BIOSURE 72201768Recalls & reported problems
Smith & Nephew, Inc. · 6 MM X 20 MM BIOSURE HA · Class 2 · Implant
All models of BIOSURE (68)
Data updated Nov 27, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
Quick summary
This Smith & Nephew BIOSURE bone screw is used to anchor ligaments and tendons during knee surgery. Most of the 31 reports describe the screw breaking during insertion or fixation in knee procedures, sometimes leaving fragments that had to be removed, along with a few packaging and labeling problems. No deaths and 12 injuries were reported. One open recall, started in November 2025, covers removal of an affected screw lot over a labeling error.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 30
- Deaths reported
- 0
- Injuries reported
- 12
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 30 adverse event reports that list this exact device (DI 03596010611642), received between Nov 1, 2017 and May 11, 2026 — classified as 12 injuries, 18 malfunctions.
The most frequently reported specific device problem is “Break” (27 reports).
1 FDA recall record name this device's identifier, 1 still open.
1 other recall cover devices cleared under the same FDA submission (K080358); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Recalls naming this device
- Nov 24, 2025Class IIOpen, Classified
Removal of affected lot of screws due to labeling error.
Z-0957-2026 · Root cause: Packaging process control · Recall details · FDA record ↗
Most reported problems
Device problems
- Break27
- Packaging Problem2
- Component Missing1
- Delivered as Unsterile Product1
- Expiration Date Error1
- Nonstandard Device1
- Wrong Label1
Patient problems
- No Clinical Signs, Symptoms or Conditions26
- No Known Impact Or Consequence To Patient2
- Device Embedded In Tissue or Plaque1
- Foreign Body In Patient1
- No Information1
Recent adverse event reports
The 25 most recent of 30 reports, in the reporter's own words as published by the FDA.
MalfunctionMay 11, 2026Break
IT WAS REPORTED THAT DURING AN ACL RECONSTRUCTION PROCEDURE, THE BIOSURE HA SCREW ANCHOR BROKE AT THE VERY BEGINNING OF THE BONE TUNNEL WHILE INSERTING IT TO ANCHOR THE LIGAMENT FOR ANTERIOR REINFORCEMENT. THE FRAGMENT WAS REMOVED USING ARTHROSCOPIC FORCEPS AND KELLY FORCEPS. THE PROCEDURE WAS RESUMED WITHOUT ANY DELAY, USING AN EQUIVALENT S+N BACK UP DEVICE. NO HARM TO THE PATIENT OR FURTHER COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25145680 ↗MalfunctionMar 15, 2026Break
IT WAS REPORTED THAT, DURING AN ACL PROCEDURE, A BIOSURE SCREW BROKE DURING GRAFT FIXATION. THE SCREW WAS REMOVED FROM THE PATIENT, LEAVING NOTHING INSIDE. THE PROCEDURE WAS RESUMED, AFTER A 30-MINUTE DELAY, AS THE SURGEON HAD TO PERFORM A SECOND PASS FOR SCREW FIXATION, USING AN EQUIVALENT S+N BACK-UP. NO ADDITIONAL ANESTHESIA WAS REQUIRED. NO FURTHER COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24601482 ↗MalfunctionNov 14, 2025Break; Nonstandard Device; Break; Nonstandard Device
IT WAS REPORTED THAT DURING A LIGAMENTOPLASTY, WHEN ATTEMPTED TO INSERT AND SCREW THE BIOSURE SCREW INTO THE PATIENT'S KNEE. THE SCREW BROKE INSIDE THE PATIENT. ALL THE BROKEN PARTS WERE REMOVED FROM THE PATIENT. UPON INSPECTION, IT WAS FOUND THAT THIS SCREW DID NOT MATCH THE SCREWS NORMALLY USED (DIFFERENT COMPOSITION). THE SURGERY WAS COMPLETED USING AN EQUIVALENT SN BACK-UP DEVICE. IT IS UNKNOWN IF THERE WAS SURGICAL DELAY. NO FURTHER COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23556652 ↗MalfunctionNov 3, 2025Break; Packaging Problem; Wrong Label; Break; Packaging Problem; Wrong Label
IT WAS REPORTED THAT THE BOX OF A BIOSURE 6MM X 20MM HAD PACKAGED A BIOSURE SCREW 5 X 20MM INSTEAD AND WAS FOUND DURING THE KNEE MEDIAL PATELLA FEMORAL LIGAMENT RECONSTRUCTION, AND ADDITIONALLY, IT BROKE WHEN BEING INSERTED INTO THE PATIENT. THE PROCEDURE WAS RESUMED, AFTER A 2-MINUTE DELAY, USING AN EQUIVALENT S+N BACK-UP IN THE ORIGINALLY DRILLED BONE HOLE. NO FURTHER COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23460400 ↗MalfunctionFeb 25, 2025Break; Break
IT WAS REPORTED THAT DURING AN ACL PROCEDURE WITH MENISCAL SUTURE, THE BIOSURE HA SCREW BROKE AT THE TIME OF FIXATION AS IT WAS RESECTED. THE PROCEDURE WAS COMPLETED WITH A SMITH AND NEPHEW BACK UP DEVICE WITH NO SURGICAL DELAY. NO FURTHER COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21464726 ↗MalfunctionFeb 4, 2025Break; Break
IT WAS REPORTED THAT DURING THE ACL PROCEDURE, THE BIOSURE HA SCREW BROKE DURING FIXATION. THE FRAGMENTS WERE REMOVED FROM INSIDE THE PATIENT. THE PROCEDURE WAS COMPLETED WITH A S+N BACK UP DEVICE. THERE WAS NO SURGICAL DELAY AND NO FURTHER COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21303609 ↗MalfunctionOct 14, 2024Break; Break
IT WAS REPORTED THAT DURING AN ARTHROSCOPIC ACL SURGERY, THE BIOSURE SCREW BROKE WHILE FIXING THE GRAFT. THE BROKEN PIECES WERE COMPLETELY REMOVED FROM THE PATIENT. NO SURGICAL DELAY WAS REPORTED; HOWEVER, IT IS UNKNOWN HOW THE PROCEDURE WAS COMPLETED. NO FURTHER COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20444077 ↗MalfunctionSep 19, 2024Break; Break
IT WAS REPORTED THAT DURING AN ARTHROSCOPY, THE TIP OF THE BIOSURE SCREW GOT BROKEN WHILE SCREWING INTO THE BONE. ALL THE BROKEN PIECES WERE REMOVED FROM THE BONE WITH THE HELP OF TWEEZERS. THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITH A DELAY OF LESS THAN 30 MINUTES USING A SMITH AND NEPHEW BACK UP DEVICE. NO FURTHER COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20257143 ↗MalfunctionMar 21, 2024Break; Break
IT WAS REPORTED THAT DURING AN ARTHROSCOPY, THE BIOSURE SCREW DECRUSTED (CHIPPED OFF) WHEN INSERTING IT INTO THE BONE. THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITHOUT SURGICAL DELAY. IT UNKNOWN IF THERE WAS A SMITH AND NEPHEW BACK UP DEVICE AVAILABLE. NO FURTHER COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18952193 ↗MalfunctionFeb 21, 2024Break; Break
IT WAS REPORTED THAT DURING AN ACL RECONSTRUCTION PROCEDURE, THE BIOSURE TIP BROKE AFTER BEING INSERTED. IT IS UNKNOWN HOW THE PROCEDURE WAS COMPLETED. THERE WAS NO SURGICAL DELAY AND NO FURTHER COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18753304 ↗MalfunctionDec 27, 2023Break; Break
IT WAS REPORTED THAT A PROCEDURE, THE SURGEON DID NOT FOLLOW THE INSTRUCTIONS FOR USE AND PROCEEDED TO USE THE MALE WITHOUT CLEANING THE ROUTE IN THE SOFT PARTS; WHICH RESULTED IN THE BIOSURE HA MAKING A FALSE TRAJECTORY AND ENDED UP BREAKING. THE BROKEN PIECES WERE TOTALLY REMOVED. SURGERY CONTINUED WITHOUT ANY DELAY, WITH A BACK-UP DEVICE. NO FURTHER COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18405727 ↗MalfunctionDec 26, 2023Break; Break
IT WAS REPORTED THAT DURING A MEDIAL COLLATERAL LIGAMENT PROCEDURE, THE TIP OF THE BIOSURE SCREW BROKE AFTER THE DOCTOR APPLIED TOO MUCH FORCE. THE FRAGMENT WAS REMOVED WITH THE HELP OF TWEEZERS. THE PROCEDURE WAS COMPLETED USING A BACK-UP DEVICE. THERE WAS NO SURGICAL DELAY AND NO FURTHER COMPLICATIONS WERE REPORTED
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18394926 ↗MalfunctionDec 6, 2023Delivered as Unsterile Product; Packaging Problem; Delivered as Unsterile Product; Packaging Problem
IT WAS REPORTED THAT DURING SET UP, THE INNER PACKING OF THE BIOSURE HA HAD AN AIR LEAKAGE. THERE WAS A SMITH AND NEPHEW BACK-UP DEVICE AVAILABLE AND NO SURGICAL DELAY WAS REPORTED. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18278240 ↗MalfunctionSep 29, 2023Component Missing; Component Missing
IT WAS REPORTED THAT DURING AN ARTHROSCOPY PROCEDURE, WHEN THE BIOSURE SCREW SEALED PACKAGE WAS OPENED, NO SCREWS WERE FOUND INSIDE. THE PROCEDURE WAS SUCCESSFULLY COMPLETED USING A BACK-UP DEVICE, A DELAY GREATER THAN 30 MINUTES WAS REPORTED. NO FURTHER COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17843904 ↗MalfunctionJun 12, 2023Break; Break
IT WAS REPORTED THAT DURING AN ACL; HOWEVER UNKNOWN IF INTERNAL OR EXTERNAL TO THE PATIENT; A BIOSURE SCREW BROKE. SURGERY RESUMED WITH A BACK-UP DEVICE; HOWEVER IT IS UNKNOWN IF THERE WAS ANY DELAY. NO FURTHER COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17113389 ↗MalfunctionJun 12, 2023Break; Break
IT WAS REPORTED THAT, DURING AN ARTHROSCOPY, THE BIORCI SCREW BROKE. IT IS UNKNOWN IF THERE WAS ANY DELAY, HOWEVER, A BACK-UP DEVICE WAS AVAILABLE. NO FURTHER COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17107865 ↗InjuryFeb 27, 2023Break; Break
IT WAS REPORTED THAT DURING AN ACL RECONSTRUCTION SURGERY, TWO BIOSURE SCREWS BROKE WHILE IMPLANTATION. ALL THE PIECES WERE REMOVED FROM THE PATIENT BY TWEEZERS. THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITH A NON-SIGNIFICANT SURGICAL DELAY USING A BACK-UP DEVICE IN THE ORIGINALLY DRILLED BONE HOLE. NO FURTHER COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16444471 ↗InjuryNov 17, 2022Break; Break
IT WAS REPORTED THAT, DURING AN ANKLE RECONSTRUCTION, THE BIOSURE SCREW BROKE INSIDE THE PATIENT. THE FRACTURE PIECES WERE REMOVED FROM THE PATIENT BY TWEEZERS. THE SURGERY WAS RESUMED AFTER A NON-SIGNIFICANT SURGICAL DELAY USING A BACK-UP DEVICE. NO FURTHER COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15817609 ↗InjuryNov 11, 2022Break; Break
IT WAS REPORTED THAT DURING A RECONSTRUCTION OF ANKLE, THE SCREW BROKE. ALL FRACTURED PIECES WERE REMOVED FROM THE PATIENT BY TWEEZERS. THE ORIGINALLY DRILLED BONE HOLE WAS USED. THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITH NON-SIGNIFICANT DELAY USING A S+N BACK-UP DEVICE. PATIENT STATUS IS GOOD. NO FURTHER COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15785060 ↗InjuryOct 20, 2022Break; Break
IT WAS REPORTED THAT DURING AN ACL PROCEDURE, THE TIP OF THE 6X20MM BIOSURE HA SCREW BROKE WHEN THE DOCTOR WAS PUTTING IT ON THE PATIENT. IT WAS COMPLETELY REMOVED USING ANOTHER SCREW. THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITH NO DELAY USING A BACK-UP DEVICE. NO FURTHER COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15643342 ↗InjurySep 2, 2022Break; Break
IT WAS REPORTED THAT DURING A LIGAMENT RECONSTRUCTION OPERATION, THE BIOSURE SCREW BROKE. THE PIECES WERE RETRIEVED BY TWEEZERS. THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITH A NON-SIGNIFICANT SURGICAL DELAY USING A BACK-UP DEVICE IN THE ORIGINALLY DRILLED BONE HOLE. NO FURTHER COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15352319 ↗InjuryAug 11, 2022Break; Break
IT WAS REPORTED THAT DURING AN ACL PROCEDURE, A BIOSURE SCREW BROKE IN HALF AND PART OF IT WAS LEFT INSIDE THE PATIENT. PROCEDURE GOT FINISHED WITHOUT ANY DELAY WITH A BACK-UP DEVICE. NO FURTHER COMPLICATIONS WERE REPORTED.
Patient problems: Device Embedded In Tissue or Plaque; Foreign Body In Patient
MAUDE report 15215138 ↗InjuryJul 11, 2022Break; Break
IT WAS REPORTED THAT DURING A ACL RECONSTRUCTION SURGERY, THE BIOSURE BROKE. ALL THE BROKEN PIECES WERE REMOVED FROM THE PATIENT BY TWEEZERS. PATIENT BONE QUALITY WAS NORMAL. THE PROCEDURE WAS COMPLETED WITH A S+N BACK UP DEVICE. THERE WAS A NON-SIGNIFICANT DELAY AND NO FURTHER COMPLICATIONS WERE REPORTED. THE BACK UP DEVICE WAS USED IN THE ORIGINAL BONE HOLE, NO VOID WAS LEFT IN THE PATIENT. AN AWL WAS USED TO PREPARE THE INSERTION SITE, THE INSTRUMENT SIZE WAS 6MM AND THE TUNNEL SIZE WAS 6MM. THE INSERTION SITE WAS CLEARED OF ALL DEBRIS. PATIENT STATUS IS GOOD AND THE SURGEON WAS ULTIMATELY SATISFICED WITH THE SURGICAL OUTCOME.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14982204 ↗InjuryJul 4, 2022Break; Break
IT WAS REPORTED THAT, DURING AN UNKNOWN PROCEDURE, THE BIOSURE SCREW BROKE. SURGERY WAS RESUMED WITH A BACK UP DEVICE INSTEAD, BUT IT IS UNKNOWN WHETHER IT CAUSED OR NOT A SURGICAL DELAY. NO FURTHER COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14911471 ↗InjuryMay 17, 2022Break; Break
IT WAS REPORTED THAT DURING AN ACL RECONSTRUCTION PROCEDURE, THE BIOSURE SCREW BROKE. ALL THE BROKEN PIECES WERE REMOVED FROM THE PATIENT WITH TWEEZERS. THE PROCEDURE WAS COMPLETED WITHOUT SURGICAL DELAY USING A BACK-UP DEVICE IN THE ORIGINALLY DRILLED BONE HOLE. NO FURTHER COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14418960 ↗
FDA clearance and classification
Cleared through 510(k) K080358 on Apr 23, 2008 (decision: Substantially Equivalent). Submitted by Smith & Nephew Inc., Endoscopy Div. as “Plla, Ha Screw”. The FDA review took 72 days. FDA record ↗
The FDA regulates product code HWC (screw, fixation, bone) under 21 CFR 888.3040 as a Class II device. 21 CFR 888.3040 ↗
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Recalls under the same FDA submission
These recalls involve devices cleared under K080358. One submission often covers many models, so check the product and lot details in each record.
- Z-1225-2012 · Feb 6, 2012 · TerminatedSmith & Nephew BIOSURE HA Interference Screw, 7 mm x 25 mm Catalog Number: 72201772. The BIOSURE HA Interference Screw is indicated for fixation of bone-tendon-bone or soft tis
Common questions about the BIOSURE 72201768
Has the BIOSURE 72201768 been recalled?
Yes. 1 FDA recall record names this device's identifier, and it is still open. The recall, Z-0957-2026 (Class II), began on Nov 24, 2025: Removal of affected lot of screws due to labeling error. If you use this device, follow the manufacturer's notice and your healthcare provider's advice.
What problems have been reported with the BIOSURE 72201768?
The FDA's MAUDE database lists 30 adverse event reports for this device, received between Nov 1, 2017 and May 11, 2026. In those reports, the most common specific device problems are “Break” (27), “Packaging Problem” (2) and “Component Missing” (1); the most common patient problems are “Device Embedded In Tissue or Plaque” (1) and “Foreign Body In Patient” (1). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the BIOSURE 72201768?
None of the reports are classified as deaths, 12 as injuries and 18 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the BIOSURE 72201768 FDA approved?
It was cleared through 510(k) K080358 on Apr 23, 2008. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.
Is the BIOSURE 72201768 MRI safe?
According to the device's FDA GUDID record, its labeling does not contain MRI safety information. Ask the manufacturer or the radiology team before an MRI.
Who makes the BIOSURE 72201768, and is it still sold?
The FDA's device database lists Smith & Nephew, Inc. as the company responsible for it (brand BIOSURE, model 72201768). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other Screw, fixation, bone devices
- NA 03.043.029Synthes GmbH · Screw, fixation, bone498 reports
- OSTEORAPTOR 72201995Smith & Nephew, Inc. · Screw, fixation, bone248 reports
- Control 360 SK12Treace Medical Concepts, Inc. · Screw, Fixation, Bone195 reports
- HEALIX PEEK ORTHOCORD 222205Medos International Sàrl · Screw, fixation, bone186 reports
- PRECICE® Bone Transport System STP3.5-200Nuvasive Specialized Orthopedics, Inc. · Screw, Fixation, Bone3 recalls0 reports
- PRECICE® Bone Transport System STP3.5-325Nuvasive Specialized Orthopedics, Inc. · Screw, Fixation, Bone3 recalls0 reports
- PRECICE® Bone Transport System STP3.5-550Nuvasive Specialized Orthopedics, Inc. · Screw, Fixation, Bone3 recalls0 reports
- PRECICE® Bone Transport System STP3.5-425Nuvasive Specialized Orthopedics, Inc. · Screw, Fixation, Bone3 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026 · each recall links to the FDA recall notice
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page