TWINFIX 72202624Recalls & reported problems
Smith & Nephew, Inc. · TWINFIX ULTRA 4.5 MM HIDROXYAPATITE ANCHOR WITH TWO No.2 SUTURES COBRAID-BLUE, COBRAID-BLACK STERILE · Class 2 · Implant
All models of TWINFIX (48)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 14
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 14 adverse event reports that list this exact device (DI 03596010647566), received between Nov 1, 2021 and Jul 10, 2025 — classified as 2 injuries, 12 malfunctions.
The most frequently reported specific device problem is “Break” (14 reports).
40 other recalls cover devices cleared under the same FDA submission (K151105); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Break14
Patient problems
- No Clinical Signs, Symptoms or Conditions14
Recent adverse event reports
The 14 most recent of 14 reports, in the reporter's own words as published by the FDA.
MalfunctionJul 10, 2025Break; Break
IT WAS REPORTED THAT DURING AN ARTHROSCOPY, THE TWINFIX ULTRA ANCHOR BROKE WHEN PUTTING INTO THE BONE. THE PROCEDURE WAS COMPLETED WITH A S+N BACK UP DEVICE. THERE WAS A SURGICAL DELAY OF LESS THAN 30 MINUTES AND NO FURTHER COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22468378 ↗MalfunctionJul 8, 2025Break; Break
IT WAS REPORTED THAT, DURING AN ARTHROSCOPY, THE TWINFIX ULTRA ANCHOR BROKE WHILE INSERTING IT INTO THE BONE. SURGERY WAS COMPLETED WITH A SMITH AND NEPHEW BACK-UP DEVICE INSTEAD. THERE WAS A SURGICAL DELAY OF LESS THAN 30 MINUTES, AND NO FURTHER COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22438236 ↗MalfunctionJul 8, 2025Break; Break
IT WAS REPORTED THAT, DURING AN ARTHROSCOPY, THE TWINFIX ULTRA ANCHOR BROKE WHILE INSERTING IT INTO THE BONE. SURGERY WAS COMPLETED WITH A SMITH AND NEPHEW BACK-UP DEVICE INSTEAD. THERE WAS A SURGICAL DELAY OF LESS THAN 30 MINUTES, AND NO FURTHER COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22437972 ↗MalfunctionJun 11, 2025Break; Break
IT WAS REPORTED THAT DURING AN ARTHROSCOPY, THE TWINFIX ANCHOR GOT BROKEN WHILE PUTTING IT INTO THE BONE. THE PROCEDURE WAS COMPLETED USING A BACK-UP DEVICE. THERE WAS A DELAY LESS THAN 30 MINUTES AND NO FURTHER COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22181722 ↗MalfunctionApr 2, 2025Break; Break
IT WAS REPORTED THAT DURING AN ARTHROSCOPY, THE TWINFIX ULTRA HA ANCHOR GOT BROKEN WHILE PUTTING INTO THE BONE. THE PROCEDURE WAS COMPLETED WITH A SMITH AND NEPHEW BACK UP DEVICE. THERE WAS A DELAY LESS THAN 30 MINUTES AND NO FURTHER COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21758984 ↗MalfunctionMar 10, 2025Break; Break
IT WAS REPORTED THAT DURING AN ARTHROSCOPY, THE TWINFIX ULTRA ANCHOR BROKE WHILE BEING PLACED INTO THE BONE. THE PROCEDURE WAS COMPLETED WITH A SURGICAL DELAY OF LESS THAN 30 MINUTES USING A SMITH AND NEPHEW BACKUP DEVICE, AND NO FURTHER COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21564396 ↗MalfunctionFeb 21, 2025Break; Break
IT WAS REPORTED THAT DURING AN ARTHROSCOPY, THE TWINFIX ULTRA ANCHOR BROKE WHILE IT WAS INSERTED INRO THE BONE. THE PROCEDURE WAS COMPLETED WITH A SMITH AND NEPHEW BACK UP DEVICE. THERE WAS A DELAY LESS THAN 30 MINUTES AND NO FURTHER COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21438698 ↗MalfunctionFeb 19, 2025Break; Break
IT WAS REPORTED THAT DURING AN ARTHROSCOPY, THE TWINFIX ULTRA ANCHOR GOT BROKEN WHILE TIGHTENING. THE PROCEDURE WAS COMPLETED WITH NON-SIGNIFICANT SURGICAL DELAY USING A BACK-UP DEVICE. NO FURTHER COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21413278 ↗MalfunctionDec 12, 2024Break; Break
IT WAS REPORTED THAT DURING AN ARTHROSCOPY, THE TWINFIX ULTRA ANCHOR BROKE WHILE BEING PLACED ONTO THE BONE. THE PROCEDURE WAS COMPLETED WITH A S+N BACK UP DEVICE. THERE WAS A SURGICAL DELAY LESS THAN 30 MINUTES AND NO FURTHER COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20912934 ↗MalfunctionJun 7, 2024Break; Break
IT WAS REPORTED THAT DURING A SHOULDER ARTHROSCOPY, THE TIP OF THE TWINFIX ULTRA ANCHOR WAS FOUND TO BE BROKEN WHEN IT WAS IN THE PATIENT BONE, THE RESIDUE WAS REMOVED FROM THE PATIENT. THE PROCEDURE WAS COMPLETED WITH A SMITH AND NEPHEW BACK UP DEVICE WITH NO SURGICAL DELAY. NO FURTHER COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19485111 ↗MalfunctionApr 1, 2024Break; Break
IT WAS REPORTED THAT DURING AN ARTHROSCOPY, THE TWINFIX ULTRA ANCHOR BROKE WHILE INSERTING IN THE BONE AFTER DRILLING. THE PROCEDURE WAS FINISHED WITH A SMITH AND NEPHEW BACK UP DEVICE. THERE WAS A DELAY LESS THAN 30 MINUTES AND NO FURTHER COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19011448 ↗MalfunctionOct 30, 2023Break; Break
IT WAS REPORTED THAT DURING AN ARTHROSCOPY, WHILE THE SURGEON WAS PLACING THE TWINFIX ULTRA ANCHOR IN THE BONE, IT BROKE; THE ANCHOR CAME OUT OF THE PATIENT WITH THE INSERTER ITSELF. THE PROCEDURE WAS COMPLETED WITHOUT SURGICAL DELAY USING A S+N BACK UP DEVICE IN THE ORIGINAL BONE HOLE. NO FURTHER COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18031017 ↗InjuryJul 29, 2022Break; Break
IT WAS REPORTED THAT DURING AN ARTHROSCOPY PROCEDURE, THE THREADS OF TWO (2) 4.5MM TWINFIX ULTRA HA ANCHORS SHEERED FROM THE TIP END WHILE SCREWING AND THE SURGEON WAS UNABLE TO TIGHTEN THE ANCHOR. THE DEVICES WERE REMOVED FORM THE PATIENT WITH TWEEZERS AND GRASPERS. THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITH NO SURGICAL DELAY USING A BACK-UP DEVICES IN THE ORIGINALLY DRILLED BONE HOLES. NO FURTHER COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15135864 ↗InjuryNov 1, 2021Break; Break
IT WAS REPORTED THAT, DURING A SHOULDER ARTHROSCOPY, INSTRUMENTS INSIDE PATIENT, THE TWINFIX BROKE DURING INSERTION, AFTER TAPPING. THE PROCEDURE WAS COMPLETED WITH THE SAME DEVICE. NON-SIGNIFICANT DELAY AND NO PATIENT COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12729447 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K151105. One submission often covers many models, so check the product and lot details in each record.
- Z-0175-2021 · Sep 9, 2020 · Open, ClassifiedMICRORAPTOR REGENESORB SUTURE ANCHOR- Absorbable Suture anchor Product Number: 72204984
- Z-0148-2021 · Sep 9, 2020 · Open, ClassifiedOSTEORAPTOR 2.3 W 1UB II BLUE-Absorbable Suture anchor Product Number: 72201991
- Z-0136-2021 · Sep 9, 2020 · Open, ClassifiedTWINFIX 5.0MM AB SUTURE ANCHOR WITH TWO Absorbable Suture anchors Product Number: 7210309
- Z-0145-2021 · Sep 9, 2020 · Open, ClassifiedULTRA FAST-FIX AB ASSEMBLY-REVERSE CURVE-Absorbable Suture anchor Product Number: 72201495
- Z-0154-2021 · Sep 9, 2020 · Open, ClassifiedOSTEORAPTOR 2.9 W 1UB II BLUE--Absorbable Suture anchor Product Number: 72202165
- Z-0156-2021 · Sep 9, 2020 · Open, ClassifiedTWINFIX ULTRA PLLA/HA 5.5 W/3 UB-Absorbable Suture anchor Product Number: 72202603
- Z-0166-2021 · Sep 9, 2020 · Open, ClassifiedOSTEORAPTOR CRVD 2.3 SA UB COBRD BLACK- Absorbable Suture anchor Product Number: 72203291
- Z-0168-2021 · Sep 9, 2020 · Open, ClassifiedHEALICOIL RSB SA 4.75MM W/1 UT & 1 UB BL-- Absorbable Suture anchor Product Number: 72203704
- Z-0142-2021 · Sep 9, 2020 · Open, ClassifiedTWINFIX AB 5.0 SUTR ANCHR W/2 38 DURAB-Absorbable Suture anchors Product Number: 72200785
- Z-0160-2021 · Sep 9, 2020 · Open, ClassifiedTWINFIX ULTRA PLLA HA 5.5MM W/ NEEDLES- Absorbable Suture anchor Product Number: 72202616
- Z-0155-2021 · Sep 9, 2020 · Open, ClassifiedTWINFIX ULTRA 5.5MM PLLA/HA 2 UBWHT/BL--Absorbable Suture anchor Product Number: 72202602
- Z-0152-2021 · Sep 9, 2020 · Open, ClassifiedOSTEORAPTOR 2.9 W/ 2 UB WHITE / BLUE -Absorbable Suture anchor Product Number: 72201995
- Z-0159-2021 · Sep 9, 2020 · Open, ClassifiedTWINFIX ULTRA PLLA HA 4.5MM W/ NEEDLES- Absorbable Suture anchor Product Number: 72202612
- Z-0171-2021 · Sep 9, 2020 · Open, ClassifiedHEALICOIL RG SA 5.5MM W/3 UB-BL CB BL BK- Absorbable Suture anchor Product Number: 72203707
- Z-0139-2021 · Sep 9, 2020 · Open, ClassifiedTWINFIX 6.5MM AB PRELOADED ULTRABRAID- Absorbable Suture anchors Product Number: 7210712
…and 25 more.
Other FASTENER, FIXATION, BIODEGRADABLE, SOFT TISSUE devices
- GRYPHON ORTHOCORD 210813Medos International Sàrl · FASTENER, FIXATION, BIODEGRADABLE, SOFT TISSUE712 reports
- HEALIX ADVANCE 222330Depuy Mitek, LLC · FASTENER, FIXATION, BIODEGRADABLE, SOFT TISSUE1 recall607 reports
- HEALIX ADVANCE ORTHOCORD 222295Medos International Sàrl · FASTENER, FIXATION, BIODEGRADABLE, SOFT TISSUE640 reports
- HEALICOIL 72205137Smith & Nephew, Inc. · Fastener, fixation, biodegradable, soft tissue427 reports
- HEALIX ADVANCE ORTHOCORD 222296Medos International Sàrl · FASTENER, FIXATION, BIODEGRADABLE, SOFT TISSUE320 reports
- TWINFIX 72202597Smith & Nephew, Inc. · FASTENER, FIXATION, BIODEGRADABLE, SOFT TISSUE288 reports
- Arthrex® AR-2324BCCArthrex, Inc. · FASTENER, FIXATION, BIODEGRADABLE, SOFT TISSUE254 reports
- HEALIX ADVANCE 222331Depuy Mitek, LLC · FASTENER, FIXATION, BIODEGRADABLE, SOFT TISSUE1 recall196 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page