DYONICS POWERMAX ELITE 72200616Recalls & reported problems
Smith & Nephew, Inc. · DYONICS POWERMAX ELITE HANDPIECE WITH BLADE MULTI-POSITIONING, HAND CONTROLS, SUCTION LEVER, 10 FT 3 M CORD · Class 1
All models of DYONICS POWERMAX ELITE (2)
Data updated Nov 27, 2026 from official FDA records · Editorial policy
Quick summary
This Smith & Nephew Dyonics PowerMAX Elite is a powered surgical saw handpiece with accessories. Its 4,304 adverse event reports mostly describe overheating, mechanical jams and unusual noise, and some describe burns. The reports include no deaths and 31 injuries. No recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 4,288
- Deaths reported
- 0
- Injuries reported
- 31
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 4,288 adverse event reports that list this exact device (DI 03596010555472), received between Aug 14, 2017 and Jun 30, 2026 — classified as 31 injuries, 4,257 malfunctions.
The most frequently reported specific device problem is “Overheating of Device” (2,519 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Most reported problems
Device problems
- Overheating of Device2,519
- Mechanical Jam1,391
- Mechanical Problem367
- Protective Measures Problem254
- Insufficient Device Problem Information172
- Noise, Audible150
- Electrical Shorting64
- Physical Resistance/Sticking53
Patient problems
- No Clinical Signs, Symptoms or Conditions2,895
- No Patient Involvement635
- No Known Impact Or Consequence To Patient633
- No Information93
- No Consequences Or Impact To Patient28
- Burn(s)7
- Insufficient Information5
- Partial thickness (Second Degree) Burn4
Recent adverse event reports
The 25 most recent of 4,288 reports, in the reporter's own words as published by the FDA.
MalfunctionJun 30, 2026Overheating of Device
IT WAS REPORTED THAT ONE (1) MOTOR DRIVE UNIT, HAND CNTRL, PWRMX EL WAS RUNNING HOT. IT IS UNKNOWN WHETHER THIS PROBLEM WAS NOTICED IN A SURGICAL ENVIRONMENT OR NOT; THEREFORE, PATIENT INVOLVEMENT HAS NOT BEEN CONFIRMED. HOWEVER, THERE WAS NO DELAY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25662320 ↗MalfunctionJun 28, 2026Overheating of Device
IT WAS REPORTED THAT, MDU WAS OVERHEATING WHEN USED. AS THIS WAS NOTICED UPON SERVICE AND REPAIR ACTIVITIES, THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25648552 ↗MalfunctionJun 1, 2026Overheating of Device
IT WAS REPORTED THAT, BEFORE A SURGERY, THE MDU WAS OVERHEATING. THE PROCEDURE WAS PERFORMED, WITHOUT ANY DELAY, USING AN EQUIVALENT S+N BACKUP. NO FURTHER COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25356468 ↗MalfunctionJun 1, 2026Mechanical Jam
IT WAS REPORTED THAT, BEFORE A KNEE ARTHROSCOPY, IT WAS NOTICED THAT A MOTOR DRIVE UNIT MALFUNCTIONED IMMEDIATELY UPON STARTING AND THERE WAS NO SOLUTION. TESTS WERE PERFORMED AND THE DEVICE WAS RESTARTED MULTIPLE TIMES BUT NOTHING WORKED. THE CONTROLLER DISPLAYED "BLADE JAMMED". THE PROCEDURE WAS PERFORMED, AFTER A DELAY GREATER THAN 30 MINUTES, USING THE BASKET AND THE 90° HANDPIECE. NO FURTHER COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25352869 ↗MalfunctionMay 31, 2026Overheating of Device
IT WAS REPORTED THAT THE MDU WAS BROKEN, GOT WARM. IT IS UNKNOWN WHETHER THIS PROBLEM WAS NOTICED IN A SURGICAL ENVIRONMENT OR NOT; THEREFORE, PATIENT INVOLVEMENT HAS NOT BEEN CONFIRMED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25341660 ↗MalfunctionMay 27, 2026Overheating of Device
IT WAS REPORTED THAT, AN UNKNOWN HANDPIECE WAS FAULTY AND GOT WARM WHEN TRYING TO USE IT. AS THIS WAS NOTICED UPON INSPECTION, THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25311092 ↗MalfunctionMay 26, 2026Overheating of Device
IT WAS REPORTED THAT, THERE WAS ABNORMAL HEATING OBSERVED ON A MOTOR DRIVE UNIT. AS THIS WAS NOTICED UPON ROUTINE INSPECTION, THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25295337 ↗MalfunctionMay 18, 2026Overheating of Device
IT WAS REPORTED THAT THE MDU WAS GETTING HOT. IT IS UNKNOWN WHETHER THIS PROBLEM WAS NOTICED IN A SURGICAL ENVIRONMENT OR NOT; THEREFORE, PATIENT INVOLVEMENT HAS NOT BEEN CONFIRMED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25202264 ↗MalfunctionMay 15, 2026Activation Failure
IT WAS REPORTED THAT, DURING A ROTATOR CUFF SURGERY, ONE (1) MDU WAS STOPPED WORKING. THE PROCEDURE WAS RESUMED, AFTER A 40 MINUTE DELAY, USING AN EQUIVALENT S+N BACK-UP. ADDITIONAL ANESTHESIA WAS NOT REQUIRED AS CONSEQUENCE OF THE SURGICAL PROLONGATION. NO FURTHER COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25185174 ↗MalfunctionMay 14, 2026Electrical /Electronic Property Problem; Overheating of Device; Mechanical Jam
IT WAS REPORTED THAT, DURING SURGERY, AN MDU OVERHEATED. THE PROCEDURE WAS RESUMED, WITHOUT ANY DELAY, USING AN EQUIVALENT S+N BACK UP PRODUCT. NO FURTHER COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25175290 ↗MalfunctionMay 11, 2026Mechanical Problem; Overheating of Device; Mechanical Jam; Noise, Audible
IT WAS REPORTED THAT, BEFORE A ROTATOR CUFF REPAIR, IT WAS NOTICED THAT ONE (1) MOTOR DRIVE UNIT WAS NOT FUNCTIONING CORRECTLY. THE MDU FREQUENTLY JAMMED, EMITTED AN UNUSUAL NOISE AS IF THE MOTOR WERE BEING STRAINED, DID NOT OSCILLATE CORRECTLY, AND OVERHEATED. THE PROCEDURE WAS PERFORMED, AFTER 1-HOUR DELAY, USING AN EQUIVALENT S+N BACK-UP. ADDITIONAL ANESTHESIA WAS REQUIRED AS CONSEQUENCE OF THE SURGICAL PROLONGATION. NO FURTHER COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25145213 ↗MalfunctionMay 11, 2026Electrical /Electronic Property Problem; Mechanical Problem; Overheating of Device
IT WAS REPORTED THAT THE MDU WAS OVERHEATING. IT IS UNKNOWN WHETHER THIS PROBLEM WAS NOTICED IN A SURGICAL ENVIRONMENT OR NOT; THEREFORE, PATIENT INVOLVEMENT HAS NOT BEEN CONFIRMED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25142742 ↗MalfunctionMay 11, 2026Overheating of Device
IT WAS REPORTED THAT THE MDU BECAME VERY HOT DURING USE. IT IS UNKNOWN WHETHER THIS PROBLEM WAS NOTICED IN A SURGICAL ENVIRONMENT OR NOT; THEREFORE, PATIENT INVOLVEMENT HAS NOT BEEN CONFIRMED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25136304 ↗MalfunctionMay 5, 2026Electrical /Electronic Property Problem; Overheating of Device; Mechanical Jam; Unexpected Shutdown
IT WAS REPORTED THAT THE MDU WENT DOWN AND WAS OVER HEATING. AS THIS WAS NOTICED UPON INSPECTION, THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25089768 ↗MalfunctionMay 3, 2026Device Alarm System; Overheating of Device
IT WAS REPORTED THAT BEFORE A MENISECTOMY, WHEN CONNECTING THE MDU TO THE CONSOLE, IT BEGAN TO HEAT UP AND SUBSEQUENTLY TRIGGERED A HANDPIECE MALFUNCTION ALARM. IT IS UNKNOWN HOW THE PROCEDURE WAS PERFORMED; HOWEVER, THERE WAS NO DELAY. NO FURTHER COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25063072 ↗MalfunctionApr 29, 2026Electrical /Electronic Property Problem; Mechanical Problem; Overheating of Device
IT WAS REPORTED THAT, BEFORE A SURGERY, THE MDU OVERHEATED. THE PROCEDURE WAS PERFORMED, WITHOUT ANY DELAY, USING AN EQUIVALENT S+N BACKUP. NO FURTHER COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25024316 ↗MalfunctionApr 19, 2026Overheating of Device
IT WAS REPORTED THAT BEFORE A KNEE SCOPE PROCEDURE, ONE DYONICS POWERMAX ELITE HANDPIECE HEATED UP AND STOPPED WORKING. THE PROCEDURE WAS PERFORMED, WITHOUT ANY DELAY, USING AN EQUIVALENT SN BACK UP DEVICE. NO FURTHER COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24921351 ↗MalfunctionApr 16, 2026Overheating of Device; Mechanical Jam
IT WAS REPORTED THAT, BEFORE A SHOULDER SURGERY, IT WAS NOTICED THAT ONE (1) MDU OVERHEATED AND DID NOT OSCILLATE. THE PROCEDURE WAS PERFORMED, WITHOUT ANY DELAY, USING AN EQUIVALENT S+N BACK-UP. NO FURTHER COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24905027 ↗MalfunctionApr 13, 2026Overheating of Device; Mechanical Jam
IT WAS REPORTED THAT, BEFORE A SHOULDER SURGERY, ONE (1) MOTOR DRIVE UNIT OVERHEATED AND DID NOT OSCILLATE. THE PROCEDURE WAS PERFORMED, WITHOUT ANY DELAY, USING AN EQUIVALENT S+N BACK-UP. NO FURTHER COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24860624 ↗MalfunctionApr 8, 2026Overheating of Device; Electrical Shorting; Activation Failure; Noise, Audible
IT WAS REPORTED THAT DURING A MENISCUS REPAIR SURGERY, THE MDU EMITTED A LOUD NOISE AND BECAME HOT WHEN CONNECTED AND POWERED ON. WHEN ATTEMPTED TO BE USED AGAIN, THE DYONICS MACHINE DISPLAYED A ¿SHORT CIRCUIT DETECTED¿ ERROR MESSAGE, AND THE MDU COULD NOT BE POWERED ON. THE PROCEDURE WAS RESUMED, AFTER A DELAY GREATER THAN 30 MINUTES, USING A COMPETITOR DEVICE. NO FURTHER COMPLICATIONS WERE REPORTED. THE PATIENT IS STABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24817614 ↗MalfunctionApr 5, 2026Overheating of Device; Mechanical Jam
IT WAS REPORTED THAT, BEFORE A SURGERY, IT WAS NOTICED THAT THE MOTOR DRIVE UNIT WAS OVERHEATING. THE PROCEDURE WAS PERFORMED, WITHOUT ANY DELAY, USING A SIMILAR SN BACK UP DEVICE. NO FURTHER COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24786753 ↗MalfunctionApr 3, 2026Overheating of Device
IT WAS REPORTED THAT AN MDU BECAME VERY HOT. IT IS UNKNOWN WHETHER THIS PROBLEM WAS NOTICED IN A SURGICAL ENVIRONMENT OR NOT; THEREFORE, PATIENT INVOLVEMENT HAS NOT BEEN CONFIRMED. NO FURTHER INFORMATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24776554 ↗MalfunctionApr 1, 2026Melted; Overheating of Device
IT WAS REPORTED THAT DURING A SHOULDER SURGERY, WHEN THE RPM OF ONE (1) MOTOR DRIVE UNIT WAS INCREASED TO 8,000¿10,000, HEAT BUILT UP INSIDE AND THE PLASTIC ON THE SHAVER BLADE MELTED. THE PROCEDURE WAS RESUMED, WITHOUT ANY DELAY, USING A SIMILAR S+N BACK-UP PRODUCT. NO FURTHER COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24754933 ↗MalfunctionMar 31, 2026Overheating of Device
IT WAS REPORTED THAT, DURING A KNEE ARTHROSCOPY, A MOTOR DRIVE UNIT WOULD HEAT UP. THE PROCEDURE WAS RESUMED, AFTER A LESS THAN 5-MINUTE DELAY, USING AN EQUIVALENT S+N BACK-UP. NO FURTHER COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24742499 ↗MalfunctionMar 30, 2026Overheating of Device; Unexpected Shutdown
IT WAS REPORTED THAT, DURING A SHOULDER ARTHROSCOPIC PROCEDURE, THE MDU STOPPED WORKING AND THE SURGEON COMPLAINED THAT THE DEVICE WAS OVERHEATING AND BURNING HIS FINGERS. THE PROCEDURE WAS RESUMED, WITH A NON SIGNIFICANT DELAY, USING AN EQUIVALENT S+N BACK-UP. NO FURTHER COMPLICATIONS WERE REPORTED.
Patient problems: Burn(s)
MAUDE report 24737092 ↗
FDA clearance and classification
The FDA regulates product code HAB (saw, powered, and accessories) under 21 CFR 878.4820 as a Class I device. 21 CFR 878.4820 ↗
Devices of this type are generally exempt from premarket notification (510(k)), so a model can be sold without an FDA clearance of its own; it still has to meet FDA registration, labeling and reporting rules.
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Common questions about the DYONICS POWERMAX ELITE 72200616
Has the DYONICS POWERMAX ELITE 72200616 been recalled?
No FDA recall record in our data names this exact device identifier (DI 03596010555472), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.
What problems have been reported with the DYONICS POWERMAX ELITE 72200616?
The FDA's MAUDE database lists 4,288 adverse event reports for this device, received between Aug 14, 2017 and Jun 30, 2026. In those reports, the most common specific device problems are “Overheating of Device” (2,519), “Mechanical Jam” (1,391) and “Mechanical Problem” (367); the most common patient problems are “Burn(s)” (7) and “Partial thickness (Second Degree) Burn” (4). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the DYONICS POWERMAX ELITE 72200616?
None of the reports are classified as deaths, 31 as injuries and 4,257 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the DYONICS POWERMAX ELITE 72200616 FDA approved?
Devices of this type (product code HAB) are generally exempt from FDA premarket notification, so this model has no FDA clearance number of its own. It is listed in the FDA's device database (GUDID) by Smith & Nephew, Inc.
Who makes the DYONICS POWERMAX ELITE 72200616, and is it still sold?
The FDA's device database lists Smith & Nephew, Inc. as the company responsible for it (brand DYONICS POWERMAX ELITE, model 72200616). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other SAW, POWERED, AND ACCESSORIES devices
- NA 72200616SSmith & Nephew, Inc. · SAW, POWERED, AND ACCESSORIES486 reports
- NA 72201500Smith & Nephew, Inc. · SAW, POWERED, AND ACCESSORIES136 reports
- PLATINUM 72205321Smith & Nephew, Inc. · SAW, POWERED, AND ACCESSORIES117 reports
- Zimmer Biomet™ 89-8509-450-60Zimmer Surgical SA · Saw, Powered, And Accessories83 reports
- NA 72200617Smith & Nephew, Inc. · SAW, POWERED, AND ACCESSORIES77 reports
- NA 72202530Smith & Nephew, Inc. · SAW, POWERED, AND ACCESSORIES53 reports
- DYONICS INCISOR 72203013Smith & Nephew, Inc. · SAW, POWERED, AND ACCESSORIES46 reports
- NA 72203521Smith & Nephew, Inc. · SAW, POWERED, AND ACCESSORIES40 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.