IN HER URINE WHICH PROMPTED THE STAFF TO ADMIT HER TO THE EMERGENCY ROOM. NO OTHER INFORMATION IS AVAILABLE.
Patient problems: Hypervolemia; Pulmonary Edema; Unspecified Kidney or Urinary Problem
MAUDE report 12124368 ↗InjuryJun 29, 2021Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT AFTER A FLUENT PROCEDURE THAT INVOLVED A PATIENT WHO HAD RETAINED PRODUCTS OF CONCEPTION THE PHYSICIAN HAD TO PERFORM A HYSTERECTOMY ON THE PATIENT. NO OTHER INFORMATION AVAILABLE.
Patient problems: Incomplete Induced Abortion
MAUDE report 12088984 ↗MalfunctionFeb 19, 2021Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT ON JANUARY 18TH, DURING A FLUENT PROCEDURE THE OUT-FLOPAK BROKE APART ON 3 OCCASIONS. NO INJURIES TO THE PATIENT OR STAFF WERE REPORTED AND THE PROCEDURE WAS COMPLETED WITH A REPLACEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11352442 ↗DeathOct 29, 2020Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT ON SEPTEMBER 28TH DOCTOR PERFORMED A MYOSURE LITE PROCEDURE ON AN (B)(6) PATIENT WHO HAS HIGH BLOOD PRESSURE AND WAS POST MENOPAUSAL PATIENT TO OBTAIN SOME SAMPLES. THE DEFICIT DID NOT EXCEED 100 ON THE SYSTEM AND THE TECHS DID NOT DO A MANUAL COUNT. TWO DAYS LATER THE PATIENT PASSED AWAY FROM A PULMONARY EMBOLISM. NO FURTHER INFORMATION WAS PROVIDED.
Patient problems: Pulmonary Embolism
MAUDE report 10759216 ↗InjuryOct 3, 2019Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT DURING THE PROCEDURE, FOUR TIMES WHEN AN 800 ML FLUID DEFICIT WAS REACHED, THE MACHINE ALARMED AND ZEROED OUT THE DEVICE. PROCEDURE CARRIED ON ALL FOUR TIMES. AFTER THE FOURTH TIME, THEY SET THE DEFICIT LIMIT TO 2500 ML. THE ALARM SOUNDED ONCE 2500 ML WAS REACHED, AND THE DEVICE WAS ZEROED OUT AGAIN AND THE PROCEDURE CONTINUED. THE WASTE BAG FILLED UP. A NON-HOLOGIC DEVICE WAS USED TO SUCTION OUT THE FLUID. SIX 3 L BAGS WERE USED. THEY ATTEMPTED TO USE THIS INFORMATION TO OBTAIN A MANUAL COUNT. THE FINAL TOTAL DEFICIT WAS 9,800 ML. THE PATIENT WAS BROUGHT TO THE ICU FOLLOWING THE PROCEDURE. SHE HAD INTERSTITIAL SWELLING OF THE FACE AND SWELLING OF THE TONGUE. SHE WAS RELEASED FROM THE HOSPITAL A DAY AFTER THE PROCEDURE.
Patient problems: Hypervolemia
MAUDE report 9153882 ↗
FDA clearance and classification
Cleared through 510(k) K180825 on Aug 3, 2018 (decision: Substantially Equivalent). Submitted by Hologic, Inc. as “Fluent Fluid Management System”. The FDA review took 126 days. FDA record ↗
The FDA regulates product code HIG (insufflator, hysteroscopic) under 21 CFR 884.1700 as a Class II device. 21 CFR 884.1700 ↗
Common questions about the Fluent Console FLT-100
Has the Fluent Console FLT-100 been recalled?
No FDA recall record in our data names this exact device identifier (DI 15420045508552), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.
What problems have been reported with the Fluent Console FLT-100?
The FDA's MAUDE database lists 13 adverse event reports for this device, received between Oct 3, 2019 and Feb 16, 2026. In those reports, the most common specific device problem is “Use of Device Problem” (4); the most common patient problems are “Hypervolemia” (7), “Pulmonary Edema” (2) and “Abdominal Pain” (1). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the Fluent Console FLT-100?
1 report is classified as a death, 11 as injuries and 1 as a malfunction. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the Fluent Console FLT-100 FDA approved?
It was cleared through 510(k) K180825 on Aug 3, 2018. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.
Who makes the Fluent Console FLT-100, and is it still sold?
The FDA's device database lists Hologic, Inc. as the company responsible for it (brand Fluent Console, model FLT-100). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other INSUFFLATOR, HYSTEROSCOPIC devices
- Fluent Fluid Management System FLT-112SHologic, Inc. · INSUFFLATOR, HYSTEROSCOPIC140 reports
- Aquilex® Fluid Control System AQL-100CBSW.O.M. World of Medicine GmbH · Insufflator, Hysteroscopic1 recall18 reports
- Aquilex Fluid Control System AQL-100PBSW.O.M. World of Medicine GmbH · Insufflator, Hysteroscopic1 recall8 reports
- ENDOMAT UP210Karl Storz GmbH & Co. KG · Insufflator, hysteroscopic4 deaths17 reports
- Aquilex® Fluid Control System AQL-100CSW.O.M. World of Medicine GmbH · Insufflator, Hysteroscopic15 reports
- Hysteroscopic Fluid Management System Stand 7210165Smith & Nephew, Inc. · INSUFFLATOR, HYSTEROSCOPIC9 reports
- Fluent Pro Fluid Management System FLT-200Hologic, Inc. · Insufflator, hysteroscopic1 death6 reports
- Hysteroscopic Fluid Management 7210164Smith & Nephew, Inc. · INSUFFLATOR, HYSTEROSCOPIC6 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page