NA 72202088Recalls & reported problems
Smith & Nephew, Inc. · DIRECT-VIEW ARTHROSCOPE AUTOCLAVABLE, 4.0 MM OUTER DIAMETER X 70 DEGREE DIRECTION OF VIEW, 115 DEGREE FIELD OF VIEW, 160 MM WORKING LENGTH, J-LOCK CANNULA CONNECTION · Class 2
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 13
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 13 adverse event reports that list this exact device (DI 03596010616876), received between Apr 2, 2018 and Nov 28, 2024 — classified as 1 injury, 12 malfunctions.
The most frequently reported specific device problem is “Break” (4 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Break4
- Crack4
- No Display/Image2
- Poor Quality Image2
- Device Inoperable1
- Insufficient Device Problem Information1
- Material Deformation1
- Material Separation1
Patient problems
- No Clinical Signs, Symptoms or Conditions6
- No Known Impact Or Consequence To Patient6
- No Patient Involvement2
- No Information1
Recent adverse event reports
The 13 most recent of 13 reports, in the reporter's own words as published by the FDA.
MalfunctionNov 28, 2024Break; Break
IT WAS REPORTED DURING A HIP ARTHROSCOPY, THE DOCTOR ACCIDENTALLY HIT ARTHROSCOPE OPTICAL WITH ANOTHER TWEEZER BREAKING IT, AND NO PIECES WERE LEFT INSIDE THE PATIENT. THE PROCEDURE WAS COMPLETED WITH NON-SIGNIFICANT SURGICAL DELAY USING A BACK-UP DEVICE. NO FURTHER COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20805693 ↗MalfunctionApr 19, 2021Crack; Crack
IT WAS REPORTED THAT, THE "HD ARTHROSCOPE (4MM X 70°)" WAS BLURRY. NO CASE REPORTED; THEREFORE, THERE WAS NO PATIENT INVOLVEMENT. RESULTS OF INVESTIGATION HAVE CONCLUDED THAT THIS UNIT HAD A CRACKED DISTAL LENS WHICH MAKES IT A REPORTABLE EVENT. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED ACCORDINGLY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11684772 ↗MalfunctionApr 2, 2021Break; Break
IT WAS REPORTED THAT THE SCOPE HAD A BROKEN EXTERNAL LENS, THAT'S ALL THE INFORMATION PROVIDED. NO CASE REPORTED; THEREFORE, THERE WAS NO PATIENT INVOLVEMENT. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED ACCORDINGLY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11611275 ↗MalfunctionMar 30, 2021Crack; Poor Quality Image; Material Deformation; Crack; Poor Quality Image; Material Deformation
IT WAS REPORTED THAT THE A SCOPE ACLAVE HD WAS SCRATCHED/BLURRY. THERE WAS NO HOSPITAL INVOLVED; THEREFORE, THERE WAS NO PATIENT INVOLVEMENT EITHER. PRELIMINARY RESULTS OF INVESTIGATION HAVE CONCLUDED THAT THIS UNIT HAD CRACKED DISTAL LENS WHICH MAKES IT A REPORTABLE EVENT. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED ACCORDINGLY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11588471 ↗MalfunctionOct 20, 2020No Display/Image; No Display/Image
IT WAS REPORTED THAT DURING SET UP THE SCOPE WAS COMPLETELY BLACK AND WILL NOT PROVIDE ANY VIEW. IT IS UNKNOWN HOW WAS THE PROCEDURE COMPLETED AND A DELAY GREATER THAN 30 MINUTES WAS REPORTED. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED ACCORDINGLY.
Patient problems: No Clinical Signs, Symptoms or Conditions; No Known Impact Or Consequence To Patient
MAUDE report 10706133 ↗MalfunctionOct 20, 2020Insufficient Device Problem Information; Insufficient Device Problem Information
IT WAS REPORTED THAT DURING THE SET UP BEFORE A HIP SCOPE, THE SCOPE HAD AN UNSPECIFIED FUNCTIONAL FAILURE. A DELAY GREATER THAN 30 MINUTES WAS REPORTED AND THE PATIENT WAS ALREADY UNDER ANESTHESIA. ANOTHER SMITH AND NEPHEW BACK-UP DEVICE WITH A DIFFERENT PART NUMBER WAS USED TO COMPLETE THE SURGERY. NO OTHER COMPLICATIONS WERE REPORTED. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED ACCORDINGLY.
Patient problems: No Information
MAUDE report 10702885 ↗MalfunctionOct 20, 2020Crack; Poor Quality Image; Optical Obstruction; Crack; Poor Quality Image; Optical Obstruction
IT WAS REPORTED THAT DURING THE SET UP BEFORE A HIP SCOPE, THE SCOPE HAD A CRACK AND THE VISION WAS BLURRY. A DELAY GREATER THAN 30 MINUTES WAS REPORTED AND THE PATIENT WAS ALREADY UNDER ANESTHESIA. ANOTHER SMITH AND NEPHEW BACK-UP DEVICE WITH A DIFFERENT PART NUMBER WAS USED TO COMPLETE THE SURGERY. NO OTHER COMPLICATIONS WERE REPORTED. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED ACCORDINGLY.
Patient problems: No Clinical Signs, Symptoms or Conditions; No Patient Involvement
MAUDE report 10702872 ↗MalfunctionOct 18, 2020Break; Break
IT WAS REPORTED THAT THE DISTAL LENS WERE CRACKED AND LIGHT FIBERS DAMAGED. THE MALFUNCTION HAPPENED DURING SET UP FOR AN ARTHROSCOPY AND WAS SOLVED WITH NO DELAYS OR PATIENT HARM USING A BACK UP DEVICE. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED ACCORDINGLY.
Patient problems: No Patient Involvement
MAUDE report 10694807 ↗InjurySep 3, 2020Crack; Crack
IT WAS REPORTED THAT THE SURGEON FORCEFULLY INSERTED THE SCOPE/CANNULA VIA SLOTTED CANNULA AS A GUIDE. THERE WAS A SLIGHT INTERNAL CRACK OF DISTAL SCOPE LENS. LIKELY, DUE TO THE AMOUNT OF PRESSURE THE SLOTTED CANNULA PLACED ON THE SCOPE LENS. THE MALFUNCTION WAS SOLVED WITH A DELAY SHORTER THAN 30 MINUTES WITH NO PATIENT HARM USING A BACK UP DEVICE AND NO FRAGMENTS REMAIN IN JOINTS. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED ACCORDINGLY.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 10491718 ↗MalfunctionJan 3, 2020Material Separation; Material Separation
IT WAS REPORTED THAT DURING AN ARTHROSCOPY OF THE LEFT JOINT KNEE, SYNOVECTOMY, JOINT CAVITY CLEANING, USE OPPOSITE GRACILIS FEMORIS AND SEMITENDINOSUS TENDON TO REPAIR AND RECONSTRUCT THE ANTERIOR CRUCIATE LIGAMENT, SUTURE OF THE LATERAL MENISCUS AND SUTURE OF THE MEDIAL COLLATERAL LIGAMENT OF THE LEFT KNEE WERE PERFORMED. AFTER SURGERY, X-RAY EXAMINATION WAS PERFORMED AND SHOWED THERE WERE SUSPICIOUS METAL FOREIGN BODIES IN THE LEFT KNEE AND PATIENT SWELLING ABOUT LEFT KNEE. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED ACCORDINGLY.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 9545991 ↗MalfunctionJul 28, 2019No Display/Image; No Display/Image
IT WAS REPORTED THAT THERE WAS DARK SCREEN INDICATING A CRACK IN THE SCOPE. NO PATIENT INJURIES OR DELAY REPORTED. BACK-UP DEVICE WAS AVAILABLE TO COMPLETE THE SURGERY.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8834818 ↗MalfunctionJan 16, 2019Break
IT WAS REPORTED THAT DURING THE PROCEDURE, THE MECHANICAL BURR NICKED THE TIP OF THE SCOPE. THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITHOUT DELAY AND THERE WAS NO BACK-UP AVAILABLE. NO PATIENT INJURY OR OTHER COMPLICATIONS WERE REPORTED.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8253515 ↗MalfunctionApr 2, 2018Device Inoperable
IT WAS REPORTED THAT THE SCOPE STOPPED WORKING DURING A PROCEDURE AND IT IS UNKNOWN IF THERE WAS A LOSS OF IMAGE OR VISUALIZATION DURING THE SURGICAL PROCEDURE. PROCEDURE WAS COMPLETED WITH A BACK-UP DEVICE. NO PATIENT INJURY REPORTED.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7390790 ↗
Other ARTHROSCOPE devices
- FMS 283512Medos International Sàrl · ARTHROSCOPE1,800 reports
- Arthrex® AR-6480Arthrex, Inc. · ARTHROSCOPE949 reports
- FMS VUE 284002Depuy Mitek, LLC · ARTHROSCOPE829 reports
- MITEK 242018Medos International Sàrl · Arthroscope731 reports
- METRx® System 9560100Medtronic Sofamor Danek, Inc. · Arthroscope667 reports
- FMS VUE II 284004Medos International Sàrl · Arthroscope562 reports
- VAPR COOLPULSE 228146Depuy Mitek, LLC · ARTHROSCOPE474 reports
- FMS DUO 284580Depuy Mitek, LLC · ARTHROSCOPE458 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page