ENDOMAT UP210Recalls & reported problems
Karl Storz GmbH & Co. KG · ENDOMAT SELECT · Class 2
Data updated Nov 27, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 17
- Deaths reported
- 4
- Injuries reported
- 10
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 17 adverse event reports that list this exact device (DI 04048551376041), received between Sep 13, 2021 and Sep 4, 2026 — classified as 4 deaths, 10 injuries, 3 malfunctions.
The most frequently reported specific device problem is “Infusion or Flow Problem” (4 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Most reported problems
Device problems
- Infusion or Flow Problem4
- Insufficient Device Problem Information4
- Adverse Event Without Identified Device or Use Problem2
- Air/Gas in Device2
- Device Sensing Problem2
- Pumping Stopped2
- Improper Flow or Infusion1
- Lack of Effect1
Patient problems
- No Clinical Signs, Symptoms or Conditions9
- Unspecified Heart Problem4
- Air Embolism3
- Embolism/Embolus1
- Ischemic Heart Disease1
- Unspecified Vascular Problem1
Recent adverse event reports
The 17 most recent of 17 reports, in the reporter's own words as published by the FDA.
InjurySep 4, 2026Device Sensing Problem
It was reported that when the staff went to attach the irrigation tubing to the machine a 001B code showed and no matter how many times it was taken off and reattached the tubing would not recognize. Had to use regular irrigation method to finish the case. No negative impact in state of health reported.\\Due to the new received information the infuser pump would not work so had to use a pressure bag a s use the bigatti scope as a saline hysteroscope. The procedure was completed successfully but the procedure was prolonged >30 mins and <60 mins. No negative impact in state of health\reported.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50619732 ↗InjurySep 4, 2026Device Sensing Problem
It was reported that when the staff went to attach the irrigation tubing to the machine a 001B code showed and no matter how many times it was taken off and reattached the tubing would not recognize. Had to use regular irrigation method to finish the case. No negative impact in state of health reported.\\Due to the new received information the infuser pump would not work so had to use a pressure bag a s use the bigatti scope as a saline hysteroscope. The procedure was completed successfully but the procedure was prolonged >30 mins and <60 mins. No negative impact in state of health\reported.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50619727 ↗MalfunctionJul 10, 2026Moisture Damage
IT WAS REPORTED THAT DURING HYSTEROSCOPY ON THE AFTERNOON OF JUNE 1, THE UTERINE DISTENSION MACHINE WAS DAMAGED DUE TO FLUID INFLOW. FURTHER INFORMATION WAS RECEIVED STATING THAT UPON VERIFICATION, THE DEVICE BECAME UNUSABLE BECAUSE SALINE LEAKED FROM THE SALINE BAG AFTER THE TUBING WAS CONNECTED TO IT DURING SURGICAL PREPARATION, SPILLING ONTO THE UTERINE DISTENSION PUMP. ACCORDING TO THE INFORMATION AVAILABLE, A BACKUP DEVICE WAS REPLACED PROMPTLY AFTERWARDS TO COMPLETE THE EXAMINATION. NO NEGATIVE IMPACT IN STATE OF HEALTH REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25786910 ↗InjuryMay 8, 2026Output Problem; No Apparent Adverse Event
THE FOLLOWING WAS REPORTED: "ON (B)(6) 2026 AT 09:00, IN OPERATING ROOM 3, THE PATIENT UNDERWENT HYSTEROSCOPY AND ENDOMETRIAL LESION RESECTION UNDER GENERAL ANESTHESIA. DURING THE PROCEDURE, WHILE USING THE GYNECOLOGICAL SHAVER, THE ENDOSCOPIC IRRIGATION AND SUCTION DEVICE EXPERIENCED A DISPLAY INTERFACE MALFUNCTION, PREVENTING NAVIGATION TO THE NEXT SCREEN AND FAILING TO ASPIRATE TISSUE NORMALLY. THE DEVICE WAS IMMEDIATELY POWERED OFF AND RESTARTED, AND MAINTENANCE PERSONNEL WERE NOTIFIED. AFTER THE RESTART, THE DEVICE RETURNED TO THE NORMAL INTERFACE BUT SUBSEQUENTLY TRIGGERED A FAULT ALARM AGAIN. THE PHYSICIAN CHANGED THE SPECIMEN COLLECTION METHOD, MANUALLY USING A SUCTION CANISTER TO ASPIRATE THE TISSUE. "
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25126033 ↗MalfunctionApr 29, 2025Material Puncture/Hole; Insufficient Device Problem Information; Material Puncture/Hole; Insufficient Device Problem Information
IT WAS REPORTED THAT THE "USER IS EXPERIENCING SOME AIR BUBBLES" DURING PROCEDURE. NO NEGATIVE IMPACT IN STATE OF HEALTH REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21931271 ↗InjuryAug 22, 2024Adverse Event Without Identified Device or Use Problem; Air/Gas in Device; Adverse Event Without Identified Device or Use Problem; Air/Gas in Device
IT WAS REPORTED THAT THE SALINE SUPPLY RAN OUT DURING SURGERY, SO THE NURSE WAS INSTRUCTED TO REPLACE IT. AFTER THE REPLACEMENT, AIR WAS MIXED IN WHEN THE STOPCOCK WAS RELEASED, CAUSING THE SALINE TO FLOW AFTER A DELAY OF SEVERAL SECONDS. THE PATIENT DEVELOPED AIR EMBOLISM AFTER SURGERY, BUT RECOVERED AND WAS DISCHARGED. DUE TO THE ADVERSE CONSEQUENCES FOR THE PATIENT AND THE PROLONGED HOSPITALIZATION THE CASE IS DEEMED REPORTABLE.
Patient problems: Air Embolism
MAUDE report 20028133 ↗InjuryAug 21, 2024Air/Gas in Device; Air/Gas in Device
IT WAS REPORTED THAT THE SALINE SUPPLY RAN OUT DURING SURGERY, SO THE NURSE WAS INSTRUCTED TO REPLACE IT. AFTER THE REPLACEMENT, AIR WAS MIXED IN WHEN THE STOPCOCK WAS RELEASED, CAUSING THE SALINE TO FLOW AFTER A DELAY OF SEVERAL SECONDS. THE PATIENT DEVELOPED AIR EMBOLISM AFTER SURGERY, BUT RECOVERED AND WAS DISCHARGED. DUE TO THE ADVERSE CONSEQUENCES FOR THE PATIENT AND THE PROLONGED HOSPITALIZATION THE CASE IS DEEMED REPORTABLE.
Patient problems: Air Embolism
MAUDE report 20013009 ↗InjuryMar 7, 2024Insufficient Device Problem Information; Insufficient Device Problem Information
IT WAS REPORTED THAT THE ENDOMAT SELECT DIDN'T HAVE ADEQUATE SUCTION DURING A PROCEDURE. THE PROCEDURE COULD NOT BE CONTINUED AND THE CASE WAS SUSPENDED. NO DEATH OR (UNANTICIPATED) SERIOUS DETERIORATION IN STATE OF HEALTH REPORTED.SINCE THE SURGERY WAS SUSPENDED, THE CASE IS DEEMED AS REPORTABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18857846 ↗MalfunctionMar 7, 2024Problem with Software Installation; Insufficient Device Problem Information; Problem with Software Installation; Insufficient Device Problem Information
IT WAS REPORTED THAT THE ENDOMAT SELECT DIDN'T HAVE ADEQUATE SUCTION DURING A PROCEDURE. THE PROCEDURE COULD NOT BE CONTINUED AND THE CASE WAS SUSPENDED. NO DEATH OR (UNANTICIPATED) SERIOUS DETERIORATION IN STATE OF HEALTH REPORTED. SINCE THE SURGERY WAS SUSPENDED, THE CASE IS DEEMED AS REPORTABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18857845 ↗DeathJan 5, 2024Infusion or Flow Problem; Infusion or Flow Problem
WHEN THE SURGEON CONNECTED THE TUBING SET 030647-10 TO ENDOMAT SELECT UP210 AND STARTED USING THEM, AIR GOT MIXED IN AND THE FIELD OF VIEW BECAME POOR. AT THIS TIME, WATER WAS BEING SUPPLIED, BUT SUCTION WAS NOT SUFFICIENT. ABOUT 20 MINUTES AFTER THE START OF THE SURGERY, THE PATIENT'S HEART RATE AND BLOOD PRESSURE DECREASED AND SUDDENLY CHANGED. AN ECHOCARDIOGRAM CONFIRMED THAT AIR WAS ENTERING THE RIGHT ATRIUM.
Patient problems: Unspecified Heart Problem
MAUDE report 18455754 ↗DeathJan 5, 2024Infusion or Flow Problem; Infusion or Flow Problem
WHEN THE SURGEON CONNECTED THE TUBING SET 030647-10 TO ENDOMAT SELECT UP210 AND STARTED USING THEM, AIR GOT MIXED IN AND THE FIELD OF VIEW BECAME POOR. AT THIS TIME, WATER WAS BEING SUPPLIED, BUT SUCTION WAS NOT SUFFICIENT. ABOUT 20 MINUTES AFTER THE START OF THE SURGERY, THE PATIENT'S HEART RATE AND BLOOD PRESSURE DECREASED AND SUDDENLY CHANGED. AN ECHOCARDIOGRAM CONFIRMED THAT AIR WAS ENTERING THE RIGHT ATRIUM.
Patient problems: Unspecified Heart Problem
MAUDE report 18455749 ↗DeathJan 5, 2024Infusion or Flow Problem; Infusion or Flow Problem
WHEN THE SURGEON CONNECTED THE TUBING SET 030647-10 TO ENDOMAT SELECT UP210 AND STARTED USING THEM, AIR GOT MIXED IN AND THE FIELD OF VIEW BECAME POOR. AT THIS TIME, WATER WAS BEING SUPPLIED, BUT SUCTION WAS NOT SUFFICIENT. ABOUT 20 MINUTES AFTER THE START OF THE SURGERY, THE PATIENT'S HEART RATE AND BLOOD PRESSURE DECREASED AND SUDDENLY CHANGED. AN ECHOCARDIOGRAM CONFIRMED THAT AIR WAS ENTERING THE RIGHT ATRIUM.
Patient problems: Unspecified Heart Problem
MAUDE report 18455748 ↗DeathJan 5, 2024Infusion or Flow Problem; Lack of Effect; Infusion or Flow Problem; Lack of Effect
WHEN THE SURGEON CONNECTED THE TUBING SET 030647-10 TO ENDOMAT SELECT UP210 AND STARTED USING THEM, AIR GOT MIXED IN AND THE FIELD OF VIEW BECAME POOR. AT THIS TIME, WATER WAS BEING SUPPLIED, BUT SUCTION WAS NOT SUFFICIENT. ABOUT 20 MINUTES AFTER THE START OF THE SURGERY, THE PATIENT'S HEART RATE AND BLOOD PRESSURE DECREASED AND SUDDENLY CHANGED. AN ECHOCARDIOGRAM CONFIRMED THAT AIR WAS ENTERING THE RIGHT ATRIUM.
Patient problems: Unspecified Heart Problem; Unspecified Vascular Problem
MAUDE report 18455689 ↗InjuryMay 25, 2023Pumping Stopped; Pumping Stopped
THE EVENT OCCURRED IN GREAT BRITAIN. IT WAS REPORTED THAT THE STOPCOCK ON THE SHEATH LEAKED DURING USE AND THE PUMP STOPPED WORKING DURING SURGERY. THERE WAS A PROLONGATION OF 1 HOUR AND THEY HAD TO CONVERT TO A MINI-LAPAROTOMY (OPEN SURGERY) TO REMOVE THE PROSTATE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17004246 ↗InjuryMay 25, 2023Pumping Stopped; Pumping Stopped
THE EVENT OCCURRED IN (B)(6). IT WAS REPORTED THAT THE STOPCOCK ON THE SHEATH LEAKED DURING USE AND THE PUMP STOPPED WORKING DURING SURGERY. THERE WAS A PROLONGATION OF 1 HOUR AND THEY HAD TO CONVERT TO A MINI-LAPAROTOMY (OPEN SURGERY) TO REMOVE THE PROSTATE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17004245 ↗InjuryDec 10, 2021Adverse Event Without Identified Device or Use Problem; Insufficient Device Problem Information; Adverse Event Without Identified Device or Use Problem; Insufficient Device Problem Information
FURTHER INFORMATION RECEIVED ABOUT THE CONSEQUENCES FOR THE PATIENT: "-(B)(6) 2021: MS T TRANSFER VIA SAMU 15 IN THE HYPERBARIC CHAMBER THEN IN INTENSIVE CARE SURGICAL FOR NON-INVASIVE VENTILATION - NIGHT FROM (B)(6): PATIENT EXTUBATED IN THE NIGHT - (B)(6) 2021: MRI PERFORMED, BEING INTERPRETED BECAUSE OF THE OBSERVATION OF HYPOTONIA OF THE ARM - SUSPICION OF AN ISCHEMIC HEART PROBLEM - THE ANATOMO-PATHOLOGICAL SAMPLE IS SENT TO THE LABORATORY FOR ANALYSIS." (MFR'S. INTERNAL COMPLAINT (B)(4)).
Patient problems: Air Embolism; Ischemic Heart Disease
MAUDE report 12975621 ↗InjurySep 13, 2021Use of Device Problem; Improper Flow or Infusion; Use of Device Problem; Improper Flow or Infusion
AS PER A MANUFACTURER INCIDENT REPORT WE RECEIVED FROM THE FACTORY IN (B)(6): IT WAS REPORTED THAT THE CUSTOMER USED AN ARTHROSCOPY SETUP ENDOSCOPE SETUP FOR A HYSTEROSCOPY. SINCE THE ARTHROSCOPY SOFTWARE IS NOT INTENDED FOR THIS TYPE OF TREATMENT, THE CUSTOMER PROVOKED AN EMBOLISM. ACCORDING TO CUSTOMER'S INFORMATION, THE PATIENT HAS RECOVERED COMPLETELY.
Patient problems: Embolism/Embolus
MAUDE report 12460147 ↗
FDA clearance and classification
The FDA regulates product code HIG (insufflator, hysteroscopic) under 21 CFR 884.1700 as a Class II device. 21 CFR 884.1700 ↗
Devices of this type normally need 510(k) clearance before sale.
Common questions about the ENDOMAT UP210
Has the ENDOMAT UP210 been recalled?
No FDA recall record in our data names this exact device identifier (DI 04048551376041), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.
What problems have been reported with the ENDOMAT UP210?
The FDA's MAUDE database lists 17 adverse event reports for this device, received between Sep 13, 2021 and Sep 4, 2026. In those reports, the most common specific device problems are “Infusion or Flow Problem” (4), “Air/Gas in Device” (2) and “Device Sensing Problem” (2); the most common patient problems are “Unspecified Heart Problem” (4), “Air Embolism” (3) and “Embolism/Embolus” (1). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the ENDOMAT UP210?
4 reports are classified as deaths, 10 as injuries and 3 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Who makes the ENDOMAT UP210, and is it still sold?
The FDA's device database lists Karl Storz GmbH & Co. KG as the company responsible for it (brand ENDOMAT, model UP210). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other Insufflator, hysteroscopic devices
- Fluent Fluid Management System FLT-112SHologic, Inc. · INSUFFLATOR, HYSTEROSCOPIC140 reports
- Aquilex® Fluid Control System AQL-100CBSW.O.M. World of Medicine GmbH · Insufflator, Hysteroscopic1 recall18 reports
- Aquilex Fluid Control System AQL-100PBSW.O.M. World of Medicine GmbH · Insufflator, Hysteroscopic1 recall8 reports
- Aquilex® Fluid Control System AQL-100CSW.O.M. World of Medicine GmbH · Insufflator, Hysteroscopic15 reports
- Fluent Console FLT-100Hologic, Inc. · INSUFFLATOR, HYSTEROSCOPIC1 death13 reports
- Hysteroscopic Fluid Management System Stand 7210165Smith & Nephew, Inc. · INSUFFLATOR, HYSTEROSCOPIC9 reports
- Fluent Pro Fluid Management System FLT-200Hologic, Inc. · Insufflator, hysteroscopic1 death6 reports
- Hysteroscopic Fluid Management 7210164Smith & Nephew, Inc. · INSUFFLATOR, HYSTEROSCOPIC6 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page