Skip to content

NA 72200873SRecalls & reported problems

Smith & Nephew, Inc. · DYONICS POWER II SHAVER SYSTEM CONTROL UNIT. ALL UNITS 100-240 VAC, 50/60 HZ SERVICE REPLACEMENT · Class 2

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
19
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 19 adverse event reports that list this exact device (DI 03596010615411), received between Dec 13, 2017 and May 5, 2026 — classified as 19 malfunctions.

The most frequently reported specific device problem is “Overheating of Device” (4 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 1 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Overheating of Device4
  • Electrical Shorting3
  • Protective Measures Problem3
  • Smoking3
  • Communication or Transmission Problem2
  • Connection Problem2
  • Device Displays Incorrect Message1
  • Device Inoperable1

Patient problems

  • No Clinical Signs, Symptoms or Conditions9
  • No Known Impact Or Consequence To Patient6
  • No Information2
  • No Consequences Or Impact To Patient1
  • No Patient Involvement1

Recent adverse event reports

The 19 most recent of 19 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMay 5, 2026Electrical Shorting

    IT WAS REPORTED THAT, DURING AN ANKLE ARTHROSCOPY, A DYONICS CONTROL UNIT CAUSED THE POWER TO FAIL AND BLEW THE ELECTRICS UP. THE PATIENT HAD TO BE MOVED TO ANOTHER THEATRE FOR THE SURGERY TO CONTINUE. THE PROCEDURE WAS RESUMED, AFTER A DELAY GREATER THAN 30-MIN, USING AN EQUIVALENT S+N BACK-UP. NO FURTHER COMPLICATIONS WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25084953 ↗
  • MalfunctionOct 26, 2022Fire; Fire

    IT WAS REPORTED THAT, DURING SET UP, THE DYONICS POWER SYSTEM WAS PLUGGED INTO A STORZ CAMERA STACK AND IT MADE BANG AND SET FIRE TO THE STACK. A NON-SIGNIFICANT DELAY WAS REPORTED. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15674693 ↗
  • MalfunctionOct 19, 2022Smoking; Smoking

    IT WAS REPORTED THAT DURING SETUP THE DYONICS POWER II STARTED SMOKING. THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITHOUT DELAY USING A SMITH AND NEPHEW BACK UP DEVICE. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15636084 ↗
  • MalfunctionApr 23, 2021Overheating of Device; Overheating of Device

    IT WAS REPORTED THAT, DURING PROCEDURE, THE DYONICS POWER II CONTROL UNIT CAUGHT FIRE. THE PROCEDURE WAS COMPLETED WITH A BACKUP DEVICE WITH NO DELAY. NO PATIENT COMPLICATIONS WERE REPORTED. IT IS UNKNOWN WHAT PROCEDURE WAS BEING PERFORMED. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED ACCORDINGLY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11712814 ↗
  • MalfunctionMar 30, 2021Protective Measures Problem; Protective Measures Problem

    IT WAS REPORTED THAT THE MOTOR DRIVE UNIT HAD A "HANPIECE SENSOR FAULT" ERROR. NO CASE REPORTED; THEREFORE, THERE WAS NO PATIENT INVOLVEMENT. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED ACCORDINGLY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11587087 ↗
  • MalfunctionMar 24, 2021Overheating of Device; Overheating of Device

    IT WAS REPORTED THAT THE DYONICS POWER II CONTROL UNIT WAS HOT. NO CASE REPORTED; THEREFORE THERE WAS NO PATIENT INVOLVED. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED ACCORDINGLY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11560689 ↗
  • MalfunctionMar 16, 2021Communication or Transmission Problem; Communication or Transmission Problem

    IT WAS REPORTED THAT THE DYONICS POWER II CONTROL UNIT HAD AN UNKNOWN HANDPIECE FAILURE WITH PORT A. NO CASE REPORTED; THEREFORE, THERE WAS NO PATIENT INVOLVEMENT. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED ACCORDINGLY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11498846 ↗
  • MalfunctionMar 14, 2021Communication or Transmission Problem; Connection Problem; Communication or Transmission Problem; Connection Problem

    IT WAS REPORTED THAT DURING SET UP OR INSPECTION FOR AN UNKNOWN PROCEDURE THE PORT A IS NOT DRIVING THE HAND PIECE AND THE PORT B IS 90 DEGREES OFF IN THE DYONICS POWER II CONTROL UNIT. THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITHOUT DELAY USING A BACK-UP DEVICE. NO PATIENT INJURY OR OTHER COMPLICATIONS WERE REPORTED. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED ACCORDINGLY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11479024 ↗
  • MalfunctionDec 28, 2020Protective Measures Problem; Protective Measures Problem

    IT WAS REPORTED THAT THE DYONICS POWER HAD A HANDPIECE SENSOR FAULT ERROR- PORT B. NO CASE WAS INVOLVED; THEREFORE, THERE WAS NO PATIENT INVOLVEMENT. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED ACCORDINGLY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11083880 ↗
  • MalfunctionSep 18, 2020Protective Measures Problem; Protective Measures Problem

    IT WAS REPORTED THAT THE DYONICS WAS NOT WORKING. IT IS UNKNOWN IF THERE WAS A DELAY. IT IS UNKNOWN WHETHER THE EVENT HAPPENED DURING SURGERY AND IF THERE WAS PATIENT INVOLVEMENT. NO FURTHER INFORMATION. RESULTS OF INVESTIGATION REVEALED THAT DURING FUNCTIONAL EVALUATION THE DEVICE DISPLAYED A HANDPIECE STALL ERROR WHICH MAKES IT A REPORTABLE EVENT. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED ACCORDINGLY.

    Patient problems: No Information

    MAUDE report 10553324 ↗
  • MalfunctionSep 18, 2020Mechanical Jam; Mechanical Jam

    IT WAS REPORTED THAT THE DYONICS WAS NOT WORKING. IT IS UNKNOWN IF THERE WAS A DELAY. IT IS UNKNOWN WHETHER THE EVENT HAPPENED DURING SURGERY AND IF THERE WAS PATIENT INVOLVEMENT. NO FURTHER INFORMATION. RESULTS OF INVESTIGATION HAVE CONCLUDED THAT THIS UNIT HAD A HAND PIECE STALL ERROR WHICH MAKES IT A REPORTABLE EVENT. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED ACCORDINGLY.

    Patient problems: No Information

    MAUDE report 10553219 ↗
  • MalfunctionAug 21, 2020Smoking; Smoking

    IT WAS REPORTED THAT WHEN THE CONTROL UNIT WAS PLUGGED IN SMOKE CAME OUT OF IT. NO CASE INVOLVED. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED ACCORDINGLY.

    Patient problems: No Patient Involvement

    MAUDE report 10439528 ↗
  • MalfunctionJul 9, 2020Failure to Power Up; Failure to Power Up

    IT WAS REPORTED THAT THE CONTROL UNIT WAS NOT TURNING ON. NO CASE INVOLVED. RESULTS OF INVESTIGATION HAVE CONCLUDED THAT A HAND PIECE STALL ERROR MESSAGE WHICH MAKES IT A REPORTABLE EVENT. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED ACCORDINGLY.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 10251307 ↗
  • MalfunctionMar 14, 2018Device Inoperable

    IT WAS REPORTED THAT THE PORT A WAS NOT WORKING, A SHORT CIRCUIT ERROR MESSAGE OCCURRED. NO USER INJURY WAS REPORTED.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 7337823 ↗
  • MalfunctionJan 30, 2018Electrical Shorting

    IT WAS REPORTED BY CUSTOMER THAT DEVICE PRESENTED SHORT CIRCUIT.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7227002 ↗
  • MalfunctionJan 5, 2018Electrical Shorting

    IT WAS REPORTED THAT THE DEVICE HAS A SHORT CIRCUIT. NO PATIENT INJURY WAS REPORTED.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 7166923 ↗
  • MalfunctionDec 22, 2017Connection Problem

    IT WAS REPORTED THAT BEFORE THE PROCEDURE THE CONNECTION PORT FOR SHAVER OF THE DYONICS POWER CONTROLLER IS HAVING ISSUES. NO PATIENT INJURY REPORTED. SUBSEQUENT INVESTIGATION RESULTS LATER REVEALED A BURNT PIN SOCKET IN PORT A AND PORT B BOTH GUIDE PINS ARE BENT.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 7143997 ↗
  • MalfunctionDec 20, 2017Overheating of Device; Device Displays Incorrect Message

    IT WAS REPORTED THAT THE DEVICE GOT HOT AND A HIGH VOLTAGE MESSAGE OCCURRED. A BACKUP WAS USED TO COMPLETE THE PROCEDURE. NO PATIENT INJURIES WERE REPORTED.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 7136360 ↗
  • MalfunctionDec 13, 2017Overheating of Device; Smoking

    IT WAS REPORTED THAT THE DEVICE WAS OVERHEATING, SHORT CIRCUIT AND SMOKING. NO PATIENT INJURY REPORTED.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 7113045 ↗

Other ARTHROSCOPE devices

All ARTHROSCOPE devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page