Porges Coloplast AA6316Recalls & reported problems
Coloplast A/S · FOLYSIL® catheter Tiemann CH FR 16 · Class 2
All models of Porges Coloplast (79)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 16
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 16 adverse event reports that list this exact device (DI 03600040128770), received between Mar 1, 2017 and Mar 24, 2025 — classified as 1 injury, 15 malfunctions.
The most frequently reported specific device problem is “Burst Container or Vessel” (10 reports).
1 FDA recall record name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Recalls naming this device
- Dec 2, 2024Class IIOpen, Classified
A possible sterility issue was detected in Coloplast's facility on some Coloplast products.
Z-0844-2025 · Root cause: Packaging · Recall details · FDA record ↗
Most reported problems
Device problems
- Burst Container or Vessel10
- Failure to Deflate2
- Material Rupture2
- Device Contamination with Chemical or Other Material1
- Fracture1
- Improper or Incorrect Procedure or Method1
Patient problems
- No Clinical Signs, Symptoms or Conditions10
- Hematuria1
- No Consequences Or Impact To Patient1
- No Information1
- No Known Impact Or Consequence To Patient1
- Pain1
- Urinary Tract Infection1
Recent adverse event reports
The 16 most recent of 16 reports, in the reporter's own words as published by the FDA.
MalfunctionMar 24, 2025Improper or Incorrect Procedure or Method; Failure to Deflate; Improper or Incorrect Procedure or Method; Failure to Deflate
ACCORDING TO THE AVAILABLE INFORMATION, THE DOCTOR SUSTAINED A SMALL PUNCTURE WOUND TO THE RIGHT INDEX FINGER.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21676992 ↗MalfunctionJul 24, 2024Burst Container or Vessel; Burst Container or Vessel
ACCORDING TO THE AVAILABLE INFORMATION, DURING PATIENT INDWELLING, THE BALLOON BURST AND FELL OUT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19822786 ↗MalfunctionJul 5, 2024Fracture; Fracture
ACCORDING TO THE AVAILABLE INFORMATION WHEN THE NURSE UNPACKED THE CATHETER, THE DISTAL END SEPARATED FROM THE BODY OF THE CATHETER. A NEW CATHETER WAS USED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19681376 ↗MalfunctionApr 24, 2024Failure to Deflate; Failure to Deflate
ACCORDING TO THE AVAILABLE INFORMATION UPON TESTING, THE BALLOON WAS FILLED WITH WATER AND INFLATED BUT WAS UNABLE TO REMOVE THE WATER AND DEFLATE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19177415 ↗MalfunctionJan 4, 2024Burst Container or Vessel; Burst Container or Vessel
ACCORDING TO THE AVAILABLE INFORMATION, PRIOR TO INSERTION, THE BALLOON TEST WAS DONE WITH 15ML OF STERILE WATER WITHOUT DIFFICULTY. THE PLACEMENT OF THE CATHETER DID NOT PRESENT ANY PROBLEM WITH A GOOD RETURN OF URINE. FOLLOWING INFLATION OF THE BALLOON WITH 15ML OF WATER THE URINE BECAME PINK THEN HEMATURIC WITH BLOOD CLOTS. WHEN REMOVING THE CATHETER THE NURSE OBSERVED THE BALLON BURST. A NEW CATHETER WAS PLACED WITHOUT PAIN SUCCESSFULLY.
Patient problems: Hematuria
MAUDE report 18445224 ↗MalfunctionMar 21, 2023Burst Container or Vessel; Burst Container or Vessel
ACCORDING TO AVAILABLE INFORMATION, THIS DEVICE REQUIRED REMOVAL DUE TO A BALLOON BURST. THE ADMINISTRATIVE TECHNICIAN AT THE MEDICAL FACILITY SAID THE DEVICE BALLOON BURST AFTER PLACEMENT. IT HAPPENED IN LESS THAN A WEEK OF USING THE DEVICE BY THE USER. NO OTHER ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16585762 ↗MalfunctionFeb 3, 2023Burst Container or Vessel; Burst Container or Vessel
ACCORDING TO AVAILABLE INFORMATION, THIS DEVICE REQUIRED REPLACEMENT DUE TO BALLOON BURST. THE BALLOON BURST WHEN INFLATED TO 15 ML AS ADVISED BUT PROVIDER FOUND THAT THE BALLOON WAS ONLY ABLE TO BE INFLATED TO 7 ML WHEN TRIED ON OTHER DEVICES. NO OTHER ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16292124 ↗MalfunctionFeb 3, 2023Burst Container or Vessel; Burst Container or Vessel
ACCORDING TO AVAILABLE INFORMATION, THIS DEVICE REQUIRED REPLACEMENT DUE TO BALLOON BURST. THE BALLOON BURST WHEN INFLATED TO 15 ML AS ADVISED BUT PROVIDER FOUND THAT THE BALLOON WAS ONLY ABLE TO BE INFLATED TO 7 ML WHEN TRIED ON OTHER DEVICES. NO OTHER ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16292062 ↗MalfunctionFeb 2, 2023Material Rupture; Material Rupture
ACCORDING TO AVAILABLE INFORMATION, THIS DEVICE REQUIRED REPLACEMENT DUE TO BALLOON BURST. THE BALLOON BURST WHEN INFLATED TO 15 ML AS ADVISED BUT PROVIDER FOUND THAT THE BALLOON WAS ONLY ABLE TO BE INFLATED TO 7 ML WHEN TRIED ON OTHER DEVICES. NO OTHER ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16282619 ↗MalfunctionDec 29, 2022Device Contamination with Chemical or Other Material; Device Contamination with Chemical or Other Material
ACCORDING TO AVAILABLE INFORMATION, THIS DEVICE WAS NOT ABLE TO BE USED DUE TO CONTAMINATION. THE DEVICE HAD FIXED DIRT ON THE BENT END AND LOOSE DIRT WITHIN THE PACKAGING. NO OTHER ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16070082 ↗MalfunctionDec 17, 2021Burst Container or Vessel; Burst Container or Vessel
ACCORDING TO THE AVAILABLE INFORMATION, THE PATIENT LOST HER CATHETER BECAUSE THE BALLOON BURST. A NEW CATHETER WAS PLACED WITHOUT DIFFICULTY, THERE WAS NO PAIN FOR THE PATIENT AND THE PATIENT HAD A GOOD GENERAL STATE OF HEALTH.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13027767 ↗MalfunctionFeb 18, 2021Burst Container or Vessel; Burst Container or Vessel
ACCORDING TO THE AVAILABLE INFORMATION, WHEN THE BALLOON IS INFLATED AFTER BEING PUT IN PLACE, THE BALLOON BURSTS. THIS OCCURRED FOUR TIMES WITH CATHETERS FROM THE SAME LOT.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 11342541 ↗InjuryJan 5, 2021Burst Container or Vessel; Burst Container or Vessel
ACCORDING TO THE AVAILABLE INFORMATION, THE PATIENT EXPERIENCED FOUR CATHETER BURSTS. THE PATIENT HAD BEEN SELF-INSERTING FOLEY CATHETERS FOR FIVE YEARS. HE UTILIZED A PREFILLED 15CC SYRING FOR INFLATION, AND A STATLOCK TO HOLD THE CATHETER IN PLACE. THE CATHETERS BURST AFTER LESS THAN 1 DAY IN PLACE. THE PATIENT STATED HE HAD NOT VISITED THE DOCTOR OR EMERGENCY ROOM FOR ANY TREATMENT, BUT ON ALL FOUR OCCASIONS, IT APPEARED ALL THE PIECES CAME OUT. THE PATIENT NOTED HE HAD EXPERIENCED BLADDER INFECTIONS AS A RESULT OF THE BURSTS BECAUSE THE STERILE WATER FROM THE BALLOON ENTERED HIS BLADDER. THE PATIENT WAS NOT EXPERIENCING ANY ADDITIONAL COMPLICATIONS AT THE TIME THIS INFORMATION WAS REPORTED. ALL FOUR CATHETERS THAT BURST HAD THE SAME LOT NUMBER.
Patient problems: Urinary Tract Infection
MAUDE report 11117988 ↗MalfunctionJun 26, 2020Burst Container or Vessel; Burst Container or Vessel
ACCORDING TO THE AVAILABLE INFORMATION, WITHIN A FEW HOURS INDWELLING, THE CATHETER FELL OFF. BALLOON BURST WAS FOUND.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10202398 ↗MalfunctionDec 7, 2018Burst Container or Vessel; Burst Container or Vessel
ACCORDING TO THE AVAILABLE INFORMATION, THE BALLOON BURST. THERE WERE NO CLINICAL CONSEQUENCE.
Patient problems: No Information
MAUDE report 8143614 ↗MalfunctionMar 1, 2017Material Rupture
END USER BEGAN EXPERIENCE A HIGH LEVEL OF BLADDER PAIN. HE SOUGHT TREATMENT AT THE ER. THE CAUSE OF THE PAIN WAS RUPTURE OF THE BALLOON. THE IC IS REPLACED EVERY 4 WEEKS AT HIS UROLOGISTS' OFFICE. THE IC WAS DUE TO BE REPLACED ON TUESDAY (B)(6) 2017. UPON EXAMINATION AT THE ER, THE NURSE ATTEMPTED TO REDUCE THE BALLOON, BUT SINCE IT RUPTURED, THERE WAS NOTHING TO REDUCE. THE CATHETER WAS REMOVED AND REPLACED WITH A NEW ONE. THE ER DEPARTMENT CONDUCTED "A SCAN" (ULTRASOUND?) AND NO EVIDENCE OF MATERIAL WAS NOTED. PATIENT WAS DISCHARGED AFTER 4 HOURS.
Patient problems: Pain
MAUDE report 6367968 ↗
Other CATHETER, RETENTION TYPE, BALLOON devices
- Bardex® All-Silicone Foley Catheter 165816C. R. Bard, Inc. · CATHETER, RETENTION TYPE, BALLOON512 reports
- Bardex® Lubri-Sil® All-Silicone Foley Catheter 175816C. R. Bard, Inc. · CATHETER, RETENTION TYPE, BALLOON476 reports
- Bardex® Lubri-Sil® I.C. All-Silicone Foley Catheter 1758SI16C. R. Bard, Inc. · CATHETER, RETENTION TYPE, BALLOON1 death383 reports
- Bardex® Lubri-Sil® All-Silicone Foley Catheter 175812C. R. Bard, Inc. · CATHETER, RETENTION TYPE, BALLOON234 reports
- Bardex® Lubri-Sil® All-Silicone Temp Sensing Foley Catheter 119216MC. R. Bard, Inc. · CATHETER, RETENTION TYPE, BALLOON167 reports
- Bardex® Lubri-Sil® All-Silicone Foley Catheter 175814C. R. Bard, Inc. · CATHETER, RETENTION TYPE, BALLOON153 reports
- Bardex® All-Silicone Foley Catheter 165812C. R. Bard, Inc. · CATHETER, RETENTION TYPE, BALLOON151 reports
- SureStep™ Foley Tray System Bard® Lubri-Sil® Foley Catheter Tray A942216C. R. Bard, Inc. · CATHETER, RETENTION TYPE, BALLOON2 recalls46 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page