1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
13
Deaths reported
0
Injuries reported
1
Recalls naming it
1

Summary

The FDA's MAUDE database contains 13 adverse event reports that list this exact device (DI 03600040130087), received between Feb 26, 2019 and Oct 4, 2023 — classified as 1 injury, 12 malfunctions.

The most frequently reported specific device problem is “Burst Container or Vessel” (5 reports).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Dec 2, 2024Class IIOpen, Classified

    A possible sterility issue was detected in Coloplast's facility on some Coloplast products.

    Z-0844-2025 · Root cause: Packaging · Recall details · FDA record ↗

Most reported problems

Device problems

  • Burst Container or Vessel5
  • Deflation Problem3
  • Difficult to Remove2
  • Fluid/Blood Leak2
  • Break1
  • Detachment of Device or Device Component1
  • Fracture1
  • Inflation Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions5
  • No Information4
  • No Known Impact Or Consequence To Patient2
  • Pain2
  • Foreign Body In Patient1

Recent adverse event reports

The 13 most recent of 13 reports, in the reporter's own words as published by the FDA.

  • MalfunctionOct 4, 2023Deflation Problem; Inflation Problem; Deflation Problem; Inflation Problem

    ACCORDING TO AVAILABLE INFORMATION, THIS DEVICE WAS DIFFICULT TO INFLATE AND DEFLATE. DURING THE INFLATION TEST BEFORE DEVICE PLACEMENT, IT WAS DIFFICULT TO INJECT AND REMOVE WATER FROM THE BALLOON. NO OTHER ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17874035 ↗
  • MalfunctionJul 1, 2022Break; Break

    ACCORDING TO AVAILABLE INFORMATION, THIS DEVICE REQUIRED REPLACEMENT DUE TO A BALLOON BURST. THE CATHETER WAS PLACED AFTER INFLATION TEST AND THE BALLOON BURST WITH 15 CC. NO OTHER ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14887517 ↗
  • MalfunctionMay 11, 2022Fluid/Blood Leak; Fluid/Blood Leak

    ACCORDING TO AVAILABLE INFORMATION, THIS DEVICE EXPERIENCED A MALFUNCTION DUE TO FLUID LEAKAGE AT THE TRANSITION OF THE SILICONE CATHETER TO THE BLOCKING VALVE. NO OTHER ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14364145 ↗
  • MalfunctionMay 11, 2022Fluid/Blood Leak; Fluid/Blood Leak

    ACCORDING TO AVAILABLE INFORMATION, THIS DEVICE EXPERIENCED A MALFUNCTION DUE TO FLUID LEAKAGE AT THE TRANSITION OF THE SILICONE CATHETER TO THE BLOCKING VALVE. NO OTHER ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14364066 ↗
  • MalfunctionOct 27, 2021Deflation Problem; Difficult to Remove; Deflation Problem; Difficult to Remove

    ACCORDING TO THE AVAILABLE INFORMATION, THE BALLOON DID NOT COMPLETELY DEFLATE DURING WITHDRAWAL, CREATING A "PLEATED BLISTERING" WHICH INCREASED THE DIAMETER AND "ADDED RELIEFS" MAKING WITHDRAWAL DIFFICULT AND PAINFUL FOR THE PATIENT.

    Patient problems: Pain

    MAUDE report 12704969 ↗
  • MalfunctionOct 19, 2021Deflation Problem; Difficult to Remove; Deflation Problem; Difficult to Remove

    ACCORDING TO THE AVAILABLE INFORMATION, THE BALLOON DID NOT COMPLETELY DEFLATE DURING WITHDRAWAL. IT CREATED A "PLEATED BLISTERING" WHICH INCREASED THE DIAMETER AND "ADD RELIEFS" MAKING THE WITHDRAWAL DIFFICULT AND PAINFUL FOR THE PATIENT.

    Patient problems: Pain

    MAUDE report 12661178 ↗
  • MalfunctionAug 24, 2021Insufficient Device Problem Information; Insufficient Device Problem Information

    ACCORDING TO THE AVAILABLE INFORMATION, THE CATHETER FELL OFF AND A BALLOON BURST WAS DISCOVERED AFTER 1 DAY INDWELLING.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12362504 ↗
  • InjuryJul 28, 2021Fracture; Detachment of Device or Device Component; Fracture; Detachment of Device or Device Component

    ACCORDING TO THE AVAILABLE INFORMATION, AFTER INSERTION OF THE CATHETER, THE DISTAL TIP WAS NO LONGER IN PLACE. IT "PROBABLY DETACHED" AND WAS NOT FOUND IN THE PATIENT AFTER CYSTOSCOPY.

    Patient problems: Foreign Body In Patient

    MAUDE report 12240825 ↗
  • MalfunctionMay 15, 2020Burst Container or Vessel; Burst Container or Vessel

    ACCORDING TO THE AVAILABLE INFORMATION, THE CATHETER FELL OFF AFTER 1 DAY INDWELLING. BALLOON BURST WAS FOUND. THE LOT NUMBER OF THE RED VALVE WAS PROVIDED, 6619792.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 10062379 ↗
  • MalfunctionMar 1, 2019Burst Container or Vessel; Burst Container or Vessel

    ACCORDING TO THE AVAILABLE INFORMATION, WHEN REFILLING THE BALLOON, THE BALLOON BURST,THE RED VALVE IS LEAKING.

    Patient problems: No Information

    MAUDE report 8382030 ↗
  • MalfunctionMar 1, 2019Burst Container or Vessel; Burst Container or Vessel

    ACCORDING TO THE AVAILABLE INFORMATION, WHEN REFILLING THE BALLOON, THE BALLOON BURST, THE RED VALVE IS LEAKING.

    Patient problems: No Information

    MAUDE report 8382026 ↗
  • MalfunctionFeb 26, 2019Burst Container or Vessel; Burst Container or Vessel

    ACCORDING TO THE AVAILABLE INFORMATION, WHEN REFILLING THE BALLOON, THE BALLOON BURSTS. NO HARM TO USER.

    Patient problems: No Information; No Known Impact Or Consequence To Patient

    MAUDE report 8371537 ↗
  • MalfunctionFeb 26, 2019Burst Container or Vessel; Burst Container or Vessel

    ACCORDING TO THE AVAILABLE INFORMATION, WHEN REFILLING THE BALLOON, THE BALLOON BURSTS. NO HARM TO USER.

    Patient problems: No Information

    MAUDE report 8371536 ↗