Cannulated Screws 71177201Recalls & reported problems
Smith & Nephew, Inc. · 3.2MM X 300MM THREADED GUIDE PIN COCR · Class 2
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 03596010601599), received between Mar 9, 2021 and Mar 9, 2021 — classified as 1 injury.
The most frequently reported specific device problem is “Incorrect Measurement” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Incorrect Measurement1
Patient problems
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
InjuryMar 9, 2021Incorrect Measurement; Incorrect Measurement
IT WAS REPORTED THAT DURING SURGERY FOR NECK OF FEMUR USING SMITH AND NEPHEW CHS SYSTEM, THE SURGICAL TEAM DECIDED TO USE THE ANTI-ROTATION SCREW TREATMENT WITH SMITH AND NEPHEW 7.0MM CANNULATED SCREW SYSTEM. FOLLOWING CORRECT POSITIONING OF THE 3.2MM X 300MM THREADED GUIDE PIN, A MEASUREMENT WAS TAKEN USING THE 6.5/8.0MM DIRECT MEASURING GAUGE. REGISTRAR READ 105MM FROM END OF THE GUIDE PIN AND REQUESTED A 7.0MM X 105MM CANNULATED SCREW 16MM THREAD STAINLESS STEEL TO BE OPENED FOR USE. DURING INSERTION OF THE 105MM SCREW, IT WAS NOTED THAT THE 105MM SCREW WAS APPROX 10MM TOO LONG. A 7.0MM X 95MM CANNULATED SCREW 16MM THREAD WAS THEN REQUESTED AND INSERTED OVER THE GUIDE PIN. THE 95MM SCREW WAS NOTED TO BE OF THE CORRECT LENGTH. FINALLY, THE CHS IMPLANTATION WAS THEN COMPLETED WITH A SURGICAL DELAY OF LESS THAN 30 MINUTES AND NO INJURY TO THE PATIENT. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED ACCORDINGLY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11444216 ↗
Other PLATE, FIXATION, BONE devices
- Lapiplasty System 2 SK14Treace Medical Concepts, Inc. · Plate, Fixation, Bone260 reports
- PRECICE® Plating System LP4.5-PRF4BNuvasive Specialized Orthopedics, Inc. · Plate, Fixation, Bone3 recalls6 reports
- PRECICE® Plating System LP4.5-MLT4BNuvasive Specialized Orthopedics, Inc. · Plate, Fixation, Bone3 recalls5 reports
- PRECICE® Plating System LP4.5-PLF4BNuvasive Specialized Orthopedics, Inc. · Plate, Fixation, Bone3 recalls3 reports
- PRECICE® Plating System LP4.5-MRT4BNuvasive Specialized Orthopedics, Inc. · Plate, Fixation, Bone3 recalls3 reports
- PRECICE® Plating System LP4.5-DRF4BNuvasive Specialized Orthopedics, Inc. · Plate, Fixation, Bone3 recalls1 report
- PRECICE® Plating System LP4.5-PLF40Nuvasive Specialized Orthopedics, Inc. · Plate, Fixation, Bone3 recalls0 reports
- PRECICE® Plating System LP4.5-PRT42Nuvasive Specialized Orthopedics, Inc. · Plate, Fixation, Bone3 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page