Skip to content

Peri-Loc 71803128Recalls & reported problems

Smith & Nephew, Inc. · PERI-LOC VOLAR DISTAL RADIUS LOCKING PLATE STANDARD HD 9 HOLE SHAFT RIGHT 135MM · Class 2 · Implant

All models of Peri-Loc (102)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 03596010583031), received between Nov 13, 2020 and Nov 13, 2020 — classified as 1 injury.

The most frequently reported specific device problem is “Break” (1 report).

1 other recall cover devices cleared under the same FDA submission (K092015, K210837, K213126); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Most reported problems

Device problems

  • Break1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1
  • No Known Impact Or Consequence To Patient1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • InjuryNov 13, 2020Break; Break

    IT WAS REPORTED THAT, DURING THE OPERATION OF A DISTAL RADIUS COMBINED WITH CADRE FRACTURE, THE STEEL PLATE BROKE DURING SHAPING. PROCEDURE CONTINUED WITH A BACKUP DEVICE FROM SMITH AND NEPHEW. IT IS UNKNOWN IF THERE WAS A DELAY OR THE OUTCOME OF THE PATIENT. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED ACCORDINGLY.

    Patient problems: No Clinical Signs, Symptoms or Conditions; No Known Impact Or Consequence To Patient

    MAUDE report 10836616 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K092015, K210837, K213126. One submission often covers many models, so check the product and lot details in each record.

  • Z-2036-2017 · Mar 22, 2017 · Terminated
    PERI-LOC VOLAR DISTAL RADIUS - RIGHT 3H STD HEAD LOCKING PLATE, 62 MM LENGTH, SST PLATE, REF 71823114, QTY: (1) NON-STERILE, Rx only, Smith & nephew, Inc., Memphis TN 38116

Other PLATE, FIXATION, BONE devices

All PLATE, FIXATION, BONE devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page