Skip to content

ACCU-PASS 7210425Recalls & reported problems

Smith & Nephew, Inc. · ACCU-PASS SUTURE SHUTTLE 45 DEGREE, UPBEND, STERILE · Class 1

All models of ACCU-PASS (7)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
10
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 10 adverse event reports that list this exact device (DI 03596010534460), received between Jun 30, 2020 and May 15, 2024 — classified as 2 injuries, 8 malfunctions.

The most frequently reported specific device problem is “Mechanical Jam” (5 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 1 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Mechanical Jam5
  • Delivered as Unsterile Product2
  • Break1
  • Failure to Eject1
  • Firing Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions8
  • No Known Impact Or Consequence To Patient2
  • Injury1
  • No Information1

Recent adverse event reports

The 10 most recent of 10 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMay 15, 2024Delivered as Unsterile Product; Delivered as Unsterile Product

    IT WAS REPORTED THAT AN ACCU-PASS HAD PIN HOLES FROM THE INSIDE ALONG THE FLAT SIDE OF THE PACKAGING. AS THE DEFICIENCY WAS OBSERVED WHILE PERFORMING THE INVESTIGATION FOR THE DEVICE, THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19324133 ↗
  • MalfunctionJun 7, 2021Firing Problem; Firing Problem

    IT WAS REPORTED THAT, DURING A HIP ARTHROSCOPY, THE "ACCUPASS 45° (UPBEND, STERILE)" COULD NOT ROLL THE MONOFILAMENT OUT OF THE DEVICE WHEN IT WAS USED. THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITHOUT SIGNIFICANT DELAY USING A BACK-UP DEVICE. NO PATIENT INJURIES OR OTHER COMPLICATIONS WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11952151 ↗
  • MalfunctionFeb 15, 2021Mechanical Jam; Mechanical Jam

    IT WAS REPORTED THAT THE ACCU-PASS STR SHUTTLE 45 DEG UP DID NOT ADVANCE THE NYLON MONOFILAMENT. THE PROCEDURE WAS COMPLETED WITH A DELAY OF 30 MINUTES OR LESS USING A BACK-UP DEVICE. NO PATIENT INJURY OR OTHER COMPLICATIONS WERE REPORTED. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED ACCORDINGLY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11328034 ↗
  • MalfunctionFeb 15, 2021Mechanical Jam; Mechanical Jam

    IT WAS REPORTED THAT THE ACCU-PASS STR SHUTTLE 45 DEG UP DID NOT ADVANCE THE NYLON MONOFILAMENT. THE PROCEDURE WAS COMPLETED WITH A DELAY OF 30 MINUTES OR LESS USING A BACK-UP DEVICE. NO PATIENT INJURY OR OTHER COMPLICATIONS WERE REPORTED. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED ACCORDINGLY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11328027 ↗
  • MalfunctionFeb 15, 2021Mechanical Jam; Mechanical Jam

    IT WAS REPORTED THAT THE ACCU-PASS STR SHUTTLE 45 DEG UP DID NOT ADVANCE THE NYLON MONOFILAMENT. THE PROCEDURE WAS COMPLETED WITH A DELAY OF 30 MINUTES OR LESS USING A BACK-UP DEVICE. NO PATIENT INJURY OR OTHER COMPLICATIONS WERE REPORTED. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED ACCORDINGLY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11328023 ↗
  • MalfunctionFeb 4, 2021Break; Break

    IT WAS REPORTED THAT DURING AN ARTHROSCOPY, THE ACCU-PASS 45-DEGREE, UPBEND, STERILE METAL SHAFT CAME OUT OF THE HANDLE. THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITHOUT DELAY USING A BACK-UP SMITH AND NEPHEW DEVICE. NO PATIENT INJURY OR OTHER COMPLICATIONS WERE REPORTED. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED ACCORDINGLY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11278661 ↗
  • InjuryNov 19, 2020Delivered as Unsterile Product; Delivered as Unsterile Product

    IT WAS REPORTED THAT DURING MENISCUS REPAIR PROCEDURE, WHEN THE OUT OF BOX WAS OPENED, THE STERILIZATION PACK SEAL OF THE ACCU-PASS 45-DEGREE DID NOT CLOSE FROM THE FIRST. THE DEVICE WAS USED ON THE PATIENT. A BACKUP DEVICE WAS AVAILABLE TO COMPLETE THE PROCEDURE WITH NO DELAY OR PATIENT INJURIES. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED ACCORDINGLY.

    Patient problems: No Clinical Signs, Symptoms or Conditions; No Information

    MAUDE report 10865731 ↗
  • InjurySep 23, 2020Mechanical Jam; Mechanical Jam

    IT WAS REPORTED THAT, DURING A PROCEDURE, THE MONOFILAMENT WOULD NOT COME OUT OF THE ACCUPASS, RENDERING IT UNUSABLE. THERE WAS A DIFFERENT S&N DEVICE AVAILABLE, WHICH WAS USED AS BACKUP . NO DELAY OR PATIENT INJURY WAS REPORTED. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED ACCORDINGLY.

    Patient problems: Injury; No Clinical Signs, Symptoms or Conditions

    MAUDE report 10577890 ↗
  • MalfunctionSep 4, 2020Mechanical Jam; Mechanical Jam

    IT WAS REPORTED THAT DURING THE PROCEDURE USING AN ACCU-PASS 45-DEGREE, WHEN THE NEEDLE WAS INSERTED INTO THE MENISCUS, THE MONOFILAMENT WOULD NOT COME OUT; THEREFORE, IT WAS NOT ABLE TO BE USED. THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITHOUT DELAY USING A BACK-UP DEVICE. NO PATIENT INJURY OR OTHER COMPLICATIONS WERE REPORTED. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED ACCORDINGLY.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 10497512 ↗
  • MalfunctionJun 30, 2020Failure to Eject; Failure to Eject

    IT WAS REPORTED THAT DURING SURGERY, THE MONOFILAMENT WOULD NOT COME OUT OF THE ACCU-PASS HENCE, IT COULD NOT BE USED. THE PROCEDURE WAS COMPLETED WITHOUT DELAY USING A S+N BACKUP DEVICE. NO PATIENT INJURY OR OTHER COMPLICATIONS WERE REPORTED. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED ACCORDINGLY.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 10212262 ↗

Other ORTHOPEDIC MANUAL SURGICAL INSTRUMENT devices

All ORTHOPEDIC MANUAL SURGICAL INSTRUMENT devices →

Sources and updates