Porges Coloplast recallZ-0844-2025 · Coloplast Manufacturing US, LLC · initiated Dec 2, 2024
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Coloplast Manufacturing US, LLC or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
A possible sterility issue was detected in Coloplast's facility on some Coloplast products.
FDA root cause category: Packaging
What the company is doing
An URGENT MEDICAL DEVICE VOLUNTARY RECALL notification letter dated 12/2/24 was sent to customers. Distributors, it is imperative that you take immediate action to cascade this recall notice to all relevant subsidiaries and customers who may have received the affected product within your distribution network. Customers affected by this recall are kindly advised to immediately inspect their internal inventory and quarantine all products and samples of the items with the lot numbers mentioned in Appendix 1 and then proceed to safe destruction. Please contact Customer Service for any assistance: Email: [email protected] Phone: +1 855 745 8549 Expenses will be refunded by Coloplast A/S upon receipt of the completed Certificate of Destruction provided in Appendix 2.
Recall details
- Recall number
- Z-0844-2025
- Status
- Open, Classified
- Initiated by the company
- Dec 2, 2024
- Posted by the FDA
- Jan 10, 2025
- Recalling firm
- Coloplast Manufacturing US, LLC
- Quantity
- 226,452 units
- Product code
- EZL
- FDA submissions
- K233411
- Distribution
- US Nationwide. Canada.
Product description
Folysil Silicone Foley Catheters, Device Reference Numbers AA6106, AA6108, AA6110, AA6112, AA6114, AA6116, AA6118, AA6120, AA6122, AA6124, AA6308, AA6310, AA6312, AA6314, AA6316, AA6318, AA6320, AA6322, AA6324, AA6408, AA6410, AA6412, AA6414, AA6416, AA6418, AA6420, AA6422, AA6424, AA6C18, AA6C20, AA6C22, and AA6C24.
Code information (models, lots, serial numbers)
Show the full code information (13k characters)
Device Reference Number AA6106, GTIN 3600040126851, Lot Numbers 7278611, 7326825, 7469914, 7528770, 7540052, 7616917, 7680676, 7726724, 7748128, 7794789, 7874610, 7883406, 7948568, 7987184, 7987186, 7998885, 8097945, 8097947, 8097949, 8097950, 8149951, 8149954, 8156925, 8326511, 8326512, 8392428, 8452914, 8452915, 8519853, 8559628, 8572973, 8572977, 8728905, 8742746, 8796890, 8829455, 8899165, 8911320, 8996565, 9139298, 9177631, 9177632, 9192348, 9294521, 9294523, 9294524, 9315917, 9374195, 9404781, 9404782, 9419436, 9429227, 9429228, 9445801, 9453606, 9700652, 9711505, 9778601, 9778603, 9778606, 9792251, 9792257, 9817387, 9829471, 9843572, 9843573, 9940969 ; Device Reference Number AA6108, GTIN 3600040126981, Lot Numbers 7278611, 7326825, 7469914, 7528770, 7540052, 7616917, 7680676, 7726724, 7748128, 7794789, 7874610, 7883406, 7948568, 7987184, 7987186, 7998885, 8097945, 8097947, 8097949, 8097950, 8149951, 8149954, 8156925, 8326511, 8326512, 8392428, 8452914, 8452915, 8519853, 8559628, 8572973, 8572977, 8728905, 8742746, 8796890, 8829455, 8899165, 8911320, 8996565, 9139298, 9177631, 9177632, 9192348, 9294521, 9294523, 9294524, 9315917, 9374195, 9404781, 9404782, 9419436, 9429227, 9429228, 9445801, 9453606, 9700652, 9711505, 9778601, 9778603, 9778606, 9792251, 9792257, 9817387, 9829471, 9843572, 9843573, 9940969; Device Reference Number AA6110, GTIN 3600040127117, Lot Numbers 7255806, 7270683, 7270685, 7278618, 7303190, 7427087, 7440872, 7469982, 7487187, 7503933, 7564542, 7596635, 7629271, 7703729, 7713958, 7774417, 7774419, 7794951, 8133792, 8156966, 8156972, 8297224, 8313405, 8350384, 8519861, 8539757, 8679575, 8781871, 8796821, 8839854, 8839855, 8985535, 8996599, 9008443, 9031094, 9067753, 9077697, 9090464, 9102793, 9213533, 9213536, 9240194, 9240195, 9240196, 9240197, 9248068, 9261075, 9282360, 9282362, 9294615, 9294624, 9304466, 9315932, 9315933, 9341162, 9351850, 9351851, 9374230, 9385336, 9396721, 9405092, 9420752, 9420753, 9429249, 9445822, 9445823, 9472404, 9472405, 9553577, 9553578, 9610617, 9634462, 9634464, 9634465, 9656853, 9656854, 9656855, 9656856, 9656857, 9666697, 9666698, 9683046, 9683047, 9690316, 9700707, 9711520, 9792840, 9792841, 9843580, 9843581 ; Device Reference Number AA6112, GTIN 3600040127247, Lot Numbers 7174166, 7174170, 7174172, 7338746, 7338747, 7338751, 7470039, 7576132, 7576144, 7585050, 7585053, 7607531, 7670529, 7680781, 7894977, 7929763, 7929766, 7929768, 7929772, 8030455, 8030458, 8122983, 8122990, 8289326, 8297221, 8313525, 8326570, 8326572, 8361200, 8472134, 8506837, 8559544, 8592891, 8679869, 8728871, 8796859, 8948101, 8985560, 9041086, 9102788, 9141405, 9177670, 9192341, 9240188, 9282330, 9282342, 9294599, 9341164, 9351839, 9351840, 9495897, 9529685, 9622813, 9634315, 9682510, 9682512, 9682513, 9690266, 9792211, 9792214, 9792217, 9792219, 9792220, 9792222, 9792236 ; Device Reference Number AA6114, GTIN 3600040127377, Lot Numbers 7174266, 7238934, 7238937, 7290670, 7303277, 7328602, 7328604, 7362452, 7387284, 7396140, 7396147, 7441024, 7680769, 7735761, 7782360, 7782371, 7794999, 7795001, 7942148, 7942154, 7969037, 7998909, 8040737, 8040751, 8098020, 8289321, 8361123, 8361133, 8408149, 8418184, 8442278, 8592968, 8620070, 8679825, 8679833, 8679834, 8728885, 8853718, 8884537, 8911244, 8923138, 8948148, 8975016, 8985570, 8985571, 8985589, 9031124, 9031133, 9102804, 9122861, 9226662, 9226666, 9248080, 9351817, 9385277, 9419405, 9419408, 9453470, 9460169, 9485166, 9529697, 9558587, 9558600, 9585508, 9585518, 9597173, 9597181, 9610603, 9610606, 9700656, 9700662, 9737166, 9748487, 9748831, 9748832, 9755424, 9755427, 9755429, 9755439, 9780645, 9780646, 9780648, 9780661, 9780679, 9780683, 9792890, 9792891, 9792893 ; Device Reference Number AA6116, GTIN 3600040127506, Lot Numbers 7193434, 7255904, 7314786, 7328608, 7328617, 7328618, 7406746, 7416574, 7447516, 7447523, 7487323, 7487328, 7494334, 7607523, 7680801, 7714102, 7794958, 7810306, 7883541, 7883545, 7942168, 8030469, 8087140, 8087162, 8111962, 8200198, 8268459, 8350405, 8361178, 8361186, 8418190, 8431498, 8442273, 8470200, 8472145, 8519842, 8529333, 8559608, 8572962, 8679849, 8679855, 8691267, 8728891, 8829399, 8829402, 8865298, 8884578, 8923523, 8948167, 8997529, 9008469, 9055014, 9077729, 9112123, 9141408, 9201829, 9240174, 9261058, 9261059, 9282296, 9294589, 9316028, 9341175, 9404784, 9429218, 9445790, 9453593, 9478195, 9495829, 9529688, 9540968, 9553558, 9558616, 9558618, 9585503, 9585504, 9610582, 9610598, 9622842, 9634401, 9634405, 9634423, 9634433, 9656837, 9666685, 9700700, 9700701, 9755483, 9755497 ; Device Reference Number AA6118, GTIN 3600040127612, Lot Numbers 7161607, 7229862, 7278686, 7290713, 7350612, 7607542, 7657482, 7657485, 7735781, 7735785, 7794982, 7794987, 7836019, 7919589, 7919593, 7919604, 7948649, 7998889, 7998893, 8244807, 8244816, 8255400, 8289336, 8341241, 8460199, 8559580, 8572905, 8700341, 8742738, 8742739, 8749952, 8829462, 8948269, 9008685, 9008690, 9041121, 9067845, 9090781, 9102910, 9122821, 9143141, 9261278, 9261283, 9261293, 9304728, 9304733, 9362306, 9362321, 9472350, 9472368, 9559982, 9585591, 9585598 ; Device Reference Number AA6120, GTIN 3600040127742, Lot Numbers 7174152, 7314709, 7314711, 7350584, 7362408, 7597839, 7607515, 7774397, 7883536, 7883538, 7942098, 7948628, 8350320, 8350322, 8442250, 8884440, 8985523, 9041132, 9201860, 9429193, 9429194, 9460760, 9460768, 9597150, 9711497, 9792240, 9792243 ; Device Reference Number AA6122, GTIN 3600040127872, Lot Numbers 7405871, 7416497, 7748381, 7942074, 8350342, 8493250, 8519854, 8592989, 8728906, 8985443, 9159829, 9404773, 9445789, 9453600, 9453601, 9485195, 9829431 ; Device Reference Number AA6124, GTIN 3600040128008, Lot Numbers 7416543, 7516929, 7773792, 7850835, 8442227, 8442228, 8442229, 9008561, 9067848, 9656792 ; Device Reference Number AA6308, GTIN 3600040128268, Lot Numbers 7210663, 7440856, 7447457, 7470016, 7516834, 7528782, 7703710, 7761577, 7782071, 7810060, 7883513, 7919523, 7919524, 7977718, 7977720, 8111946, 8111947, 8268401, 8268407, 8326502, 8341188, 8452349, 8559594, 8572917, 8679560, 8817139, 8829340, 8839643, 9090446, 9122812, 9226586, 9240167, 9351797, 9351798, 9429188, 9485153, 9495821, 9634306, 9656800, 9682505, 9690261, 9940979, 9940980, 9940981, 9940982 ; Device Reference Number AA6310, GTIN 3600040128398, Lot Numbers 7224527, 7270676, 7326836, 7503928, 7564546, 7575984, 7629249, 7669034, 7694121, 7782053, 7825239, 7850714, 7850715, 7874609, 7894935, 7942054, 7977212, 8008229, 8229393, 8268433, 8313522, 8350372, 8392310, 8392311, 8442257, 8484903, 8572971, 8572972, 8742701, 8742702, 8806409, 8865389, 8884593, 8996692, 9213520, 9304403, 9351788, 9404767, 9495819, 9505078, 9622811, 9634390, 9778578, 9940983, 9940985, 9940986, 9940987, 9940988 ; Device Reference Number AA6312, GTIN 3600040128527, Lot Numbers 7174140, 7187848, 7187849, 7210666, 7278643, 7362305, 7396116, 7447574, 7540089, 7540090, 7585014, 7748371, 7748372, 7748373, 7894950, 7894951, 7894952, 8009485, 8009487, 8149941, 8149944, 8326518, 8392430, 8620093, 8700281, 8710124, 8710125, 8765639, 8765640, 8765642, 8839922, 8854150, 8865380, 8865381, 8931454, 8979454, 8979455, 9008570, 9008577, 9031600, 9077808, 9077809, 9112304, 9112305, 9112307, 9226831, 9226834, 9351873, 9351874, 9351875, 9385381, 9405093, 9420757, 9472413, 9472414, 9530032, 9541014, 9634457, 9634458, 9634459 ; Device Reference Number AA6314, GTIN 3600040128657, Lot Numbers 7161409, 7199153, 7210671, 7270715, 7303239, 7328572, 7387266, 7597841, 7657501, 7748382, 7810282, 7810284, 7810285, 7883527, 7883528, 7883530, 7942127, 8021005, 8021008, 8021010, 8133760, 8133762, 8210455, 8268417, 8326563, 8361161, 8392440, 8431470, 8442240, 8506820, 8506824, 8506827, 8506829, 8592983, 8691234, 8781858, 8948130, 8996706, 9067761, 9090475, 9142760, 9192349, 9192350, 9315900, 9351810, 9351811, 9374199, 9396686, 9404795, 9405052, 9405054, 9405055, 9405056, 9419450, 9419451, 9429239, 9429242, 9429244, 9429245, 9445846, 9498282, 9498284, 9498291, 9520729, 9583957, 9583958, 9591036, 9683091, 9683093, 9683095 ; Device Reference Number AA6316, GTIN 3600040128787, Lot Numbers 7270691, 7270693, 7270695, 7278620, 7338694, 7372033, 7372058, 7416502, 7440888, 7469922, 7502564, 7504465, 7516897, 7531041, 7543264, 7657521, 7657840, 7810293, 7810295, 7810298, 7874664, 7919607, 7919615, 7948692, 8009509, 8087237, 8156954, 8156959, 8229360, 8229362, 8279953, 8279964, 8289295, 8385832, 8431464, 8442233, 8442236, 8459903, 8506902, 8506911, 8592930, 8620035, 8710056, 8749908, 8765594, 8796808, 8829392, 8884465, 9018628, 9031089, 9031090, 9159876, 9177676, 9201834, 9248070, 9304424, 9304425, 9341146, 9341147, 9351832, 9351834, 9362253, 9362254, 9396699, 9396700, 9405074, 9405076, 9405078, 9405080, 9405082, 9420743, 9420747, 9429258, 9429259, 9429260, 9429264, 9446036, 9446037, 9498317, 9498319, 9498321, 9498325, 9498328, 9553608, 9583682, 9583685, 9583686, 9583951, 9597332, 9610632, 9666715, 9666718, 9666722, 9690346, 9711527, 9711528, 9780661 ; Device Reference Number AA6318, GTIN 3600040128916, Lot Numbers 7187853, 7187855, 7187856, 7210690, 7314706, 7338701, 7387255, 7441009, 7494299, 7540104, 7624498, 7624499, 7774405, 7774409, 7774410, 7836009, 7836010, 7873737, 7874459, 7883405, 7894833, 7909021, 7919584, 7919587, 8063418, 8149978, 8210437, 8326549, 8452921, 8519805, 8519806, 8592976, 8656364, 8710067, 8710071, 8884562, 8899156, 8899157, 8931469, 8948191, 8948192, 9018829, 9055044, 9112167, 9192342, 9248025, 9261042, 9341126, 9341127, 9351828, 9351829, 9351830, 9374215, 9405067, 9419447, 9429238, 9445816, 9472419, 9472424, 9472428, 9505200, 9505205, 9505210, 9597325, 9683098 ; Device Reference Number AA6320, GTIN 3600040129043, Lot Numbers 7470031, 7564265, 7564548, 7694153, 7748259, 7859432, 7859433, 7968997, 7968998, 8111929, 8111931, 8350380, 8408104, 8529305, 8620018, 8620019, 8765651, 8796853, 9102809, 9159909, 9177740, 9261105, 9461967, 9634310, 9656805 ; Device Reference Number AA6322, GTIN 3600040129173, Lot Numbers 7278629, 7427155, 7669083, 7883400, 8020995, 8341192, 8529360, 8806425, 8865242, 9077741, 9261332, 9529711 ; Device Reference Number AA6324, GTIN 3600040129302, Lot Numbers 7576086, 8781856 ; Device Reference Number AA6408, GTIN 3600040129548, Lot Numbers 7161423, 7255852, 7314698, 7338705, 7487153, 7494196, 7503941, 7564628, 7859393, 7919509, 8122919, 8296919, 8313327, 8473741, 8498208, 8559595, 8710047, 8763704, 8829389, 8865402, 9520259, 9529677, 9622807, 9634305 ; Device Reference Number AA6410, GTIN 3600040129654, Lot Numbers 7238889, 7328578, 7504537, 7727224, 7875897, 7968423, 8268274, 8279805, 8289260, 8313332, 8385000, 8572855, 8829332, 8911104, 9041019, 9090452, 9112040, 9122839, 9304740, 9351786, 9404766, 9485150, 9585499, 9597191, 9610626, 9622874, 9645254 ; Device Reference Number AA6412, GTIN 3600040129760, Lot Numbers 7469902, 7487192, 7575986, 7616927, 7657453, 7831467, 7850833, 7894943, 7969018, 8063393, 8484904, 8763741, 8865370, 8996597, 9540949, 9597147 ; Device Reference Number AA6414, GTIN 3600040129876, Lot Numbers 7188119, 7199212, 7440901, 7528790, 7576075, 7714004, 7883386, 7987202, 8157026, 8418146, 8506830, 8620089, 8829335, 8884416, 8931371, 9041046, 9404761, 9634304 ; Device Reference Number AA6416, GTIN 3600040129982, Lot Numbers 7350530, 7447481, 7564676, 7942073, 8030404, 8279949, 8289288, 8341195, 8392320, 8442223, 8506815, 8620005, 8691165, 8765660, 8806426, 8884587, 8911347, 8996693, 9112189, 9404768, 9656818, 9690317 ; Device Reference Number AA6418, GTIN 3600040130094, Lot Numbers 7504609, 7774381, 8728852, 8806464, 8839747, 8865340, 8931610, 9018807, 9315892, 9341121, 9405065, 9711524 ; Device Reference Number AA6420, GTIN 3600040130209, Lot Numbers 7405944, 7528861, 7748339, 8097883, 8174966, 8200073, 8221042 ; Device Reference Number AA6422, GTIN 3600040130315, Lot Numbers 7387180, 7494256, 7504533, 7835619, 7998860, 8072753, 8122937, 8289258, 8361032, 8572867, 8724746, 8853703, 8923157, 8931625, 9018923, 9253805, 9315714, 9420936, 9530112, 9683051 ; Device Reference Number AA6424, GTIN 3600040130421, Lot Numbers 7908254, 8097924, 8111874, 8539720, 8700221, 8865324, 9520261, 9711496 ; Device Reference Number AA6C18, GTIN 3600040132005, Lot Numbers 7576074, 7735670, 7835647, 8529292, 8763748, 8865410, 9090707, 9177753 ; Device Reference Number AA6C20, GTIN 3600040132135, Lot Numbers 7362291, 7405869, 7714003, 8679900 ; Device Reference Number AA6C22, GTIN 3600040132265, Lot Numbers 8055955, 8114931, 8114941, 8114965, 9690288 ; Device Reference Number AA6C24, GTIN 3600040132395, Lot Numbers 8114942, 8498233, 8603437, 8865331, 9132275, 9558659 ;
Devices named in this recall
32 devices in the FDA device registry whose identifier appears in the recall's code information.
- Porges Coloplast AA6114Coloplast A/S · CATHETER, RETENTION TYPE, BALLOON1 recall1 death68 reports
- Porges Coloplast AA6116Coloplast A/S · CATHETER, RETENTION TYPE, BALLOON1 recall67 reports
- Porges Coloplast AA6106Coloplast A/S · CATHETER, RETENTION TYPE, BALLOON1 recall29 reports
- Porges Coloplast AA6118Coloplast A/S · CATHETER, RETENTION TYPE, BALLOON1 recall29 reports
- Porges Coloplast AA6112Coloplast A/S · CATHETER, RETENTION TYPE, BALLOON1 recall28 reports
- Porges Coloplast AA6120Coloplast A/S · CATHETER, RETENTION TYPE, BALLOON1 recall22 reports
- Porges Coloplast AA6110Coloplast A/S · CATHETER, RETENTION TYPE, BALLOON1 recall19 reports
- Porges Coloplast AA6316Coloplast A/S · CATHETER, RETENTION TYPE, BALLOON1 recall16 reports
- Porges Coloplast AA6314Coloplast A/S · CATHETER, RETENTION TYPE, BALLOON1 recall13 reports
- Porges Coloplast AA6418Coloplast A/S · CATHETER, RETENTION TYPE, BALLOON1 recall13 reports
- Porges Coloplast AA6108Coloplast A/S · CATHETER, RETENTION TYPE, BALLOON1 recall12 reports
- Porges Coloplast AA6318Coloplast A/S · CATHETER, RETENTION TYPE, BALLOON1 recall9 reports
- Porges Coloplast AA6416Coloplast A/S · CATHETER, RETENTION TYPE, BALLOON1 recall9 reports
- Porges Coloplast AA6414Coloplast A/S · CATHETER, RETENTION TYPE, BALLOON1 recall7 reports
- Porges Coloplast AA6C18Coloplast A/S · CATHETER, RETENTION TYPE, BALLOON1 recall4 reports
- Porges Coloplast AA6122Coloplast A/S · CATHETER, RETENTION TYPE, BALLOON1 recall3 reports
- Porges Coloplast AA6312Coloplast A/S · CATHETER, RETENTION TYPE, BALLOON1 recall3 reports
- Porges Coloplast AA6310Coloplast A/S · CATHETER, RETENTION TYPE, BALLOON1 recall2 reports
- Porges Coloplast AA6320Coloplast A/S · CATHETER, RETENTION TYPE, BALLOON1 recall2 reports
- Porges Coloplast AA6C20Coloplast A/S · CATHETER, RETENTION TYPE, BALLOON1 recall2 reports
- Porges Coloplast AA6308Coloplast A/S · CATHETER, RETENTION TYPE, BALLOON1 recall1 report
- Porges Coloplast AA6410Coloplast A/S · CATHETER, RETENTION TYPE, BALLOON1 recall1 report
- Porges Coloplast AA6124Coloplast A/S · CATHETER, RETENTION TYPE, BALLOON1 recall0 reports
- Porges Coloplast AA6322Coloplast A/S · CATHETER, RETENTION TYPE, BALLOON1 recall0 reports
- Porges Coloplast AA6324Coloplast A/S · CATHETER, RETENTION TYPE, BALLOON1 recall0 reports
- Porges Coloplast AA6408Coloplast A/S · CATHETER, RETENTION TYPE, BALLOON1 recall0 reports
- Porges Coloplast AA6412Coloplast A/S · CATHETER, RETENTION TYPE, BALLOON1 recall0 reports
- Porges Coloplast AA6420Coloplast A/S · CATHETER, RETENTION TYPE, BALLOON1 recall0 reports
- Porges Coloplast AA6422Coloplast A/S · CATHETER, RETENTION TYPE, BALLOON1 recall0 reports
- Porges Coloplast AA6424Coloplast A/S · CATHETER, RETENTION TYPE, BALLOON1 recall0 reports
- Porges Coloplast AA6C22Coloplast A/S · CATHETER, RETENTION TYPE, BALLOON1 recall0 reports
- Porges Coloplast AA6C24Coloplast A/S · CATHETER, RETENTION TYPE, BALLOON1 recall0 reports
Product lines
Other products in the same recall event
The FDA groups related products into one recall event (event 95895).
- Z-0845-2025 · Open, ClassifiedX-Flow prostatectomy catheter, Device Reference Numbers AB6022, AB6024, AB6118, AB6120, AB6122, AB6124, AB6220, AB6222, AB6224, AB6318, AB6320, AB6322, AB6324, AB6418, AB6420, AB64…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page