TRIGEN 71674075Recalls & reported problems
Smith & Nephew, Inc. · HONEYCOMB · Class 1
Quick summary
This Smith & Nephew Honeycomb instrument is a surgical tool used to guide pins and wires during orthopedic bone fixation surgery. Reports mostly describe the plastic tip or body breaking, cracking or snapping — often noticed during surgery setup or while inserting it into the knee, and in most cases surgery continued with a backup instrument without harm to the patient. There are 36 reports, no deaths and no injuries, and no recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 36
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 36 adverse event reports that list this exact device (DI 03596010560711), received between Nov 16, 2018 and Jul 30, 2024 — classified as 36 malfunctions.
The most frequently reported specific device problem is “Break” (25 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Break25
- Crack6
- Flaked2
- Fracture2
- Material Separation1
Patient problems
- No Clinical Signs, Symptoms or Conditions27
- No Consequences Or Impact To Patient4
- No Patient Involvement3
- No Known Impact Or Consequence To Patient2
Recent adverse event reports
The 25 most recent of 36 reports, in the reporter's own words as published by the FDA.
MalfunctionJul 30, 2024Break; Break
IT WAS REPORTED THAT DURING AN INTERTAN PROCEDURE, THE HONEYCOMB BROKE AT THE MOMENT IT WAS INTRODUCED INTO THE KNEE. NO PIECES FELL INTO PATIENT. SURGERY WAS PERFORMED, WITHOUT ANY DELAY, WITH A BACK-UP DEVICE. PATIENT'S CURRENT HEALTH STATUS IS STABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19862006 ↗MalfunctionApr 19, 2023Fracture; Fracture
IT WAS REPORTED THAT DURING SURGERY, A HONEYCOMB WAS USED TO REROUTE THE GUIDE PIN FOR A TIBIA NAIL. IN THE SOFT TISSUE SLEEVE, THE HONEYCOMB BROKE IN HALF INTO TWO PARTS AND HAD TO BE REMOVED WITH A COCHAR. THE PROCEDURE WAS RESUMED, WITHOUT ANY DELAY, USING A S+N BACK-UP DEVICE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16771829 ↗MalfunctionSep 14, 2022Fracture; Fracture
IT WAS REPORTED THAT THE TIP OF A HONEYCOMB WAS FOUND FRACTURED. IT IS UNKNOWN WHETHER THIS PROBLEM WAS NOTICED IN A SURGICAL ENVIRONMENT OR NOT; THEREFORE, PATIENT INVOLVEMENT HAS NOT BEEN CONFIRMED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15422999 ↗MalfunctionJun 11, 2022Break; Break
IT WAS REPORTED THAT FOLLOWING STERILIZATION, THAT TIP OF THE HONEYCOMB WAS BROKEN AND MISSING NO CASE INVOLVED; THEREFORE, THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14671116 ↗MalfunctionFeb 2, 2022Crack; Crack
IT WAS REPORTED THAT DURING SET UP FOR SURGERY, THE END OF THE HONEYCOMB HAS CRACKED OFF. THE PROCEDURE WAS COMPLETED, WITHOUT DELAY, USING A S+N BACK-UP DEVICE. PATIENT WAS NOT HARMED AS CONSEQUENCE OF THIS PROBLEM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13429760 ↗MalfunctionJan 10, 2022Break; Break
IT WAS REPORTED THAT THE TIP OF A HONEYCOMB IS BROKEN. IT IS UNKNOWN WHETHER THIS PROBLEM WAS NOTICED IN A SURGICAL ENVIRONMENT OR NOT; THEREFORE, PATIENT INVOLVEMENT HAS NOT BEEN CONFIRMED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13213923 ↗MalfunctionDec 30, 2021Break; Break
IT WAS REPORTED THAT, DURING THE SET UP FOR AN INTERNAL FIXATION SURGERY, THE PLASTIC END OF A HONEYCOMB WAS NOTICED TO BE BROKEN. SURGERY WAS PERFORMED, WITHOUT ANY DELAY, WITH A BACK-UP DEVICE INSTEAD. PATIENT WAS NOT INJURED AS CONSEQUENCE OF THIS PROBLEM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13118973 ↗MalfunctionNov 16, 2021Material Separation; Material Separation
IT WAS REPORTED THAT, DURING SET UP FOR SURGERY, THE HONEYCOMB CAME APART. AS THIS HAPPENED BEFORE SURGERY, THERE WAS NO HARM TO PATIENT HOWEVER, THE ISSUE RESULTED ON A CHANGE IN SURGICAL TECHNIQUE. NO OTHER COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12820190 ↗MalfunctionNov 15, 2021Break; Break
IT WAS REPORTED THAT A 130 RAD DRILL GDE DROP HAS A MISSING PIN ((B)(4)), THE HEX END OF A TRIGEN SHORT HEXDRIVER IS ROUNDED ((B)(4)), THE TIP OF A HONEYCOMB IS BROKEN ((B)(4)), A TRIGEN RULER HAS BROKEN THREADS ((B)(4)), AND A LID FOR OUTER CASES HAS BROKEN CLASP ((B)(4)). AS THIS WAS NOTICED UPON FIELD INSPECTION, THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12817857 ↗MalfunctionOct 13, 2021Flaked; Flaked
IT WAS REPORTED THAT, DURING AN INTERNAL FIXATION SURGERY, A HONEYCOMB WAS CHIPPED AT THE END. THIS HAPPENED BEFORE USE IN PATIENT. SURGERY WAS PERFORMED, WITHOUT ANY DELAY, WITH A BACK-UP DEVICE INSTEAD. PATIENT WAS NOT INJURED AS CONSEQUENCE OF THIS PROBLEM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12623834 ↗MalfunctionOct 1, 2021Flaked; Flaked
IT WAS REPORTED THAT, DURING AN INTERNAL FIXATION SURGERY, A HONEYCOMB WAS CHIPPED AT THE END. THIS HAPPENED BEFORE USE IN PATIENT. SURGERY WAS PERFORMED, WITHOUT ANY DELAY, WITH A BACK-UP DEVICE INSTEAD. PATIENT WAS NOT INJURED AS CONSEQUENCE OF THIS PROBLEM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12565444 ↗MalfunctionSep 20, 2021Crack; Crack
IT WAS REPORTED THAT, THE LISTED DEVICES WERE CRACKED OR BROKEN. AS THIS WAS NOTICED UPON FIELD INSPECTION, THERE WAS NOT PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12496920 ↗MalfunctionSep 20, 2021Crack; Crack
IT WAS REPORTED THAT, THE LISTED DEVICES WERE CRACKED OR BROKEN. AS THIS WAS NOTICED UPON FIELD INSPECTION, THERE WAS NOT PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12496865 ↗MalfunctionSep 20, 2021Crack; Crack
IT WAS REPORTED THAT, THE LISTED DEVICES WERE CRACKED OR BROKEN. AS THIS WAS NOTICED UPON FIELD INSPECTION, THERE WAS NOT PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12496862 ↗MalfunctionJul 29, 2021Crack; Crack
IT WAS REPORTED THAT, UPON FIELD INSPECTION, A FLEXIBLE SHAFT W/CIR CONNECTOR ((B)(4)) AND A HONEYCOMB ((B)(4)) WERE FOUND CRACKED. ALSO, A SHORT HEXDRIVER ((B)(4)) AND A REDUCER ((B)(4)) WERE FOUND WITH THEIR THREADS DAMAGED. AS THIS WAS NOTICED UPON INSPECTION, THERE WAS NOT PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12243741 ↗MalfunctionJul 14, 2021Break; Break
IT WAS REPORTED THAT A HONEYCOMB'S TIP BROKE. IT IS UNKNOWN WHETHER THIS PROBLEM WAS NOTICED IN A SURGICAL ENVIRONMENT OR NOT; THEREFORE, PATIENT INVOLVEMENT IS UNCERTAIN.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12166200 ↗MalfunctionJul 13, 2021Break; Break
IT WAS REPORTED THAT THE TIP OF A TRIGEN HONEYCOMB IS BROKEN OFF. AS THIS WAS NOTICED UPON FIELD INSPECTION, THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12158843 ↗MalfunctionMar 30, 2021Break; Break
IT WAS REPORTED THAT DURING AN INSPECTION IN THE WAREHOUSE IT WAS NOTICED THE FAILURE OF DEFECTIVE DEVICES. THE JOURNEY II BCS ARTICULAR INSERT SIZE 1-2 9 MM LEFT ((B)(4)), THE JOURNEY II BCS ARTICULAR INSERT SIZE 1-2 9 MM RIGHT ((B)(4)), TWO UNIVERSAL INSERT SPACER SIZE 1-2 18 MM ((B)(4)), JOURNEY II CR LOCKING FEMORAL IMPACTOR BUMPER RIGHT ((B)(4)), JOURNEY UNI TIBIA INSERT SIZE 1-2/8 MM ((B)(4)), JOURNEY UNI TIBIA TRIAL SIZE 5-6/8 MM ((B)(4)),JOURNEY UNI TIBIA TRIAL SIZE 3-4/8 MM ((B)(4)), THREE HONEYCOMB ((B)(4)), JOURNEY II BCS ARTICULAR INSERT TRIAL SIZE 1-2 10 MM RIGHT ((B)(4)), THE JOURNEY II BCS ARTICULAR INSERT TRIAL SIZE 1-2 10 MM LEFT ((B)(4)) AND THE JOURNEY II BCS ARTICULAR INSERT TRIAL SIZE 1-2 12 MM LEFT ((B)(4)) WERE NOTICED CRACKED. TWO COMPRESSION SCREW HEX DRIVER ((B)(4)) AND SCREWDRIVER RELEASE HANDLE ((B)(4)) WERE NOTICED BROKEN. NO CASE INVOLVED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11593435 ↗MalfunctionMar 30, 2021Break; Break
IT WAS REPORTED THAT DURING AN INSPECTION IN THE WAREHOUSE IT WAS NOTICED THE FAILURE OF DEFECTIVE DEVICES. THE JOURNEY II BCS ARTICULAR INSERT SIZE 1-2 9 MM LEFT ((B)(4)), THE JOURNEY II BCS ARTICULAR INSERT SIZE 1-2 9 MM RIGHT ((B)(4)), TWO UNIVERSAL INSERT SPACER SIZE 1-2 18 MM ((B)(4)), JOURNEY II CR LOCKING FEMORAL IMPACTOR BUMPER RIGHT ((B)(4)), JOURNEY UNI TIBIA INSERT SIZE 1-2/8 MM ((B)(4)), JOURNEY UNI TIBIA TRIAL SIZE 5-6/8 MM ((B)(4)),JOURNEY UNI TIBIA TRIAL SIZE 3-4/8 MM ((B)(4)), THREE HONEYCOMB ((B)(4)), JOURNEY II BCS ARTICULAR INSERT TRIAL SIZE 1-2 10 MM RIGHT ((B)(4)), THE JOURNEY II BCS ARTICULAR INSERT TRIAL SIZE 1-2 10 MM LEFT ((B)(4)) AND THE JOURNEY II BCS ARTICULAR INSERT TRIAL SIZE 1-2 12 MM LEFT ((B)(4)) WERE NOTICED CRACKED. TWO COMPRESSION SCREW HEX DRIVER ((B)(4)) AND SCREWDRIVER RELEASE HANDLE ((B)(4)) WERE NOTICED BROKEN. NO CASE INVOLVED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11593413 ↗MalfunctionMar 30, 2021Break; Break
IT WAS REPORTED THAT DURING AN INSPECTION IN THE WAREHOUSE IT WAS NOTICED THE FAILURE OF DEFECTIVE DEVICES. THE JOURNEY II BCS ARTICULAR INSERT SIZE 1-2 9 MM LEFT (B)(4), THE JOURNEY II BCS ARTICULAR INSERT SIZE 1-2 9 MM RIGHT (B)(4), TWO UNIVERSAL INSERT SPACER SIZE 1-2 18 MM (B)(4), JOURNEY II CR LOCKING FEMORAL IMPACTOR BUMPER RIGHT (B)(4), JOURNEY UNI TIBIA INSERT SIZE 1-2/8 MM (B)(4), JOURNEY UNI TIBIA TRIAL SIZE 5-6/8 MM (B)(4), JOURNEY UNI TIBIA TRIAL SIZE 3-4/8 MM (B)(4), THREE HONEYCOMB (B)(4), JOURNEY II BCS ARTICULAR INSERT TRIAL SIZE 1-2 10 MM RIGHT (B)(4), THE JOURNEY II BCS ARTICULAR INSERT TRIAL SIZE 1-2 10 MM LEFT (B)(4) AND THE JOURNEY II BCS ARTICULAR INSERT TRIAL SIZE 1-2 12 MM LEFT (B)(4) WERE NOTICED CRACKED. TWO COMPRESSION SCREW HEX DRIVER (B)(4) AND SCREWDRIVER RELEASE HANDLE (B)(4) WERE NOTICED BROKEN. NO CASE INVOLVED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11593404 ↗MalfunctionMar 30, 2021Break; Break
IT WAS REPORTED THAT DURING AN INSPECTION IN THE WAREHOUSE IT WAS NOTICED THE FAILURE OF DEFECTIVE DEVICES. IT WAS NOTICED THAT THE NAIL PROX DROP ((B)(4)) HAS A DEFECTIVE ATTACHMENT. FOUR GENESIS II TIBIAL ALIGNMENT SPIKED FIXATION ROD ((B)(4)), THE SCREWDRIVER RELEASE HANDLE ((B)(4)), THE JOURNEY II BCS ARTICULAR INSERT TRIAL SIZE 5-6 9MM RIGHT ((B)(4)), THE JOURNEY II BCS ARTICULAR INSERT TRIAL SIZE 5-6 9MM LEFT ((B)(4)), R3 0 DEG 32 MM TRIAL 48 MM ((B)(4)), THE HONEYCOMB ((B)(4)), THE BOX CHISEL ((B)(4)), JOURNEY II BCS ARTICULAR INSERT TRIAL SIZE 1-2 11 MM RIGHT ((B)(4)), THE LAG SCREW 3.2 GUIDE PIN SLEEVE ((B)(4)), THE MEDIUM HEXDRIVER ((B)(4)), THE SUBTROC LAG SCREW DRIVER ((B)(4)), THE ACCORD DUAL ENDED HEX DRIVER ((B)(4)) TWO THIN OSTEO BLADE 8X3 ((B)(4)), THE THIN OSTEO BLADE 12X3 ((B)(4)) AND THE JOURNEY DCF AP FEM CUT BLK 4 ((B)(4)) WERE NOTICED CRACKED. THE NAIL PROX DROP ((B)(4)), THE META ANTERIOR DROP ((B)(4)), THE MI Z HANDLE ACET REAMER ((B)(4)) AND TWO THIN OSTEO BLADE 8X3 ((B)(4)) WERE NOTICED BROKEN. NO CASE INVOLVED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11593377 ↗MalfunctionFeb 11, 2021Break; Break
IT WAS REPORTED THAT THE HONEYCOMB HAS CRACKED. THE EVENT WAS NOTICED BY THE STERILE STAFF, AFTER SURGERY, NO SURGICAL DELAY OR INJURY TO THE PATIENT WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11312296 ↗MalfunctionJan 19, 2021Break; Break
IT WAS REPORTED THAT A HONEYCOMB WAS FOUND TO BE BROKEN. IT IS UNKNOWN WHEN THIS MALFUNCTION OCCURRED AND IF THERE WAS A DELAY OR A BACK UP DEVICE. NO OTHER COMPLICATIONS WERE REPORTED AT THIS TIME.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11194125 ↗MalfunctionJan 8, 2021Break; Break
IT WAS REPORTED THAT, THE RETAINING CLIP FOR A HONEYCOMB CAME OFF, THE TWO PIECES DO NOT HOLD TOGETHER. INCIDENT OCCURRED DURING TREATMENT WITH INSTRUMENTS OUTSIDE THE PATIENT. THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITHOUT DELAY USING A BACK-UP DEVICE. NO PATIENT INJURY OR OTHER COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11141689 ↗MalfunctionDec 31, 2020Break; Break
IT WAS REPORTED THAT, A HONEYCOMB BROKE IN HALF. INCIDENT OCCURRED DURING EFIP INSPECTION FAILURE. NO CASE REPORTED; THEREFORE, THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11103267 ↗
Other ORTHOPEDIC MANUAL SURGICAL INSTRUMENT devices
- AML EXCEL 2001-42-000Depuy (Ireland) · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT2,009 reports
- NA 03.010.523Synthes GmbH · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT1,208 reports
- EXPRESSEW 214141Medos International Sàrl · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT1,166 reports
- N/A 9560524Medtronic Sofamor Danek, Inc. · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT866 reports
- INHANCE 6500-11-104Depuy (Ireland) · Orthopedic manual surgical instrument831 reports
- INHANCE 6500-11-102Depuy (Ireland) · Orthopedic manual surgical instrument567 reports
- R3 71364450Smith & Nephew, Inc. · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT520 reports
- N/A 9561524Medtronic Sofamor Danek, Inc. · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT494 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.