SMALL CANNULATED SCREWS 71103240SRecalls & reported problems
Smith & Nephew, Inc. · 3.0MM CANNULATED SCREW 40MM STERILE · Class 2 · Implant
All models of SMALL CANNULATED SCREWS (18)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 03596010623997), received between May 15, 2024 and Dec 2, 2025 — classified as 1 injury, 3 malfunctions.
The most frequently reported specific device problem is “Fracture” (3 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Fracture3
- Insufficient Device Problem Information1
Patient problems
- No Clinical Signs, Symptoms or Conditions3
- Device Embedded In Tissue or Plaque1
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
InjuryDec 2, 2025Fracture; Fracture
IT WAS REPORTED THAT DURING AN OSTEOSYNTHESIS SURGERY, AFTER IMPLANTING ONE (1) 3.0MM CANNULATED SCREW 40MM IN THE MEDIAL MALLEOLUS, THE DOCTOR FINISHES SCREWING IT IN AND WHEN IT REACHES THE STOP (THE SCREW HEAD STOPS AGAINST THE CORTEX) THE HEAD COMES LOOSE AND REMAINS IN THE SCREWDRIVER. THE PROCEDURE WAS COMPLETED, AFTER NO DELAY, USING THE SAME S+N PRODUCT. NO INJURY WAS REPORTED AS A CONSEQUENCE OF THIS ISSUE.
Patient problems: Device Embedded In Tissue or Plaque
MAUDE report 23699870 ↗MalfunctionDec 5, 2024Insufficient Device Problem Information; Insufficient Device Problem Information
IT WAS REPORTED THAT DURING PATELLA FIXATION SURGERY, ONE (1) 3.0MM CANNULATED SCREW 40MM FAILED. THE PROCEDURE WAS COMPLETED AFTER A SIGNIFICANT DELAY USING A S+N BACK-UP DEVICE. NO INJURY WAS REPORTED AS A CONSEQUENCE OF THIS ISSUE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20858568 ↗MalfunctionOct 14, 2024Fracture; Fracture
IT WAS REPORTED THAT DURING THE AN INTERNAL FIXATION, THE 3.0MM CANNULATED SCREW 40MM WAS STARTED AND THE HEAD BECAME LOOSE. THE PROCEDURE WAS RESUMED, AFTER A SIGNIFICANT DELAY, WITH A S+N BACK-UP DEVICE. PATIENT WAS NOT INJURED AS CONSEQUENCE OF THIS PROBLEM. THE PATIENT'S HEALTH CONDITION IS STABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20440469 ↗MalfunctionMay 15, 2024Fracture; Fracture
IT WAS REPORTED THAT, DURING THE OSTEOSYNTHESIS SURGERY, THE HEAD OF ONE (1) 3.0MM CANNULATED SCREW 36MM AND ONE (1) 3.0MM CANNULATED SCREW 40MM WERE BROKEN. THE PROCEDURE WAS RESUMED, AFTER A NON-SIGNIFICANT DELAY, WITH A S+N BACK-UP DEVICE. NO INJURY WAS REPORTED AS A CONSEQUENCE OF THIS ISSUE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19321363 ↗
Other PLATE, FIXATION, BONE devices
- Lapiplasty System 2 SK14Treace Medical Concepts, Inc. · Plate, Fixation, Bone260 reports
- PRECICE® Plating System LP4.5-PRF4BNuvasive Specialized Orthopedics, Inc. · Plate, Fixation, Bone3 recalls6 reports
- PRECICE® Plating System LP4.5-MLT4BNuvasive Specialized Orthopedics, Inc. · Plate, Fixation, Bone3 recalls5 reports
- PRECICE® Plating System LP4.5-MRT4BNuvasive Specialized Orthopedics, Inc. · Plate, Fixation, Bone3 recalls3 reports
- PRECICE® Plating System LP4.5-PLF4BNuvasive Specialized Orthopedics, Inc. · Plate, Fixation, Bone3 recalls3 reports
- PRECICE® Plating System LP4.5-DRF4BNuvasive Specialized Orthopedics, Inc. · Plate, Fixation, Bone3 recalls1 report
- PRECICE® Plating System LP4.5-MLT42Nuvasive Specialized Orthopedics, Inc. · Plate, Fixation, Bone3 recalls0 reports
- PRECICE® Plating System LP4.5-MRT42Nuvasive Specialized Orthopedics, Inc. · Plate, Fixation, Bone3 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page