Skip to content

NA 72202089Recalls & reported problems

Smith & Nephew, Inc. · DIRECT-VIEW ARTHROSCOPE AUTOCLAVABLE, 4.0 MM OUTER DIAMETER X 0 DEGREE DIRECTION OF VIEW, 115 DEGREE FIELD OF VIEW, 157 MM WORKING LENGTH, J-LOCK CANNULA CONNECTION · Class 2

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 03596010616883), received between Mar 30, 2021 and Oct 24, 2021 — classified as 2 injuries.

The most frequently reported specific device problem is “Break” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Most reported problems

Device problems

  • Break1
  • Detachment of Device or Device Component1

Patient problems

  • Foreign Body In Patient1
  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • InjuryOct 24, 2021Detachment of Device or Device Component; Detachment of Device or Device Component

    IT WAS REPORTED THAT DURING AN ARTHROSCOPY, PART OF THE TIP CAME OFF FROM THE SCOPE AND FALL OFF INSIDE OF THE PATIENT. THE PIECE WAS REMOVED FROM THE PATIENT. THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITH NINETY MINUTES OF DELAY USING A BACK-UP DEVICE. NO PATIENT COMPLICATIONS WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12686879 ↗
  • InjuryMar 30, 2021Break; Break

    IT WAS REPORTED THAT DURING AN ACL RECONSTRUCTION PROCEDURE, THE TIP OF THE SCOPE WAS DEFORMED AND IT HAD PARTS MISSING. AFTER FIXING THE FEMUR SIDE WITH CLBTB, AN IMAGE WAS TAKEN AND IT WAS FOUND THAT A METAL PIECE WAS LEFT INTO THE PATIENT. THAT MEANS, THE BROKEN PIECE COULD NOT BE REMOVED. THEY DECIDED TO FINISH THE OPERATION WITHOUT ANY TREATMENT. A DELAY GREATER THAN 30 MINUTES WAS REPORTED. IT IS UNKNOWN IF THERE WAS AN AVAILABLE BACK UP DEVICE. NO OTHER COMPLICATIONS WERE REPORTED. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED ACCORDINGLY.

    Patient problems: Foreign Body In Patient

    MAUDE report 11590302 ↗

Other ARTHROSCOPE devices

All ARTHROSCOPE devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page