OSTEORAPTOR 72201993Recalls & reported problems
Smith & Nephew, Inc. · OSTEORAPTOR 2.3 SUTURE ANCHOR WITH ONE ULTRABRAID No.2 SUTURE COBRAID BLUE · Class 2 · Implant
Data updated Oct 13, 2026 from official FDA records · Editorial policy
Quick summary
This Smith & Nephew OsteoRaptor is a suture anchor used in soft tissue repair. The FDA records hold 96 reports, with no deaths and 35 injuries; many describe the anchor or its suture breaking during surgery, and in many cases a back-up device was used to finish the procedure. No recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 96
- Deaths reported
- 0
- Injuries reported
- 35
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 96 adverse event reports that list this exact device (DI 03596010615541), received between Jul 24, 2017 and May 26, 2026 — classified as 35 injuries, 61 malfunctions.
The most frequently reported specific device problem is “Break” (64 reports).
40 other recalls cover devices cleared under the same FDA submission (K151105); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Break64
- Device Dislodged or Dislocated17
- Fracture5
- Crack3
- Material Deformation3
- Material Separation3
- Detachment of Device or Device Component2
- Material Fragmentation2
Patient problems
- No Clinical Signs, Symptoms or Conditions68
- No Known Impact Or Consequence To Patient16
- Perforation3
- Device Embedded In Tissue or Plaque2
- Foreign Body In Patient2
- No Consequences Or Impact To Patient2
- Rupture2
- Discomfort1
Recent adverse event reports
The 25 most recent of 96 reports, in the reporter's own words as published by the FDA.
MalfunctionMay 26, 2026Break
IT WAS REPORTED THAT, DURING A HIP LABRAL REPAIR, THE OSTEORAPTOR ANCHOR GOT FRACTURED WHILE INSERTING INTO THE BONE. THE SURGERY WAS COMPLETED USING AN EQUIVALENT S&N BACK-UP DEVICE. IT IS UNKNOWN IF THERE WAS SURGICAL DELAY. NO FURTHER COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25279605 ↗MalfunctionMay 18, 2026Fracture
IT WAS REPORTED THAT DURING SURGERY, ONE (1) OSTEORAPTOR ANCHOR WAS FRACTURED. THE PROCEDURE WAS PERFORMED USING AN EQUIVALENT S+N BACK-UP. IT IS UNKNOWN IF THERE WAS ANY SURGICAL DELAY. NO FURTHER COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25209817 ↗MalfunctionMay 5, 2026Break; Device Dislodged or Dislocated
IT WAS REPORTED THAT, DURING A HIP ARTHROSCOPY, THE SUTURE OF ONE (1) OSTEORAPTOR SUTURE ANCHOR BROKE. THE BROKEN ANCHOR WAS COMPLETELY REMOVED FROM THE PATIENT WITH TWEEZERS. THE PROCEDURE WAS RESUMED, AFTER A NON-SIGNIFICANT DELAY, USING AN EQUIVALENT S+N BACK-UP IN THE ORIGINALLY DRILLED BONE HOLE. ADDITIONAL ANESTHESIA WAS NOT REQUIRED AS CONSEQUENCE OF THE SURGICAL PROLONGATION. NO FURTHER COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25083156 ↗MalfunctionApr 14, 2026Break
IT WAS REPORTED THAT, DURING A HIP LABRUM REPAIR, ONE (1) OSTEORAPTOR SUTURE ANCHOR BROKE. THE PROCEDURE WAS RESUMED, USING AN EQUIVALENT S+N BACKUP; HOWEVER, IT IS UNKNOWN IF THERE WAS A SURGICAL DELAY. NO FURTHER COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24876279 ↗MalfunctionMar 31, 2026No Apparent Adverse Event
IT WAS REPORTED THAT DURING A HIP ARTHROSCOPY, ONE (1) OSTEORAPTOR ANCHOR WAS MISUSED BY THE DOCTOR AND IT WAS REMOVED WITH GRASPER FORCEPS. THE PROCEDURE WAS RESUMED, WITHOUT ANY DELAY, USING AN EQUIVALENT S+N BACK-UP. NO FURTHER COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24750408 ↗MalfunctionMar 23, 2026Break
IT WAS REPORTED THAT, DURING A ROTATOR CUFF REPAIR, AFTER THE INSERTION SITE WAS PREPARED WITH AN ANCHOR MATCHING INSTRUMENT AND CLEARED OF ALL DEBRIS, AN OSTEORAPTOR ANCHOR'S SUTURE BROKE WHEN TYING THE KNOT. THE BROKEN PIECES WERE REMOVED USING FORCEPS. THE PROCEDURE WAS RESUMED, AFTER A NON-SIGNIFICANT DELAY, USING AN EQUIVALENT S+N BACK-UP, IN THE ORIGINALLY DRILLED BONE HOLE. NO ADDITIONAL ANESTHESIA WAS REQUIRED. NO FURTHER COMPLICATIONS WERE REPORTED. THE PATIENT'S STATUS IS FINE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24663083 ↗MalfunctionJan 5, 2026Break
IT WAS REPORTED THAT, DURING A SURGERY, IT WAS NOTICED THAT WHEN ANCHORING AND DELIVERING THE FIRST IMPACT OF THE OSTEORAPTOR, IT BROKE IN HALF, MAKING PLACEMENT UNSUCCESSFUL. THE PROCEDURE WAS RESUMED, AFTER A NON-SIGNIFICANT DELAY, USING A SIMILAR S+N BACK UP DEVICE. NO ADDITIONAL ANESTHESIA WAS REQUIRED. NO FURTHER COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23958126 ↗MalfunctionDec 2, 2025Break; Break
IT WAS REPORTED THAT DURING AN ANKLE JOINT LIGAMENT REPAIR SURGERY, THE OSTEORAPTOR ANCHOR BROKE. THE BROKE PIECES WERE REMOVED FROM INSIDE THE PATIENT USING TWEEZERS AND A GRASPER. THE PROCEDURE WAS RESUMED USING AN EQUIVALENT S+N BACKUP INSERTED INTO THE ORIGINAL BONE HOLE, AFTER A NON-SIGNIFICANT DELAY. NO FURTHER COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23700948 ↗MalfunctionNov 27, 2025Break; Break
IT WAS REPORTED THAT DURING A HIP ARTHROPLASTY PROCEDURE, THE OSTEORAPTOR ANCHOR BROKE DUE TO THE ANGLE OF THE INJURY UPON THE IMPACT. THE BROKEN PIECES WERE REMOVED. THE PROCEDURE WAS RESUMED, WITHOUT ANY DELAY, USING A SIMILAR SN BACK-UP. NO FURTHER COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23665468 ↗MalfunctionSep 30, 2025Break; Break
IT WAS REPORTED THAT DURING A SHOULDER JOINT LABRUM REPAIR SURGERY, IT WAS NOTICED THAT THE OSTEORAPTOR ANCHOR BROKE WHILE BEING IMPLANTED .ALL THE PIECES WERE REMOVED WITH GRASPING FORCEPS AND SUCTION. THE PROCEDURE WAS RESUMED, AFTER A NON-SIGNIFICANT DELAY USING AN EQUIVALENT S+N BACK-UP. THE BACK UP WAS USED IN THE ORIGINALLY DRILLED BONE. THERE WAS NO VOID LEFT IN THE PATIENT AND THE INSTRUMENT TO PREPARE THE INSERTION SITE WAS A 2.3MM ANCHOR MATCHING INSTRUMENT. THE INSERTION SITE WAS CLEARED OF ALL DEBRIS PRIOR TO INSERTION NO FURTHER COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23181064 ↗MalfunctionMay 29, 2025Break; Break
IT WAS REPORTED THAT DURING A SURGERY FOR LIGAMENT INJURY OF THE LEFT ANKLE JOINT, AN OSTEORAPTOR ANCHOR WAS USED, BUT THE ANCHOR BROKE. THE PIECES WERE RETRIEVED WITH TWEEZERS. THE PROCEDURE WAS COMPLETED USING A BACK-UP DEVICE. THE BACK UP DEVICE WAS USED IN THE ORIGINALLY DRILLED HOLE; THERE WERE NO VOIDS LEFT IN THE PATIENT. THERE WAS A DELAY LESS THAN 30 MINUTES AND NO FURTHER COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22102722 ↗MalfunctionMay 20, 2025Fracture; Fracture
IT WAS REPORTED THAT, DURING A HIP JOINT REPAIR SURGERY, THE OSTEORAPTOR ANCHOR BROKE IN THE MIDDLE. SURGERY WAS COMPLETED WITH A NEW ANCHOR OF THE SAME MODEL PLACED IN THE ORIGINALLY DRILLED BONE HOLE, WHICH WAS PREPARED WITH A 2.3 MM INSTRUMENT AND CLEARED OF ALL DEBRIS, NO VOID WAS LEFT IN PATIENT. THE SEVERED PART WAS REMOVED FROM THE BODY, ALTHOUGH IT WAS VERY DIFFICULT. THERE WAS A SURGICAL DELAY OF LESS THAN 30 MINUTES, AND NO FURTHER COMPLICATIONS WERE REPORTED. PATIENT'S CONDITION IS FINE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22053278 ↗MalfunctionMay 15, 2025Break; Break
IT WAS REPORTED THAT DURING A SHOULDER LABRUM REPAIR, THE OSTEORAPTOR 2.3 W. UB COBRAID BLUE ANCHOR BROKE. THE PROCEDURE WAS COMPLETED WITH A SMITH AND NEPHEW BACK UP DEVICE USING THE ORIGINAL DRILLED BONE HOLE AND LEAVING NO VOID IN PATIENT. ALL THE BROKEN PARTS WERE REMOVED USING TWEEZERS. THERE WAS A DELAY LESS THAN 30 MINUTES AND NO FURTHER COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22023917 ↗MalfunctionMay 2, 2025Fracture; Material Deformation; Fracture; Material Deformation
IT WAS REPORTED THAT DURING A SHOULDER PELVIS AND LIP REPAIR, THE GUIDE 2.3 INLINE DRILL AND THE OBTURATOR 2.3 NON-CANNULATED WERE FIRST USED TO ENTER THE JOINT ALONG TO LOCATE THE IDEAL ANCHOR IMPLANTATION POINT, AND THEN PULLED OUT THE OBTURATOR 2.3 NON-CANNULATED. THE SURGEON HELD THE GUIDE 2.3 INLINE DRILL TO FIRMLY LOCATE THE POSITION. THE ASSISTANT USED THE SPECIAL DRILL BIT DRILL FLUTED 2.6MM TO REACH THE DEPTH LIMIT POSITION AND IMPLANTED THE OSTEORAPTOR ANCHOR. WHEN ONE-THIRD OF THE ANCHOR WAS IMPLANTED, IT WAS FOUND THAT THE ANCHOR SPLIT. THE ANCHOR WAS REMOVED. THE PROCEDURE WAS COMPLETED USING A BACK-UP DEVICE. THERE WAS A DELAY LESS THAN 30 MINUTES AND NO FURTHER COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21954317 ↗MalfunctionMar 11, 2025Break; Break
IT WAS REPORTED THAT DURING REPAIR SURGERY FOR TORN SHOULDER AND LABRUM, THE OSTEORAPTOR ANCHOR WAS BROKEN. ALL BROKEN PIECES WERE REMOVED USING TWEEZERS. THE PROCEDURE WAS COMPLETED WITH A SURGICAL DELAY OF LESS THAN 30 MINUTES USING A SMITH AND NEPHEW BACK UP DEVICE, IN THE ORIGINALLY DRILLED BONE HOLE. NO FURTHER COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21574697 ↗MalfunctionFeb 6, 2025Break; Break
IT WAS REPORTED THAT DURING A REPAIR OF THE ANTERIOR TALOFIBULAR LIGAMENT OF THE RIGHT ANKLE, TWO OSTEORAPTOR ANCHOR DEVICES (B)(4) BROKE. ALL BROKEN PIECES WERE REMOVED FROM THE PATIENT BY SUCTION. THE SURGERY WAS COMPLETED USING A BACK-UP DEVICE, WITH A SURGICAL DELAY OF LESS THAN 30 MINUTES. NO FURTHER COMPLICATIONS WERE REPORTED, AND THE PATIENT'S STATUS IS GOOD.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21319468 ↗MalfunctionJan 28, 2025Break; Break
IT WAS REPORTED THAT DURING AN ANKLE JOINT LIGAMENT REPAIR SURGERY, THE OSTEORAPTOR ANCHOR BROKE AND COULD NOT BE USED. THE BROKEN PIECES WERE REMOVED FROM THE INSIDE THE PATIENT USING TWEEZERS. THE PROCEDURE WAS COMPLETED PLACING A S+N BACK UP DEVICE INTO THE ORIGINALLY BONE HOLE. THERE WAS A SURGICAL DELAY OF LESS THAN 30 MINUTES AND NO FURTHER COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21251851 ↗MalfunctionNov 27, 2024Break; Break
IT WAS REPORTED THAT DURING AN ARTHROSCOPY, THE BACK END OF THE OSTEORAPTOR ANCHOR BROKE. THE PROCEDURE WAS COMPLETED USING A BACK-UP DEVICE. THERE WAS A DELAY LESS THAN 30 MINUTES AND NO FURTHER COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20797064 ↗MalfunctionOct 8, 2024Break; Break
IT WAS REPORTED THAT DURING AN ARTHROSCOPY, THE ANCHOR OF THE OSTERAPTOR BROKE. THE PROCEDURE WAS SUCCESSFULLY COMPLETED USING A SMITH AND NEPHEW BACK UP DEVICE. IT IS UNKNOWN IF THERE WAS A DELAY. ALL PIECES WERE REMOVED FROM THE PATIENT USING GRASPERS. THE BACK UP DEVICE WAS USED IN THE ORIGINALLY DRILLED BONE HOLE SO NO VOID WAS LEFT IN THE PATIENT. NO FURTHER COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20402113 ↗MalfunctionSep 18, 2024Break; Device Dislodged or Dislocated; Break; Device Dislodged or Dislocated
IT WAS REPORTED THAT DURING A HIP JOINT LABRUM REPAIR SURGERY, THE OSTEORAPTOR ANCHOR PULLED OUT AFTER BEING IMPLANTED AND WHEN THE SUTURE KNOT WAS TIGHTENED. THE PROCEDURE WAS COMPLETED WITH A SMITH AND NEPHEW BACK UP DEVICE USING THE ORIGINAL DRILLED BONE HOLE WITH A SURGICAL DELAY LESS THAN 30 MINUTES. NO VOID LEFT TO THE PATIENT AND NO FURTHER COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20245826 ↗MalfunctionAug 26, 2024Break; Break
IT WAS REPORTED THAT THE OSTEORAPTOR ANCHOR BROKE. IT IS UNKNOWN WHETHER THE EVENT HAPPENED DURING SURGERY AND IF THERE WAS PATIENT INVOLVEMENT, AND IF THERE WAS A DELAY. THERE WAS A SMITH AND NEPHEW BACK UP DEVICE AVAILABLE. NO FURTHER COMPLICATION WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20071517 ↗MalfunctionAug 7, 2024Device Dislodged or Dislocated; Device Dislodged or Dislocated
IT WAS REPORTED THAT DURING A SHOULDER ARTHROSCOPY, THE OSTERORAPTOR ANCHOR CAME LOOSE FROM THE CABLE. AFTER THE MALFUNCTION THE ANCHOR DID NOT HAVE GOOD FIXATION AND WAS REMOVED. THE PROCEDURE WAS COMPLETED WITH A S+N BACK UP DEVICE. THERE WAS NO SURGICAL DELAY AND NO FURTHER COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19927898 ↗MalfunctionMay 27, 2024Crack; Crack
IT WAS REPORTED THAT, DURING A HIP SURGERY, THE OSTEORAPTOR'S ANCHOR CRACKED WHEN IT WAS HAMMERED. THE THREAD WAS PULLED OUT, AND THE ANCHOR WAS LEFT IN THE PATIENT. FINALLY, THE BROKEN ANCHOR WAS REMOVED WITH SURGICAL TOOL. SURGERY WAS COMPLETED WITH A DELAY OF LESS THAN 30 MINUTES USING A BACK-UP DEVICE IN THE ORIGINALLY DRILLED BONE HOLE. NO FURTHER COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19400955 ↗MalfunctionApr 30, 2024Break; Break
IT WAS REPORTED THAT DURING AN ANTERIOR TALOFIBULAR LIGAMENT OF THE RIGHT ANKLE SURGERY, THE OSTEORAPTOR ANCHOR BROKE. THE PROCEDURE WAS COMPLETED WITH NON-SIGNIFICANT SURGICAL DELAY USING A BACK-UP DEVICE IN THE ORIGINALLY DRILLED BONE HOLE AND NO VOID WAS LEFT IN THE PATIENT AND ALL THE BROKEN PIECES WERE REMOVED FROM THE PATIENT. THE INSERTION SITE WAS CLEARED OF ALL DEBRIS PRIOR TO INSERTION OF THE ANCHOR. NO FURTHER COMPLICATIONS WERE REPORTED. CURRENT STATUS OF PATIENT IS FINE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19214363 ↗MalfunctionApr 23, 2024Material Deformation; Material Deformation
IT WAS REPORTED THAT THE DURING AN UNSPECIFIED ARTHROSCOPY, THE OSTEORAPTOR ANCHOR WAS DEFORMED AFTER IMPLANTATION. THE DEFORMED ANCHOR WAS REMOVED FROM THE PATIENT USING SURGICAL TOOL. NO VOID WAS LEFT IN THE PATIENT. THE PROCEDURE WAS COMPLETED WITH A S+N BACK UP DEVICE IN THE ORIGINALLY BONE HOLE. THERE WAS A DELAY LESS THAN 30 MINUTES AND NO FURTHER COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19161207 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K151105. One submission often covers many models, so check the product and lot details in each record.
- Z-0155-2021 · Sep 9, 2020 · Open, ClassifiedTWINFIX ULTRA 5.5MM PLLA/HA 2 UBWHT/BL--Absorbable Suture anchor Product Number: 72202602
- Z-0143-2021 · Sep 9, 2020 · Open, ClassifiedULTRA FAST-FIX AB ASSEMBLY - STRAIGHT-Absorbable Suture anchors Product Number: 72201493
- Z-0147-2021 · Sep 9, 2020 · Open, ClassifiedRAPTORMITE 3.7MM AB W/ NEEDLES AND TWO-Absorbable Suture anchor Product Number: 72201805
- Z-0150-2021 · Sep 9, 2020 · Open, ClassifiedOSTEORAPTOR 2.3 W. UB COBRAID BLUE-Absorbable Suture anchor Product Number: 72201993
- Z-0138-2021 · Sep 9, 2020 · Open, ClassifiedTWINFIX 5.0MM AB SUTURE ANCHOR-Absorbable Suture anchors Product Number: 7210711
- Z-0151-2021 · Sep 9, 2020 · Open, ClassifiedOSTEORAPTOR 2.9 W/ 1 UB COBRAID BLUE -Absorbable Suture anchor Product Number: 72201994
- Z-0162-2021 · Sep 9, 2020 · Open, ClassifiedTWINFIX ULTRA HA 4.5 W/2 UB BLUE & BLK- Absorbable Suture anchor Product Number: 72202624
- Z-0141-2021 · Sep 9, 2020 · Open, ClassifiedTWINFIX AB 5.0 SUTR ANCHR W/2 38 ULTRA-Absorbable Suture anchors Product Number:72200777
- Z-0153-2021 · Sep 9, 2020 · Open, ClassifiedOSTEORAPTOR 2.9 W. 2 UB WHITE / BLACK -Absorbable Suture anchor Product Number: 72201996
- Z-0165-2021 · Sep 9, 2020 · Open, ClassifiedOSTEORAPTOR CRVD 2.3 SA UB COBRD BLUE- Absorbable Suture anchor Product Number: 72203290
- Z-0156-2021 · Sep 9, 2020 · Open, ClassifiedTWINFIX ULTRA PLLA/HA 5.5 W/3 UB-Absorbable Suture anchor Product Number: 72202603
- Z-0146-2021 · Sep 9, 2020 · Open, ClassifiedBIORAPTOR 2.9MM SUTURE ANCHOR-Absorbable Suture anchor Product Number: 72201702
- Z-0158-2021 · Sep 9, 2020 · Open, ClassifiedTWINFIX ULTRA PLLA/HA 6.5 W/3 UB- Absorbable Suture anchor Product Number: 72202610
- Z-0145-2021 · Sep 9, 2020 · Open, ClassifiedULTRA FAST-FIX AB ASSEMBLY-REVERSE CURVE-Absorbable Suture anchor Product Number: 72201495
- Z-0168-2021 · Sep 9, 2020 · Open, ClassifiedHEALICOIL RSB SA 4.75MM W/1 UT & 1 UB BL-- Absorbable Suture anchor Product Number: 72203704
…and 25 more.
Other FASTENER, FIXATION, BIODEGRADABLE, SOFT TISSUE devices
- GRYPHON ORTHOCORD 210813Medos International Sàrl · FASTENER, FIXATION, BIODEGRADABLE, SOFT TISSUE712 reports
- HEALIX ADVANCE 222330Depuy Mitek, LLC · FASTENER, FIXATION, BIODEGRADABLE, SOFT TISSUE1 recall607 reports
- HEALIX ADVANCE ORTHOCORD 222295Medos International Sàrl · FASTENER, FIXATION, BIODEGRADABLE, SOFT TISSUE640 reports
- HEALICOIL 72205137Smith & Nephew, Inc. · Fastener, fixation, biodegradable, soft tissue427 reports
- HEALIX ADVANCE ORTHOCORD 222296Medos International Sàrl · FASTENER, FIXATION, BIODEGRADABLE, SOFT TISSUE320 reports
- TWINFIX 72202597Smith & Nephew, Inc. · FASTENER, FIXATION, BIODEGRADABLE, SOFT TISSUE288 reports
- Arthrex® AR-2324BCCArthrex, Inc. · FASTENER, FIXATION, BIODEGRADABLE, SOFT TISSUE254 reports
- HEALIX ADVANCE 222331Depuy Mitek, LLC · FASTENER, FIXATION, BIODEGRADABLE, SOFT TISSUE1 recall196 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 13, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 13, 2026
- Page last rebuilt: Oct 13, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page