ENDOMAT UP210Recalls & reported problems
Karl Storz GmbH & Co. KG · ENDOMAT SELECT · Class 2
- Adverse event reports
- 15
- Deaths reported
- 4
- Injuries reported
- 8
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 15 adverse event reports that list this exact device (DI 04048551376041), received between Sep 13, 2021 and Jul 10, 2026 — classified as 4 deaths, 8 injuries, 3 malfunctions.
The most frequently reported specific device problem is “Infusion or Flow Problem” (4 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Infusion or Flow Problem4
- Insufficient Device Problem Information4
- Adverse Event Without Identified Device or Use Problem2
- Air/Gas in Device2
- Pumping Stopped2
- Improper Flow or Infusion1
- Lack of Effect1
- Material Puncture/Hole1
Patient problems
- No Clinical Signs, Symptoms or Conditions7
- Unspecified Heart Problem4
- Air Embolism3
- Embolism/Embolus1
- Ischemic Heart Disease1
- Unspecified Vascular Problem1
Recent adverse event reports
The 15 most recent of 15 reports, in the reporter's own words as published by the FDA.
MalfunctionJul 10, 2026Moisture Damage
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25786910 ↗InjuryMay 8, 2026Output Problem; No Apparent Adverse Event
THE FOLLOWING WAS REPORTED: "ON (B)(6) 2026 AT 09:00, IN OPERATING ROOM 3, THE PATIENT UNDERWENT HYSTEROSCOPY AND ENDOMETRIAL LESION RESECTION UNDER GENERAL ANESTHESIA. DURING THE PROCEDURE, WHILE USING THE GYNECOLOGICAL SHAVER, THE ENDOSCOPIC IRRIGATION AND SUCTION DEVICE EXPERIENCED A DISPLAY INTERFACE MALFUNCTION, PREVENTING NAVIGATION TO THE NEXT SCREEN AND FAILING TO ASPIRATE TISSUE NORMALLY. THE DEVICE WAS IMMEDIATELY POWERED OFF AND RESTARTED, AND MAINTENANCE PERSONNEL WERE NOTIFIED. AFTER THE RESTART, THE DEVICE RETURNED TO THE NORMAL INTERFACE BUT SUBSEQUENTLY TRIGGERED A FAULT ALARM AGAIN. THE PHYSICIAN CHANGED THE SPECIMEN COLLECTION METHOD, MANUALLY USING A SUCTION CANISTER TO ASPIRATE THE TISSUE. "
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25126033 ↗MalfunctionApr 29, 2025Material Puncture/Hole; Insufficient Device Problem Information; Material Puncture/Hole; Insufficient Device Problem Information
IT WAS REPORTED THAT THE "USER IS EXPERIENCING SOME AIR BUBBLES" DURING PROCEDURE. NO NEGATIVE IMPACT IN STATE OF HEALTH REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21931271 ↗InjuryAug 22, 2024Adverse Event Without Identified Device or Use Problem; Air/Gas in Device; Adverse Event Without Identified Device or Use Problem; Air/Gas in Device
IT WAS REPORTED THAT THE SALINE SUPPLY RAN OUT DURING SURGERY, SO THE NURSE WAS INSTRUCTED TO REPLACE IT. AFTER THE REPLACEMENT, AIR WAS MIXED IN WHEN THE STOPCOCK WAS RELEASED, CAUSING THE SALINE TO FLOW AFTER A DELAY OF SEVERAL SECONDS. THE PATIENT DEVELOPED AIR EMBOLISM AFTER SURGERY, BUT RECOVERED AND WAS DISCHARGED. DUE TO THE ADVERSE CONSEQUENCES FOR THE PATIENT AND THE PROLONGED HOSPITALIZATION THE CASE IS DEEMED REPORTABLE.
Patient problems: Air Embolism
MAUDE report 20028133 ↗InjuryAug 21, 2024Air/Gas in Device; Air/Gas in Device
IT WAS REPORTED THAT THE SALINE SUPPLY RAN OUT DURING SURGERY, SO THE NURSE WAS INSTRUCTED TO REPLACE IT. AFTER THE REPLACEMENT, AIR WAS MIXED IN WHEN THE STOPCOCK WAS RELEASED, CAUSING THE SALINE TO FLOW AFTER A DELAY OF SEVERAL SECONDS. THE PATIENT DEVELOPED AIR EMBOLISM AFTER SURGERY, BUT RECOVERED AND WAS DISCHARGED. DUE TO THE ADVERSE CONSEQUENCES FOR THE PATIENT AND THE PROLONGED HOSPITALIZATION THE CASE IS DEEMED REPORTABLE.
Patient problems: Air Embolism
MAUDE report 20013009 ↗InjuryMar 7, 2024Insufficient Device Problem Information; Insufficient Device Problem Information
IT WAS REPORTED THAT THE ENDOMAT SELECT DIDN'T HAVE ADEQUATE SUCTION DURING A PROCEDURE. THE PROCEDURE COULD NOT BE CONTINUED AND THE CASE WAS SUSPENDED. NO DEATH OR (UNANTICIPATED) SERIOUS DETERIORATION IN STATE OF HEALTH REPORTED.SINCE THE SURGERY WAS SUSPENDED, THE CASE IS DEEMED AS REPORTABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18857846 ↗MalfunctionMar 7, 2024Problem with Software Installation; Insufficient Device Problem Information; Problem with Software Installation; Insufficient Device Problem Information
IT WAS REPORTED THAT THE ENDOMAT SELECT DIDN'T HAVE ADEQUATE SUCTION DURING A PROCEDURE. THE PROCEDURE COULD NOT BE CONTINUED AND THE CASE WAS SUSPENDED. NO DEATH OR (UNANTICIPATED) SERIOUS DETERIORATION IN STATE OF HEALTH REPORTED. SINCE THE SURGERY WAS SUSPENDED, THE CASE IS DEEMED AS REPORTABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18857845 ↗DeathJan 5, 2024Infusion or Flow Problem; Infusion or Flow Problem
WHEN THE SURGEON CONNECTED THE TUBING SET 030647-10 TO ENDOMAT SELECT UP210 AND STARTED USING THEM, AIR GOT MIXED IN AND THE FIELD OF VIEW BECAME POOR. AT THIS TIME, WATER WAS BEING SUPPLIED, BUT SUCTION WAS NOT SUFFICIENT. ABOUT 20 MINUTES AFTER THE START OF THE SURGERY, THE PATIENT'S HEART RATE AND BLOOD PRESSURE DECREASED AND SUDDENLY CHANGED. AN ECHOCARDIOGRAM CONFIRMED THAT AIR WAS ENTERING THE RIGHT ATRIUM.
Patient problems: Unspecified Heart Problem
MAUDE report 18455754 ↗DeathJan 5, 2024Infusion or Flow Problem; Infusion or Flow Problem
WHEN THE SURGEON CONNECTED THE TUBING SET 030647-10 TO ENDOMAT SELECT UP210 AND STARTED USING THEM, AIR GOT MIXED IN AND THE FIELD OF VIEW BECAME POOR. AT THIS TIME, WATER WAS BEING SUPPLIED, BUT SUCTION WAS NOT SUFFICIENT. ABOUT 20 MINUTES AFTER THE START OF THE SURGERY, THE PATIENT'S HEART RATE AND BLOOD PRESSURE DECREASED AND SUDDENLY CHANGED. AN ECHOCARDIOGRAM CONFIRMED THAT AIR WAS ENTERING THE RIGHT ATRIUM.
Patient problems: Unspecified Heart Problem
MAUDE report 18455749 ↗DeathJan 5, 2024Infusion or Flow Problem; Infusion or Flow Problem
WHEN THE SURGEON CONNECTED THE TUBING SET 030647-10 TO ENDOMAT SELECT UP210 AND STARTED USING THEM, AIR GOT MIXED IN AND THE FIELD OF VIEW BECAME POOR. AT THIS TIME, WATER WAS BEING SUPPLIED, BUT SUCTION WAS NOT SUFFICIENT. ABOUT 20 MINUTES AFTER THE START OF THE SURGERY, THE PATIENT'S HEART RATE AND BLOOD PRESSURE DECREASED AND SUDDENLY CHANGED. AN ECHOCARDIOGRAM CONFIRMED THAT AIR WAS ENTERING THE RIGHT ATRIUM.
Patient problems: Unspecified Heart Problem
MAUDE report 18455748 ↗DeathJan 5, 2024Infusion or Flow Problem; Lack of Effect; Infusion or Flow Problem; Lack of Effect
WHEN THE SURGEON CONNECTED THE TUBING SET 030647-10 TO ENDOMAT SELECT UP210 AND STARTED USING THEM, AIR GOT MIXED IN AND THE FIELD OF VIEW BECAME POOR. AT THIS TIME, WATER WAS BEING SUPPLIED, BUT SUCTION WAS NOT SUFFICIENT. ABOUT 20 MINUTES AFTER THE START OF THE SURGERY, THE PATIENT'S HEART RATE AND BLOOD PRESSURE DECREASED AND SUDDENLY CHANGED. AN ECHOCARDIOGRAM CONFIRMED THAT AIR WAS ENTERING THE RIGHT ATRIUM.
Patient problems: Unspecified Heart Problem; Unspecified Vascular Problem
MAUDE report 18455689 ↗InjuryMay 25, 2023Pumping Stopped; Pumping Stopped
THE EVENT OCCURRED IN GREAT BRITAIN. IT WAS REPORTED THAT THE STOPCOCK ON THE SHEATH LEAKED DURING USE AND THE PUMP STOPPED WORKING DURING SURGERY. THERE WAS A PROLONGATION OF 1 HOUR AND THEY HAD TO CONVERT TO A MINI-LAPAROTOMY (OPEN SURGERY) TO REMOVE THE PROSTATE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17004246 ↗InjuryMay 25, 2023Pumping Stopped; Pumping Stopped
THE EVENT OCCURRED IN (B)(6). IT WAS REPORTED THAT THE STOPCOCK ON THE SHEATH LEAKED DURING USE AND THE PUMP STOPPED WORKING DURING SURGERY. THERE WAS A PROLONGATION OF 1 HOUR AND THEY HAD TO CONVERT TO A MINI-LAPAROTOMY (OPEN SURGERY) TO REMOVE THE PROSTATE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17004245 ↗InjuryDec 10, 2021Adverse Event Without Identified Device or Use Problem; Insufficient Device Problem Information; Adverse Event Without Identified Device or Use Problem; Insufficient Device Problem Information
FURTHER INFORMATION RECEIVED ABOUT THE CONSEQUENCES FOR THE PATIENT: "-(B)(6) 2021: MS T TRANSFER VIA SAMU 15 IN THE HYPERBARIC CHAMBER THEN IN INTENSIVE CARE SURGICAL FOR NON-INVASIVE VENTILATION - NIGHT FROM (B)(6): PATIENT EXTUBATED IN THE NIGHT - (B)(6) 2021: MRI PERFORMED, BEING INTERPRETED BECAUSE OF THE OBSERVATION OF HYPOTONIA OF THE ARM - SUSPICION OF AN ISCHEMIC HEART PROBLEM - THE ANATOMO-PATHOLOGICAL SAMPLE IS SENT TO THE LABORATORY FOR ANALYSIS." (MFR'S. INTERNAL COMPLAINT (B)(4)).
Patient problems: Air Embolism; Ischemic Heart Disease
MAUDE report 12975621 ↗InjurySep 13, 2021Use of Device Problem; Improper Flow or Infusion; Use of Device Problem; Improper Flow or Infusion
AS PER A MANUFACTURER INCIDENT REPORT WE RECEIVED FROM THE FACTORY IN (B)(6): IT WAS REPORTED THAT THE CUSTOMER USED AN ARTHROSCOPY SETUP ENDOSCOPE SETUP FOR A HYSTEROSCOPY. SINCE THE ARTHROSCOPY SOFTWARE IS NOT INTENDED FOR THIS TYPE OF TREATMENT, THE CUSTOMER PROVOKED AN EMBOLISM. ACCORDING TO CUSTOMER'S INFORMATION, THE PATIENT HAS RECOVERED COMPLETELY.
Patient problems: Embolism/Embolus
MAUDE report 12460147 ↗
Other Insufflator, hysteroscopic devices
- Fluent Fluid Management System FLT-112SHologic, Inc. · INSUFFLATOR, HYSTEROSCOPIC140 reports
- Aquilex® Fluid Control System AQL-100CBSW.O.M. World of Medicine GmbH · Insufflator, Hysteroscopic1 recall18 reports
- Aquilex Fluid Control System AQL-100PBSW.O.M. World of Medicine GmbH · Insufflator, Hysteroscopic1 recall8 reports
- Aquilex® Fluid Control System AQL-100CSW.O.M. World of Medicine GmbH · Insufflator, Hysteroscopic15 reports
- Fluent Console FLT-100Hologic, Inc. · INSUFFLATOR, HYSTEROSCOPIC1 death13 reports
- Hysteroscopic Fluid Management System Stand 7210165Smith & Nephew, Inc. · INSUFFLATOR, HYSTEROSCOPIC9 reports
- Hysteroscopic Fluid Management 7210164Smith & Nephew, Inc. · INSUFFLATOR, HYSTEROSCOPIC6 reports
- HysteroLux™ Fluid Management System Control Unit PH304W.O.M. World of Medicine GmbH · Insufflator, Hysteroscopic6 reports