Porges Coloplast AA6314Recalls & reported problems
Coloplast A/S · FOLYSIL® catheter Tiemann CH FR 14 · Class 2
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 13
- Deaths reported
- 0
- Injuries reported
- 3
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 13 adverse event reports that list this exact device (DI 03600040128640), received between Jan 15, 2019 and Oct 30, 2024 — classified as 3 injuries, 10 malfunctions.
The most frequently reported specific device problem is “Burst Container or Vessel” (8 reports).
1 FDA recall record name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Recalls naming this device
- Dec 2, 2024Open, Classified
A possible sterility issue was detected in Coloplast's facility on some Coloplast products.
Z-0844-2025 · Root cause: Packaging · FDA record ↗
Most reported problems
Device problems
- Burst Container or Vessel8
- Separation Problem2
- Contamination1
- Deflation Problem1
- Material Rupture1
- Obstruction of Flow1
Patient problems
- No Known Impact Or Consequence To Patient5
- No Clinical Signs, Symptoms or Conditions4
- Urinary Retention2
- No Information1
- Unspecified Infection1
Recent adverse event reports
The 13 most recent of 13 reports, in the reporter's own words as published by the FDA.
InjuryOct 30, 2024Burst Container or Vessel; Burst Container or Vessel
ACCORDING TO THE AVAILABLE INFORMATION, THE PATIENT GETS A NEW CATHETER IN AND TWO DAYS LATER THE BALLOON BREAKS. THE PATIENT ENDS UP IN THE EMERGENCY TO GET A NEW CATHETER BECAUSE HE CANNOT VOID HIS URINE.
Patient problems: Urinary Retention
MAUDE report 20569179 ↗MalfunctionOct 15, 2023Burst Container or Vessel; Material Rupture; Burst Container or Vessel; Material Rupture
ACCORDING TO AVAILABLE INFORMATION, THIS DEVICE HAD A BALLOON BURST. THE NURSE INFLATED 10CC OF STERILE WATER FOR TESTING. AFTER THE TESTING, THE NURSE PLACED THE DEVICE AND INFLATED 10CC OF STERILE WATER AGAIN. THE DEVICE SLIPPED OUT FROM THE PATIENT. THE NURSE CHECKED THE DEVICE AND FOUND THAT THE BALLOON BROKE. NO OTHER ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17936842 ↗MalfunctionJul 24, 2023Contamination; Contamination
ACCORDING TO AVAILABLE INFORMATION, THIS DEVICE WAS NOT ABLE TO BE USED DUE TO A HAIR. THERE WAS A HAIR IN THE PACKAGING BEFORE IT WAS OPENED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17375851 ↗InjuryMar 28, 2023Obstruction of Flow; Obstruction of Flow
ACCORDING TO AVAILABLE INFORMATION, THIS DEVICE REQUIRED REPLACEMENT DUE TO NOT FILLING. THE DEVICE WAS CHECKED AFTER 2 HOURS AND NO URINE HAD COLLECTED IN THE BAG. THE DEVICE WAS NOT ABLE TO BE UNBLOCKED. THE PATIENT HAD A PELVIC ULTRASOUND AND A BLADDER GLOBE WAS FOUND. DURING REPLACEMENT IT WAS FOUND THAT THE DISTAL LUMEN HAD NOT COMPLETELY PUNCTURED. NO OTHER ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: Urinary Retention
MAUDE report 16627259 ↗MalfunctionFeb 16, 2023Separation Problem; Separation Problem
ACCORDING TO AVAILABLE INFORMATION, THIS DEVICE SEPARATED DURING PLACEMENT. DURING PLACEMENT OF THE DEVICE THE CRUTCH END OF THE DEVICE CAME OFF. NO OTHER ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16384052 ↗MalfunctionFeb 14, 2023Separation Problem; Separation Problem
ACCORDING TO AVAILABLE INFORMATION, THIS DEVICE SEPARATED DURING PLACEMENT. DURING PLACEMENT OF THE DEVICE THE CRUTCH END OF THE DEVICE CAME OFF. NO OTHER ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16372048 ↗MalfunctionNov 19, 2022Deflation Problem; Deflation Problem
ACCORDING TO AVAILABLE INFORMATION, THIS DEVICE REQUIRED REPLACEMENT DUE TO DEFLATION AND INFECTION. THERE WERE NO ISSUES WITH THE INFLATION TEST BEFORE PLACEMENT OF THE DEVICE. AFTER THREE DAYS INDWELLING, THE DEVICE FELL OFF AND THE BALLOON DEFLATED. A LEAK WAS FOUND AT THE TIP SIDE OF THE BALLOON. IT WAS NOTED THE PATIENT HAD A POSSIBLE INFECTION. NO OTHER ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: Unspecified Infection
MAUDE report 15829810 ↗MalfunctionSep 6, 2019Burst Container or Vessel; Burst Container or Vessel
ACCORDING TO THE AVAILABLE INFORMATION, THE INDWELLING CATHETER BALLOON BURST AFTER 2-3 DAYS WHILE IT WAS INSERTED. THE BALLOON WAS PUMPED 8-10 TIMES. THERE HAS BEEN NO HARM TO THE END USER.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8972686 ↗MalfunctionSep 6, 2019Burst Container or Vessel; Burst Container or Vessel
ACCORDING TO THE AVAILABLE INFORMATION, THE INDWELLING CATHETER BALLOON BURST AFTER 2-3 DAYS WHILE IT WAS INSERTED. THE BALLOON WAS PUMPED 8-10 TIMES. THERE HAS BEEN NO HARM TO THE END USER.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8972684 ↗MalfunctionSep 6, 2019Burst Container or Vessel; Burst Container or Vessel
ACCORDING TO THE AVAILABLE INFORMATION, THE INDWELLING CATHETER BALLOON BURST AFTER 2-3 DAYS WHILE IT WAS INSERTED. THE BALLOON WAS PUMPED 8-10 TIMES. THERE HAS BEEN NO HARM TO THE END USER.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8972683 ↗MalfunctionSep 6, 2019Burst Container or Vessel; Burst Container or Vessel
ACCORDING TO THE AVAILABLE INFORMATION, THE INDWELLING CATHETER BALLOON BURST AFTER 2-3 DAYS WHILE IT WAS INSERTED. THE BALLOON WAS PUMPED 8-10 TIMES. THERE HAS BEEN NO HARM TO THE END USER.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8972679 ↗MalfunctionSep 6, 2019Burst Container or Vessel; Burst Container or Vessel
ACCORDING TO THE AVAILABLE INFORMATION, THE INDWELLING CATHETER BALLOON BURST AFTER 2-3 DAYS WHILE IT WAS INSERTED. THE BALLOON WAS PUMPED 8-10 TIMES. THERE HAS BEEN NO HARM TO THE END USER.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8972677 ↗InjuryJan 15, 2019Burst Container or Vessel; Burst Container or Vessel
ACCORDING TO THE AVAILABLE INFORMATION, THE BALLOON FELL OFF DURING INDWELLING DUE TO BALLOON BURST, THE PHYSICIAN CONFIRMED NO RESIDUE OR PARTICLE REMAINS IN PATIENT'S BLADDER, REPLACED WITH NEW CATHETER IMMEDIATELY. NO CLINICAL CONSEQUENCES REPORTED.
Patient problems: No Information
MAUDE report 8248323 ↗
Other CATHETER, RETENTION TYPE, BALLOON devices
- Bardex® All-Silicone Foley Catheter 165816C. R. Bard, Inc. · CATHETER, RETENTION TYPE, BALLOON512 reports
- Bardex® Lubri-Sil® All-Silicone Foley Catheter 175816C. R. Bard, Inc. · CATHETER, RETENTION TYPE, BALLOON476 reports
- Bardex® Lubri-Sil® I.C. All-Silicone Foley Catheter 1758SI16C. R. Bard, Inc. · CATHETER, RETENTION TYPE, BALLOON1 death383 reports
- Bardex® Lubri-Sil® All-Silicone Foley Catheter 175812C. R. Bard, Inc. · CATHETER, RETENTION TYPE, BALLOON234 reports
- Bardex® Lubri-Sil® All-Silicone Temp Sensing Foley Catheter 119216MC. R. Bard, Inc. · CATHETER, RETENTION TYPE, BALLOON167 reports
- Bardex® Lubri-Sil® All-Silicone Foley Catheter 175814C. R. Bard, Inc. · CATHETER, RETENTION TYPE, BALLOON153 reports
- Bardex® All-Silicone Foley Catheter 165812C. R. Bard, Inc. · CATHETER, RETENTION TYPE, BALLOON151 reports
- SureStep™ Foley Tray System Bard® Lubri-Sil® Foley Catheter Tray A942216C. R. Bard, Inc. · CATHETER, RETENTION TYPE, BALLOON2 recalls46 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.